HI-TORQUE POWERTURN 1044595Recalls & reported problems

Abbott Vascular Inc. · HI-TORQUE POWERTURN Guide Wire w/Hydrophilic Coating .014" Straight Tip 300 cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648173998), received between Sep 14, 2018 and Sep 14, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Difficult to Remove1
  • Improper or Incorrect Procedure or Method1
  • Stretched1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 14, 2018Difficult to Remove; Stretched; Improper or Incorrect Procedure or Method; Deformation Due to Compressive Stress; Difficult to Remove; Stretched; Improper or Incorrect Procedure or Method; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE PEDAL ARTERY WITH MODERATE TORTUOSITY AND MODERATE CALCIFICATION. A HT POWERTURN GUIDEWIRE WAS ADVANCED TO THE LESION WITH NO REPORTED RESISTANCE. WHEN THE GUIDEWIRE REACHED THE LESION AT THE SITE OF A COLLATERAL VESSEL, A KINK OCCURRED. AN UNSPECIFIED BALLOON DILATATION CATHETER (BDC) WAS ADVANCED OVER THE WIRE AND LESION WAS DILATED. ON REMOVAL OF THE BDC, RESISTANCE WITH THE GUIDEWIRE WAS NOTED. AS A RESULT, THE GUIDEWIRE AND BDC HAD TO BE REMOVED AS A SINGLE UNIT. ON INSPECTION OF THE GUIDEWIRE OUTSIDE OF THE ANATOMY, THE COILS WERE NOTED TO BE UNRAVELED. THE CORE OF THE GUIDEWIRE WAS INTACT AND THE DEVICE REMAINED IN ONE PIECE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN ADDITIONAL NON-ABBOTT GUIDEWIRE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7878470 ↗