MITRACLIP CDS0601-XTRRecalls & reported problems

Abbott Vascular Inc. · MitraClip XTR Clip Delivery System · Class 3 · Implant

All models of MitraClip (19) · MITRACLIP (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Abbott MitraClip XTR clip delivery system appears in 878 adverse event reports, including 30 deaths and 615 injuries. Many describe the valve leak not being fully reduced, clips getting caught or moving unexpectedly, and tissue tears, with some mentioning leaflets detaching. One open recall, started in September 2022, concerns clips that may fail to lock, which can let severe valve leakage return.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
875
Deaths reported
30
Injuries reported
614
Recalls naming it
1

Summary

The FDA's MAUDE database contains 875 adverse event reports that list this exact device (DI 08717648226366), received between Oct 10, 2018 and Jul 31, 2026 — classified as 30 deaths, 614 injuries, 231 malfunctions.

The most frequently reported specific device problem is “Incomplete Coaptation” (344 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 3 reportsFeb 2024: 5 reportsMar 2024: 4 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 6 reportsSep 2024: 2 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 3 reportsJan 2025: 6 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 4 reportsJul 2025: 3 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 4 reportsDec 2025: 3 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 4 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 8, 2022Class IIOpen, Classified

    An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

    Z-0115-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incomplete Coaptation344
  • Adverse Event Without Identified Device or Use Problem154
  • Unintended Movement99
  • Difficult to Open or Close92
  • Off-Label Use83
  • Difficult or Delayed Positioning62
  • Difficult to Remove59
  • Device Damaged by Another Device53

Patient problems

  • No Consequences Or Impact To Patient301
  • Mitral Regurgitation187
  • Tissue Damage132
  • Mitral Valve Insufficiency/ Regurgitation87
  • Unspecified Tissue Injury69
  • Dyspnea53
  • Foreign Body In Patient38
  • Mitral Valve Stenosis34

Recent adverse event reports

The 25 most recent of 875 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2026Entrapment of Device

    It was reported that on (b)(6)2026, a patient presented with grade 5 degenerative mitral regurgitation (MR) and a prolapsed posterior leaflet for a MitraClip procedure. During the procedure, a MitraClip was advanced within the patient. While positioning the clip it became caught on the chordae tendinea. Troubleshooting was performed and the clip remained entraped on the leaflets when tissue damage was visualized. The procedure was discontinued, the MR remained at grade 5. There were no clinically significant delays reported.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 50101260 ↗
  • InjuryJul 20, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that on (b)(6) 2026, a patient presented with grade 4+ degenerative mitral regurgitation (MR) for a MitraClip procedure. During the procedure, a MitraClip was advanced within the patient. During grasping attempts the MitraClip XTR caused a tear on the posterior leaflet. The Mitraclip was removed from the patient and the procedure was discontinued without implanting any MitraClips. The final MR remained at grade 4+. There were no clinically significant delays reports.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25871292 ↗
  • InjuryJul 7, 2026Incomplete Coaptation; Device Damaged by Another Device

    It was reported that on (b)(6) 2026, a patient presented with grade 4+ degenerative mitral regurgitation (MR), posterior leaflet flail, and a prolapsed posterior leaflet for a MitraClip procedure. During the procedure, a MitraClip was successfully implanted. A second MitraClip was then selected to further reduce the MR. When the second MitraClip was placed on the leaflet, the first MitraClip had a single leaflet detachment. The first MitraClip was no longer attached to the posterior leaflet. The second MitraClip was removed from the patient and the procedure was discontinue. The final MR remained at grade 4+. There were no adverse patient sequela or clinically significant delays reported.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 25740635 ↗
  • InjuryJun 23, 2026Difficult to Open or Close

