Skip to content

HI-TORQUE CONNECT 1012589Recalls & reported problems

Abbott Vascular Inc. · Hi-Torque Connect Guide Wire .018" 300 cm · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648174995), received between Aug 25, 2022 and Aug 25, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Advance1
  • Difficult to Remove1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2022Difficult to Remove; Material Separation; Difficult to Advance; Difficult to Remove; Material Separation; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE 90% STENOSED, HEAVILY TORTUOUS POPLITEAL ARTERY. WHEN ADVANCING THE CONNECT 18 GUIDE WIRE TO THE LOWER LIMB ARTERIES, RESISTANCE WAS NOTED, AND THE TIP OF THE GUIDE SEPARATED IN THE GUIDE CATHETER. RESISTANCE WAS NOTED DURING REMOVAL, SO THE GUIDE WIRE AND GUIDE CATHETER WERE REMOVED, WITH FORCE, TOGETHER, AS A SINGLE UNIT. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER CONNECT 18 WAS USED TO CONTINUE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15291612 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →