HI-TORQUE CONNECT 1012592Recalls & reported problems
Abbott Vascular Inc. · Hi-Torque Connect Flex Guide Wire .018" 300 cm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08717648175022), received between Apr 13, 2020 and Jul 4, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Material Separation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation4
- Difficult to Advance3
- Difficult to Remove3
- Material Frayed2
- Entrapment of Device1
- Improper or Incorrect Procedure or Method1
- Material Too Soft/Flexible1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 4, 2024Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT LEFT PULMONARY ARTERY. WHEN ATTEMPTING TO ADVANCE THE CARDIOMEMS DELIVERY SYSTEM OVER THE .018 CONNECT FLEX GUIDE WIRE RESISTANCE WAS FELT AND THE SYSTEM WAS NOT ABLE TO ADVANCE FURTHER. THE WIRE AND DELIVERY SYSTEM WERE REMOVED. THE GUIDE WIRE WAS NOTED TO HAVE COATING SHEARED OFF. A NEW .018 GUIDEWIRE WAS USED ALONG WITH CARDIOMEMS DEVICE AND THE SENSOR WAS DEPLOYED SUCCESSFULLY. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19674340 ↗MalfunctionSep 5, 2023Material Frayed; Material Frayed
IT WAS REPORTED THE PROCEDURE WAS TO TREAT THE SUPERFICIAL FEMORAL ARTERY POPLITEAL WITH LASER ATHERECTOMY. DURING THE PROCEDURE VIA ANGIOGRAM VISUALIZATION, THE 018 HT CONNECT FLEX GUIDE WIRE POLYMER WAS NOTED TO BE FRAYED. A NEW CONNECT FLEX GUIDE WIRE WAS USED AND THE SAME ISSUE OCCURRED. A NON-ABBOTT WIRE WAS USED TO COMPLETE THE PROCEDURE. NO COATING WAS LEFT IN THE ANATOMY. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY DURING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17681516 ↗MalfunctionSep 5, 2023Material Frayed; Material Frayed
IT WAS REPORTED THE PROCEDURE WAS TO TREAT THE SUPERFICIAL FEMORAL ARTERY POPLITEAL WITH LASER ATHERECTOMY. DURING THE PROCEDURE VIA ANGIOGRAM VISUALIZATION, THE 018 HT CONNECT FLEX GUIDE WIRE POLYMER WAS NOTED TO BE FRAYED. A NEW CONNECT FLEX GUIDE WIRE WAS USED AND THE SAME ISSUE OCCURRED. A NON-ABBOTT WIRE WAS USED TO COMPLETE THE PROCEDURE. NO COATING WAS LEFT IN THE ANATOMY. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY DURING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17681143 ↗MalfunctionMar 14, 2023Difficult to Remove; Material Separation; Difficult to Advance; Difficult to Remove; Material Separation; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE SUPERFICIAL FEMORAL ARTERY THAT WAS RESTENOSED AND NON CALCIFIED. A CONNECT FLEX WIRE TIP BROKE OFF OFF IN THE RIGHT SFA IN THE NON-ABBOTT STENT OCCLUSION (PREVIOUSLY IMPLANTED IN A SEPARATE PROCEDURE AT AN UNKNOWN DATE). THE PROCEDURE WAS DELAYED FOR AN HOUR AS HE HAD TO USE A SNARE TO RETRIEVE THE BROKEN PIECE. THERE WAS RESISTANCE IN THE LESION WHILE TRYING TO CROSS IT AND COME APART DUE TO RESISTANCE DURING WITHDRAWAL. THE PROCEDURE WAS COMPLETED BY CROSSING WITH A DIFFERENT WIRE WHERE THEY THEN LAYERED THE IN-STENT AREA AND USED A DRUG COATED BALLOON. THE PATIENT WAS SUCCESSFULLY TREATED WITH NO ADVESE SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16536964 ↗MalfunctionSep 7, 2022Difficult to Remove; Material Separation; Difficult to Remove; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A POSTERIOR TIBIAL ARTERY. DURING AN ATTEMPT TO RECANALIZE THE POSTERIOR TIBIAL ARTERY, THE TIP OF THE 018 HT CONNECT FLEX GUIDE WIRE (GW) SEPARATED IN THE ANATOMY. THE GUIDE WIRE WAS REMOVED WITH RESISTANCE. A SNARE WAS USED TO REMOVE THE TIP FROM THE ANATOMY. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. A NON-ABBOTT GUIDE WIRE AND ARMADA BALLOON DILATATION CATHETER (BDC) WERE USED TO SUCCESSFULLY COMPETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15368459 ↗InjuryMay 21, 2021Difficult to Remove; Material Separation; Improper or Incorrect Procedure or Method; Difficult to Advance; Difficult to Remove; Material Separation; Improper or Incorrect Procedure or Method; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE LEFT POSTERIOR TIBIAL ARTERY. A CONNECT FLEX GUIDE WIRE WAS BEING USED WHEN THE TIP SEPARATED. THE SEPARATED TIP WAS EMBEDDED IN THE VESSEL USING AN UNSPECIFIED STENT. THERE WAS NO ADVERSE PATIENT SEQUELAE AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 11866712 ↗InjuryApr 13, 2020Entrapment of Device; Material Separation; Material Too Soft/Flexible; Entrapment of Device; Material Separation; Material Too Soft/Flexible
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A TIBIAL LESION. A 0.18 HI-TORQUE CONNECT FLEX PERIPHERAL GUIDE WIRE WAS BEING ADVANCED WHEN IT PROLAPSED INSIDE THE ANATOMY AT THE TARGET LESION. THE PHYSICIAN ATTEMPTED TO REMOVE THE GUIDE WIRE, BUT THE TIP BECAME STUCK IN A COLLATERAL VESSEL. RESISTANCE WAS NOTED AND THE TIP OF THE GUIDE WIRE BROKE OFF AND REMAINS IN THE SMALL COLLATERAL VESSEL IN THE PATIENT. THE PROXIMAL END OF THE GUIDE WIRE WAS REMOVED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A COMMAND 14 GUIDEWIRE. THERE WAS NO ADVERSE PATIENT SEQUELA OR A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 9955724 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page