    It was reported that on (b)(6) 2026, a patient presented with grade 4+ functional mitral regurgitation (MR) for a MitraCli procedure. During the procedure, a MitraClip was advanced within the patient. While testing the clip, it was identified that the clip would not close. Troubleshooting was performed unsuccessfully. The procedure was discontinued, the final MR remained at grad 4+. There were no adverse patient effects or clinically significant delays reported.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25600369 ↗
  • InjuryJun 3, 2026Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2026, A PATIENT PRESENTED WITH GRADE 4 FUNCTIONAL MITRAL REGURGITATION (MR) AND A RESTRICTED POSTERIOR LEAFLET FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, TWO MITRACLIPS WERE IMPLANTED SUCCESSFULLY. THE PROCEDURE WAS DISCONTINUED, THE FINAL MR WAS REDUCED TO GRADE 1. THERE WERE NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAYS REPORTED. IT WAS REPORTED THAT ON (B)(6) 2026, A FOLLOW-UP ECHOCARDIOGRAM WAS PERFORMED AND IT WAS IDENTIFIED THAT THE PATIENTS MR HAD INCREASED TO GRADE 3. THE PATIENT WAS TRANSITIONED INTO A PROCEDURE FOR A SECONDARY MITRACLIP PROCEDURE. DURING THE PROCEDURE, IT WAS IDENTIFIED THAT ONE OF THE PREVIOUSLY IMPLANTED CLIP HAD A SINGLE LEAFLET DETACHMENT (SLDA) AND WAS NO LONGER ATTACHED TO THE ANTERIOR LEAFLET. AN ADDITIONAL CLIP WAS IMPLANTED BESIDE THE SLDA CLIP FOR STABILIZATION. THE PROCEDURE WAS DISCONTINUED, THE FINAL MR WAS REDUCED TO GRADE 1. THERE WERE NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAYS REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25382848 ↗
  • InjuryApr 21, 2026Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2026, A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT DEGENERATIVE MITRAL REGURGITATION (MR) WITH A GRADE OF 4+ ON A PATIENT WITH A PROLAPSED POSTERIOR LEAFLET, THICK LEAFLETS, A CALCIFIED VALVE ANNULUS, AND A PREVIOUS TRANSCATHETER AORTIC VALVE IMPLANT (TAVI) SURGERY. A STEERABLE GUIDE CATHETER (SGC) AND A MITRACLIP XTR WERE INSERTED WITHOUT ISSUES. HOWEVER, LIMITED BENDING MANIPULATION WITH THE SGC OCCURRED, WHICH RESULTED IN DIFFICULTY WITH POSITIONING OF THE SGC AND UNEVEN CLAMPING WITH THE MITRACLIP XTR. THE MITRACLIP XTR WAS ULTIMATELY DEPLOYED ON THE MITRAL VALVE, REDUCING MR TO A GRADE OF 1+. AFTER THE SGC WAS REMOVED FROM THE PATIENT, THE SGC WAS TESTED AND THE SGC TIP WAS BENDING NORMALLY. ON (B)(6) 2026, ONE WEEK POST PROCEDURE, THE PATIENT DEVELOPED CHEST TIGHTNESS. AN ULTRASOUND WAS PERFORMED AND REVEALED THAT THE CLIP HAD DETACHED FROM THE POSTERIOR LEAFLET AND REMAINED ATTACHED TO THE ANTERIOR LEAFLET (SINGLE LEAFLET DEVICE ATTACHMENT/SLDA), CAUSING MR TO INCREASE TO A GRADE OF 4+. THE PATIENT WAS REPORTED TO BE STABLE.

    Patient problems: Angina; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 24942067 ↗
  • InjuryApr 16, 2026Unintended Movement

    IT WAS REPORTED THAT ON (B)(6) 2026, A PATIENT PRESENTED WITH MITRAL REGURGITATION (MR) FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, A MITRACLIP WAS SUCCESSFULLY PLACED ON THE LEAFLETS AND DEPLOYED. AFTER DEPLOYMENT, THE CLIP OPENED TO APPROXIMATELY 30 DEGREES. THE PROCEDURE WAS DISCONTINUED. THERE WERE NO ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAYS REPORTED. ON (B)(6) 2026, A TRANSTHORACIC ECHOCARDIOGRAM (TTE) WAS PERFORMED ON THE PATIENT AND IT WAS IDENTIFIED THAT THE MR RETURNED. THE PATIENT'S HEART RATE AND BLOOD PRESSURE REMAINED STABLE, THE PATIENT WAS TREATED WITH MEDICATION AND REMAINS HOSPITALIZED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 24905174 ↗
  • MalfunctionApr 14, 2026Difficult to Open or Close

