HI-TORQUE VERSATURN 1013317JRecalls & reported problems
Abbott Vascular Inc. · HI-TORQUE VersaTurn F Guide Wire Fully Coated .014 J Tip 190 cm · Class 2
All models of HI-TORQUE VERSATURN (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08717648198472), received between Jul 31, 2018 and Sep 17, 2026 — classified as 5 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation4
- Component Misassembled2
- Entrapment of Device2
- Break1
- Difficult to Advance1
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
- Peeled/Delaminated1
Patient problems
- Foreign Body In Patient4
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Vessel Or Plaque, Device Embedded In1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2026Difficult to Remove; Material Separation
It was reported that the HT Versaturn guidewire was used in a 100% in-stent restenosis in the left circumflex coronary artery. The guidewire was able to cross the lesion; however, the guidewire broke during removal. A snare was used to successfully retrieve the separated segment of the wire. There were no adverse patient effects. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50802915 ↗MalfunctionApr 3, 2024Break; Difficult to Advance; Break; Difficult to Advance
IT WAS REPORTED THAT THE DRAGONFLY OPTIS IMAGING CATHETER WAS USED AND AFTER OPTICAL COHERENCE TOMOGRAPHY (POST SOFT RADIOPAQUE PORTION), THE VERSATURN GUIDE WIRE WAS ATTEMPTED TO BE RE-ADVANCED IN ANOTHER VESSEL AND COULD NOT BE ADVANCED. THE GUIDE WIRE WAS REMOVED INDEPENDENTLY AND WAS NOTED TO HAVE UNRAVELED BUT WAS STILL INTACT. ANOTHER VERSATURN GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ISSUE WITH THE DRAGONFLY OPTIS IMAGING CATHETER. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. SUBSEQUENT TO THE INITIALLY FILED REPORT, THE FOLLOWING INFORMATION WAS NOTED UPON FURTHER REVIEW: THE GUIDE WIRE WAS REMOVED INDEPENDENTLY AND WAS NOTED TO HAVE UNRAVELED (PAST SOFT RADIOPAQUE PORTION) BUT WAS STILL INTACT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19037789 ↗InjuryJul 13, 2023Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY CALCIFIED, MILDLY TORTUOUS LEFT ANTERIOR DESCENDING ARTERY. DURING USE WITH THE 014 HT VERSATURN GUIDE WIRE, IT WAS NOTED THAT THE TEFLON COATING WAS PEELING AND COMING OFF. ANOTHER NON-ABBOTT GUIDE WIRE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT SEQUELA REPORTED AND NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 17317559 ↗MalfunctionDec 28, 2020Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE MODERATELY CALCIFIED, MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING CORONARY ARTERY. DURING USE, AN UNKNOWN MARKER WAS FOUND ON A VERSATURN F GUIDE WIRE. UNSPECIFIED DEVICES WERE USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11083853 ↗MalfunctionDec 1, 2020Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MODERATELY CALCIFIED, MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING ARTERY. AN UNKNOWN MARKER WAS SEEN ON THE VERSATURN F GUIDE WIRE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10928642 ↗InjuryAug 13, 2020Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT PRESENTED EXPERIENCING A NON-ST-ELEVATION MYOCARDIAL INFARCTION. THE LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY WAS 99% OCCLUDED. THE DIAGONAL BRANCH WAS 70% OCCLUDED. THE VERSATURN GUIDE WIRE WAS ADVANCED INTO THE DIAGONAL ARTERY TO KEEP THE PATHWAY TO ARTERY OPEN AND A NON-ABBOTT DRUG ELUTING STENT (DES) WAS IMPLANTED IN THE LAD, JAILING THE VERSATURN GUIDE WIRE. DURING REMOVAL OF THE GUIDE WIRE, THE DISTAL PORTION OF THE WIRE BROKE OFF. A NON-ABBOTT STENT WAS IMPLANTED IN THE DIAGONAL TO TREAT THE LESION AND TO STABILIZE THE DISTAL PORTION OF THE SEPARATED WIRE. THE PROXIMAL PORTION OF THE SEPARATED GUIDE WIRE WAS EMBEDDED INTO THE VESSEL WALL OF THE LAD. THE FINAL OUTCOME OF THE PROCEDURE WAS GOOD. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient; Vessel Or Plaque, Device Embedded In
MAUDE report 10405202 ↗InjuryNov 7, 2018Entrapment of Device; Material Separation; Entrapment of Device; Material Separation
IT WAS REPORTED THAT DURING A LEFT ANTERIOR DESCENDING (LAD) ARTERY INTERVENTION, THE VERSATURN GUIDE WIRE WAS POSITIONED IN THE DIAGONAL BRANCH AS A VESSEL STABILIZER. A NON-ABBOTT STENT WAS IMPLANTED AND POST-DILATATION WAS COMPLETED. DURING REMOVAL OF THE VERSATURN WITHOUT RESISTANCE, THE GUIDE WIRE SEPARATED DISTALLY AND WAS ENTRAPPED UNDER THE IMPLANTED STENT AND SLIGHTLY PROTRUDING INTO THE DIAGONAL BRANCH. ANOTHER NON-ABBOTT STENT WAS IMPLANTED TO STABILIZE THE DETACHED PORTION OF THE VERSATURN WIRE. THE PATIENT WAS REPORTED AS OK AND THE OUTCOME REPORTED AS NORMAL. THE PATIENT WAS DISCHARGED THE NEXT DAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 8048496 ↗InjuryJul 31, 2018Entrapment of Device; Material Separation; Improper or Incorrect Procedure or Method; Entrapment of Device; Material Separation; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A DE NOVO LESION, BELOW THE BIFURCATION OF THE MARGINAL AND LEFT ANTERIOR DESCENDING CORONARY ARTERIES. ON REMOVAL OF A HIGH TORQUE VERSATURN GUIDEWIRE, RESISTANCE WITH A NON-ABBOTT STENT WAS FELT. THE GUIDEWIRE WAS REMOVED WITH FORCE AND A SEPARATION OCCURRED AT THE DISTAL RADIOPAQUE PORTION OF THE WIRE. THE TIP OF THE GUIDEWIRE REMAINED CAUGHT BY THE STENT. RETRIEVAL OF THE SEPARATED PORTION OF THE GUIDEWIRE WITH AN ADDITIONAL UNSPECIFIED WIRE WAS ATTEMPTED BUT FAILED. THE SEPARATED PORTION OF THE GUIDEWIRE REMAINED IN THE ANATOMY. THE PATIENT CONDITION WAS REPORTED TO BE GOOD. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 7735813 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page