    IT WAS REPORTED THAT ON (B)(6) 2026, A PATIENT PRESENTED WITH MITRAL REGURGITATION (MR) FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, THERE WAS A LEAK IN THE STEERABLE GUIDE CATHETER (SGC) FROM THE HEMOSTASIS VALVE. THE SGC WAS REMOVED FROM THE PROCEDURE AND A REPLACEMENT SGC WAS SELECTED. A MITRACLIP WAS ADVANCED WITHIN THE PATIENT AND ATTEMPTED TO BE PLACED ON THE LEAFLETS. IT WAS IDENTIFIED THAT THE POSTERIOR LEAFLET WOULD NOT DESCEND. TROUBLESHOOTING WAS PERFORMED UNSUCCESSFULLY AND THE CLIP WAS ATTEMPTED TO BE REMOVED FROM THE PROCEDURE. IT WAS NOT POSSIBLE TO REMOVE THE CLIP AND THE PATIENT WAS CONVERTED TO SURGERY. THERE WERE NO CLINICALLY SIGNIFICANT DELAYS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873587 ↗
  • MalfunctionApr 8, 2026Leak/Splash

    IT WAS REPORTED THAT ON (B)(6) 2026, A PATIENT PRESENTED WITH GRADE 4+ DEGENERATIVE MITRAL REGURGITATION (MR) FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, A MITRACLIP WAS ADVANCED WITHIN THE STEERABLE GUIDE CATHETER (SGC) WHEN AIR BUBBLES WERE VISUALIZED WITHIN THE SGC. TROUBLESHOOTING WAS PREFORMED AND IT WAS IDENTIFIED THAT THE HEMOSTATIC VALVE WAS DAMAGED. THE MITRACLIP AND SGC WAS REMOVED FROM THE PATIENT, REPLACEMENT DEVICES WERE SELECTED. THE REPLACEMENT MITRACLIP WAS SUCCESSFULLY IMPLANTED. THE FINAL MR WAS REDUCED TO <1. THERE WERE NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821755 ↗
  • InjuryJan 29, 2026Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2026, A PATIENT PRESENTED WITH GRADE 4+ MIXED MITRAL REGURGITATION (MR), THICK LEAFLETS WITH SEVERE REFLUX AND EXTENSIVE PROLAPSE FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, A MITRACLIP WAS IMPLANTED AND THE REFLUX WAS CONTROLLED. THE FINAL MR WAS GRADE 1-2. THERE WERE NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAYS REPORTED. ON (B)(6) 2026, IT WAS DETERMINED THAT A SINGLE LEAFLET DEVICE ATTACHMENT HAS OCCURRED AND CLIP WAS NO LONGER ATTACHED TO POSTERIOR LEAFLET AND MR HAS INCREASED AND PATIENT RETURNED WITH CONGESTIVE HEART FAILURE (CHF). SURGICAL TREATMENT OF THORACOTOMY AND MITRAL VALVE REPLACEMENT WAS PERFORMED ON (B)(6) 2026.

    Patient problems: Heart Failure/Congestive Heart Failure; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 24214743 ↗
  • InjuryJan 15, 2026Difficult or Delayed Positioning; Incomplete Coaptation

    N/A.

    Patient problems: Angina; Dyspnea; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 24081831 ↗
  • InjuryJan 12, 2026Incomplete Coaptation; Unintended Movement

    IT WAS REPORTED ON (B)(6) 2025 THAT THE PATIENT PRESENTED WITH GRADE 4 MIXED MITRAL REGURGITATION (MR), RESTRICTED AND FLAIL ANTERIOR LEAFLET FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, PLUS KNOB WAS UNABLE TO RESPOND DUE TO THE MECHANICAL JAM. AN ATTEMPT WAS MADE TO DEPLOY THE CLIP. THE CLIP OPENED TO 60 DEGREE POST DEPLOYMENT AND HAD SINGLE LEAFLET DEVICE ATTACHMENT(SLDA) FROM THE POSTERIOR LEAFLET. CHORDAL RUPTURE WAS OBSERVED. THE MR REMAINED UNCHANGED POST PROCEDURE. THE PATIENT WAS REFERRED FOR THORACIC SURGERY. THERE WAS NO CLINICALLY SIGNIFICANT DELAY.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; Unspecified Tissue Injury

    MAUDE report 24041173 ↗
  • MalfunctionJan 5, 2026Incomplete Coaptation

    IT WAS REPORTED ON (B)(6) 2025 THAT THE PATIENT PRESENTED WITH GRADE 4+ FUNCTIONAL MITRAL REGURGITATION (MR), TWO VENTRICULAR ANEURYSMS AND CARDIAC TRANSPOSITION FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, IMAGING WAS DIFFICULT. ONE MITRACLIP WAS SUCCESSFULLY IMPLANTED. DURING DEPLOYMENT OF SECOND MITRACLIP, PARTIAL CLIP DETACHMENT FROM ANTERIOR LEAFLET WAS OBSERVED. THE PROCEDURE WAS TERMINATED WITH REMOVAL OF THE SECOND CLIP AND NO CHANGE IN MR POST PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23956556 ↗
  • InjuryDec 24, 2025Positioning Failure; Entrapment of Device; Incomplete Coaptation; Positioning Failure; Entrapment of Device; Incomplete Coaptation

    IT WAS REPORTED THAT A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT FUNCTIONAL MITRAL REGURGITATION (MR) WITH A GRADE OF 4, LARGE GAP, AND ENLARGED ATRIUM. A MITRACLIP XTR WAS INSERTED AND DEPLOYED ON THE MITRAL VALVE. A SECOND CLIP, A MITRACLIP XTR (50717R1017) WAS INSERTED AND ADVANCED UNDER THE VALVE. HOWEVER, DIFFICULTIES CAPTURING THE ANTERIOR AND POSTERIOR LEAFLETS OCCURRED, THE CLIP BECAME CAUGHT IN CHORDAE. TROUBLESHOOTING WAS PERFORMED, AND THE CLIP WAS UNABLE TO BE FREED FROM CHORDAE. THEREFORE, THE CLIP WAS DEPLOYED WHERE IT WAS STUCK, ATTACHED TO THE ANTERIOR LEAFLET AND DETACHED FROM THE POSTERIOR LEAFLET (SINGLE LEAFLET DEVICE ATTACHMENT/SLDA). NO ADDITIONAL CLIPS WERE IMPLANTED, AND THE MR REMAINED AT A GRADE OF 4. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 23889694 ↗
  • InjuryDec 22, 2025Incomplete Coaptation; Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2025, A PATIENT PRESENTED WITH GRADE 4+ MIXED MITRAL REGURGITATION (MR), PROLAPSED ANTERIOR LEAFLET, THIN AND FRIABLE LEAFLETS FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, TWO MITRACLIPS WERE IMPLANTED SUCCESSFULLY. THE FINAL MR WAS GRADE 1+. THERE WERE NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAYS REPORTED. ON (B)(6) 2025, CARDIAC AUSCULTATION REVEALED A MURMUR AT THE MITRAL VALVE AREA. ECHOCARDIOGRAPHY WAS DONE AND IT WAS DETERMINED THAT SINGLE LEAFLET DEVICE ATTACHMENT (SLDA) HAS OCCURRED WITH ONE OF THE MITRACLIP DEVICES THAT HAD BEEN IMPLANTED AND MITRACLIP HAD DETACHED FROM THE POSTERIOR LEAFLET. MR WAS GRADE 4+ AFTER SLDA. IT WAS NOTED THAT THE PATIENT REQUIRED SURGICAL PROCEDURE FOR VALVE REPLACEMENT AND REMOVAL OF THE MITRACLIP. IT WAS NOTED THAT ON 01 DECEMBER 2025 DURING MORNING ROUNDS, THE PATIENT WAS ALERT AND IN STABLE CONDITION. PER THE PHYSICIAN, THE CAUSE OF THE SLDA WAS DETERMINED TO HAVE BEEN RELATED TO PRE-EXISTING PATIENT ANATOMY OF THIN AND FRIABLE LEAFLETS.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 23871226 ↗
  • InjuryDec 10, 2025Incomplete Coaptation; Device Damaged by Another Device; Incomplete Coaptation; Device Damaged by Another Device

    IT WAS REPORTED THAT ON (B)(6) 2025, A PATIENT PRESENTED WITH GRADE 4 DEGENERATIVE MITRAL REGURGITATION FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE A MITRACLIP XTR WAS SUCCESSFULLY IMPLANTED ON THE ANTERIOR-3/POSTERIOR-3 (A3/P3) LEAFLET SEGMENT. A SECOND MITRACLIP NTR WAS ADVANCED TO FURTHER REDUCE MR. THE SECOND CLIP HAD CHALLENGES CAPTURING THE LEAFLETS AND CAUSED THE FIRST CLIP TO HAVE A SINGLE LEAFLET DETACHMENT (SLDA). THE PROCEDURE WAS DISCONTINUED AND THE PATIENT UNDERWENT A SURGICAL MITRAL VALVE REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23767053 ↗
  • InjuryNov 17, 2025Incomplete Coaptation; Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2025, A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT DEGENERATIVE MITRAL REGURGITATION (MR) WITH A GRADE OF 4 AND PROLAPSED POSTERIOR LEAFLET. A STEERABLE GUIDE CATHETER (SGC) AND A MITRACLIP XTR WAS INSERTED WITHOUT ISSUES. IT WAS NOTED THAT DURING THE PROCEDURE, DURING MANIPULATION OF THE SGC, THERE WAS INSUFFICIENT SENSITIVITY, PREVENTING PRECISE CONTROL. GRASPING WAS PERFORMED WITH THE MITRACLIP XTR. HOWEVER, THE CLIP WAS UNABLE TO FIRMLY GRASP THE LEAFLETS. THE CLIP WAS ULTIMATELY IMPLANTED. TO FURTHER DECREASE MR, A SECOND CLIP, A MITRACLIP XTR WAS INSERTED AND DEPLOYED ON THE MITRAL VALVE, REDUCING MR TO A GRADE OF 1-2. ON (B)(6) 2025, AN ECHOCARDIOGRAM WAS PERFORMED AND REVEALED THAT THE CLIP HAD DETACHED FROM ONE LEAFLET AND REMAINED ATTACHED TO THE OTHER LEAFLET (SINGLE LEAFLET DEVICE ATTACHMENT/SLDA), CAUSING MR TO INCREASE.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 23569469 ↗
  • InjuryNov 13, 2025Incomplete Coaptation; Incomplete Coaptation

    IT WAS REPORTED THAT ON (B)(6) 2025, A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT FUNCTIONAL MITRAL REGURGITATION (MR) WITH A GRADE OF 5, EJECTION FRACTION OF 21%, AND CARDIAC TRANSPOSITION. THREE MITRACLIPS WERE INSERTED AND DEPLOYED ON THE MITRAL VALVE, REDUCING MR TO A GRADE OF 1.5+. ON (B)(6) 2025, A FOLLOW UP ULTRASOUND WAS PERFORMED AND REVEALED A POSTERIOR MITRAL LEAFLET PERFORATION CAUSED BY THE FIRST CLIP. THE MITRACLIP XTR (50605R1031). THE XTR HAD DETACHED FROM THE ANTERIOR LEAFLET AND REMAINED ATTACHED TO THE POSTERIOR LEAFLET (SINGLE LEAFLET DEVICE ATTACHMENT/SLDA), CAUSING MR TO INCREASE TO A GRADE OF 4+.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; Unspecified Tissue Injury

    MAUDE report 23547930 ↗
  • InjuryNov 4, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT ON (B)(6) 2025, A PATIENT PRESENTED WITH GRADE 3-4 FUNCTIONAL MITRAL REGURGITATION (MR) FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, A MITRACLIP XTR WAS ATTEMPTED TO BE PLACED BUT WAS UNABLE TO ADEQUATELY GRASP THE LEAFLETS. TROUBLESHOOTING WAS PREFORMED AND MULTIPLE GRASPING ATTEMPTS WERE MADE WHEN IT WAS IDENTIFIED THAT THE LEAFLET TISSUE HAD BEEN DAMAGED. THE CLIP WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS DISCONTINUED. THE FINAL MR REMAINED AT GRADE 3-4. THERE WERE NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAYS REPORTED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 23468136 ↗
  • InjuryNov 4, 2025Incomplete Coaptation; Incomplete Coaptation

    N/A.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 23467723 ↗
  • InjuryAug 18, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT DEGENERATIVE MITRAL REGURGITATION (MR) WITH A GRADE OF 4+ AND PROLAPSED POSTERIOR LEAFLET. A MITRACLIP XTR WAS INSERTED, ADVANCED INTO THE LEFT VENTRICLE (LV), AND GRASPING WAS PERFORMED. HOWEVER, DIFFICULTIES VISUALIZING THE CLIP OCCURRED, AND THE CLIP BECAME CAUGHT IN CHORDAE. TROUBLESHOOTING WAS PERFORMED, AND THE CLIP WAS FREED. HOWEVER, CHORDAE RUPTURE AND LEAFLET DAMAGE WERE OBSERVED. THEREFORE, THE CLIP WAS REMOVED FROM THE PATIENT. HOWEVER, AFTER REMOVING THE STEERABLE GUIDE CATHETER (SGC), IT WAS OBSERVED THAT THE INTERATRIAL SEPTUM WAS DAMAGED. THE PROCEDURE WAS DISCONTINUED. NO CLIPS WERE IMPLANTED, AND THE MR REMAINED AT A GRADE OF 4+. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 22829692 ↗
  • DeathAug 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2025, A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT FUNCTIONAL MITRAL REGURGITATION (MR) GRADE 4+. PATIENT HAD DILATED CARDIOMYOPATHY, LEFT VENTRICULAR END-SYSTOLIC DIAMETER 70MM, AND POSTERIOR LEAFLET RESTRICTION. A XTR CLIP WAS IMPLANTED REDUCING MR TO GRADE 1+. FIVE DAYS AFTER THE PROCEDURE (B)(6) 2025, THE PATIENT DIED FROM CEREBRAL HEMORRHAGE. THERE WERE NO REPORTED DEVICE ISSUE WITH THE XTR OR THE STEERABLE GUIDE CATHETER DURING THE INITIAL PROCEDURE. POST PROCEDURE THE CLIP WAS REPORTED TO REMAIN STABLE ON BOTH LEAFLETS AND NO TISSUE DAMAGE WAS OBSERVED. THE CAUSE OF THE CEREBRAL HEMORRHAGE WAS REPORTED TO BE UNKNOWN.

    Patient problems: Intracranial Hemorrhage

    MAUDE report 22815722 ↗
  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED ON (B)(6) 2025, A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT MIXED MITRAL REGURGITATION (MR) WITH A GRADE OF 4+. TWO CLIPS WERE IMPLANTED. HOWEVER, MR REMAINED UNCHANGED POST-PROCEDURE. ON (B)(6) 2025, AT A FOLLOW-UP APPOINTMENT, MR SHOWED NO SIGNS OF IMPROVEMENT AND TISSUE DAMAGE OBSERVED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; Unspecified Tissue Injury

    MAUDE report 22741838 ↗
  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED ON (B)(6) 2025, A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT MIXED MITRAL REGURGITATION (MR) WITH A GRADE OF 4+. TWO CLIPS WERE IMPLANTED. HOWEVER, MR REMAINED UNCHANGED POST-PROCEDURE. ON (B)(6) 2025, AT A FOLLOW-UP APPOINTMENT, MR SHOWED NO SIGNS OF IMPROVEMENT AND TISSUE DAMAGE OBSERVED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; Unspecified Tissue Injury

    MAUDE report 22741634 ↗
  • InjuryJul 25, 2025Difficult or Delayed Positioning; Incomplete Coaptation; Difficult or Delayed Positioning; Incomplete Coaptation

    IT WAS REPORTED ON (B)(6) 2025 THAT THE PATIENT WAS PRESENTED WITH GRADE 4+ DEGENERATIVE MITRAL REGURGITATION (MR), SMALL CARDIAC CHAMBERS AND PROLAPSED POSTERIOR LEAFLET FOR A MITRACLIP PROCEDURE. DURING THE PROCEDURE, A SINGLE LEAFLET DEVICE ATTACHMENT AND ATRIAL SEPTAL INJURY WAS OBSERVED REQUIRING SURGICAL INTERVENTION. THE MR REMAINED UNCHANGED POST PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; Unspecified Tissue Injury

    MAUDE report 22607799 ↗

FDA clearance and classification

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: To repair the mitral valve by coapting the valve leaflets to prevent mitral regurgitation. These devices are different from the classified device (annuloplasty rings) in that a cardiotomy is not required to place them, some of them are placed percutaneously, and they definitely are not rings that are sewn on to the valve annulus.

Common questions about the MITRACLIP CDS0601-XTR

Has the MITRACLIP CDS0601-XTR been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0115-2023 (Class II), began on Sep 8, 2022: An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions.… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the MITRACLIP CDS0601-XTR?

The FDA's MAUDE database lists 875 adverse event reports for this device, received between Oct 10, 2018 and Jul 31, 2026. In those reports, the most common specific device problems are “Incomplete Coaptation” (344), “Unintended Movement” (99) and “Difficult to Open or Close” (92); the most common patient problems are “Mitral Regurgitation” (187), “Tissue Damage” (132) and “Mitral Valve Insufficiency/ Regurgitation” (87). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MITRACLIP CDS0601-XTR?

30 reports are classified as deaths, 614 as injuries and 231 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MITRACLIP CDS0601-XTR MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the MITRACLIP CDS0601-XTR, and is it still sold?

The FDA's device database lists Abbott Vascular Inc. as the company responsible for it (brand MITRACLIP, model CDS0601-XTR). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Mitral valve repair devices devices

All Mitral valve repair devices devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page