MitraClipRecalls & reported problems across all models
3 companies · 19 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one MitraClip model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9,056
- Deaths reported
- 224
- Injuries reported
- 4,992
- Recalls
- 2
We found 19 models of MitraClip in the FDA's device registry (GUDID). Together they are named in 9,056 adverse event reports in the FDA's MAUDE database, received between Aug 4, 2016 and Sep 30, 2026 — classified as 224 deaths, 4,992 injuries, 3,840 malfunctions.
The most frequently reported specific device problem is “Incomplete Coaptation” (2,334 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
MitraClip recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 8, 2022Open, Classified
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
Z-0116-2023 · Root cause: Process control · Recall details · FDA record ↗
- Sep 8, 2022Open, Classified
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
Z-0115-2023 · Root cause: Process control · names 5 of these models · Recall details · FDA record ↗
Most reported MitraClip problems
Device problems
- Incomplete Coaptation2,334
- Unintended Movement1,546
- Adverse Event Without Identified Device or Use Problem1,297
- Difficult to Open or Close1,207
- Leak/Splash737
- Difficult to Remove670
- Difficult or Delayed Positioning579
- Positioning Failure511
Patient problems
- No Clinical Signs, Symptoms or Conditions3,873
- Mitral Valve Insufficiency/ Regurgitation1,299
- Unspecified Tissue Injury1,099
- No Consequences Or Impact To Patient1,087
- Mitral Regurgitation483
- Tissue Damage417
- Dyspnea321
- Foreign Body In Patient273
MitraClip models
19 models, most reported first. Each model has its own page with its reports, recalls and device details.
- MITRACLIP CDS0701-XTWAbbott Vascular Inc. · Mitral valve repair devices1 recall33 deaths1,524 reports
- MitraClip™ CDS0706-XTWSt. Jude Medical, Inc. · Mitral valve repair devices31 deaths1,341 reports
- MITRACLIP CDS0701-NTWAbbott Vascular Inc. · Mitral valve repair devices1 recall22 deaths959 reports
- MITRACLIP SGC0701Abbott Vascular Inc. · CATHETER, STEERABLE17 deaths963 reports
- MITRACLIP CDS0601-XTRAbbott Vascular Inc. · Mitral valve repair devices1 recall30 deaths873 reports
- MITRACLIP CDS0601-NTRAbbott Vascular Inc. · Mitral valve repair devices1 recall21 deaths641 reports
- MITRACLIP CDS0701-NTAbbott Vascular Inc. · Mitral valve repair devices1 recall12 deaths430 reports
- MitraClip™ CDS0706-NTWSt. Jude Medical, Inc. · Mitral valve repair devices13 deaths448 reports
- MitraClip™ CDS0801-XTWSt. Jude Medical, Inc. · Mitral valve repair devices5 deaths335 reports
- MITRACLIP CDS0701-XTAbbott Vascular Inc. · Mitral valve repair devices1 recall6 deaths271 reports
- MITRACLIP CDS0501Abbott Vascular Inc. · Mitral valve repair devices8 deaths312 reports
- MitraClip™ CDS0706-XTSt. Jude Medical, Inc. · Mitral valve repair devices7 deaths282 reports
- MitraClip™ SGC0801St. Jude Medical, Inc. · Catheter, percutaneous2 deaths177 reports
- MitraClip™ CDS0706-NTSt. Jude Medical, Inc. · Mitral valve repair devices4 deaths170 reports
- MitraClip™ CDS0801-XTSt. Jude Medical, Inc. · Mitral valve repair devices3 deaths115 reports
- MITRACLIP CDS0201Abbott Vascular · Mitral valve repair devices7 deaths82 reports
- MitraClip™ CDS0801-NTWSt. Jude Medical, Inc. · Mitral valve repair devices1 death69 reports
- MitraClip™ CDS0801-NTSt. Jude Medical, Inc. · Mitral valve repair devices1 death49 reports
- MITRACLIP SGC0101Abbott Vascular · CATHETER, STEERABLE1 death15 reports
Recent MitraClip adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Incomplete Coaptation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952127 ↗InjurySep 30, 2026Unintended Movement
It was reported that a MitraClip procedure was performed to treat degenerative mitral regurgitation (MR) with a grade of 4+ on a patient with thick leaflets and a pre-existing flail. A MitraClip XTW was inserted and placed on the mitral valve. However, during deployment, after pulling the lock line lever, the clip relaxed slightly, but within normal limits. The clip was deployed on the mitral valve. However, after deployment, the clip opened from approximately 20 to 60 degrees. It was noted that the clip remained stable on the leaflets. To stabilize the clip, a second clip was then inserted and deployed medial and adjacent to the first clip, reducing MR to a grade of 1. There was no clinically significant delay in the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952061 ↗InjurySep 30, 2026Incomplete Coaptation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951811 ↗MalfunctionSep 29, 2026Leak/Splash; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940222 ↗MalfunctionSep 29, 2026Difficult to Open or Close; Unintended Movement
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940086 ↗InjurySep 29, 2026Positioning Failure; Difficult to Remove
It was reported that a MitraClip procedure was performed to treat functional mitral regurgitation (MR) with a grade of 4 on a patient with a restricted posterior leaflet. A MitraClip XTW was inserted and grasping was performed. However, difficulties visualizing the clip occurred, he leaflets were unable to be grasped or captured, and the clip became caught in chordae. Troubleshooting was performed, and the clip was able to be freed from chordae, and the clip was removed from the patient. However, at the end of the procedure, a small flail was observed likely due to subvalvular interaction. The procedure was discontinued with no clips implanted. There was no clinically significant delay in the procedure.
Patient problems: Unspecified Tissue Injury
MAUDE report 50938411 ↗MalfunctionSep 29, 2026Break; Difficult or Delayed Positioning; Difficult to Open or Close
It was reported that on 04 September 2026 a patient presented with grade 4 functional mitral regurgitation (MR) for a Triclip procedure. During the procedure, posterior gripper would not lower or raise. Troubleshooting was performed but was unsuccessful. It was decided to remove the clip from the patient and was observed that gripper line was broken. Another mitraclip XT was implanted successfully. Imaging was difficult. The final MR was grade 2. All steps were performed as per IFU. There were no adverse patient effects or clinically significant delays reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936002 ↗InjurySep 29, 2026Incomplete Coaptation
It was reported that a MitraClip procedure was performed to treat functional mitral regurgitation (MR) with a grade of 4. A MitraClip XTW was inserted and deployed on the mitral valve. A second clip was then inserted and advanced to the left ventricle and positioned medial to the first clip. However, it was observed that the clip was too far away. Therefore, the clip was inverted and brought back into the atrium. Once the clip returned to the atrium, it was observed that the first clip detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/SLDA). To stabilize the clip, the second clip, The MitraClip XT, was repositioned and deployed on the lateral side of the first clip. To further stabilize the SLDA clip, a third clip was then inserted and deployed on the medial side of the first clip. The procedure was completed with three clips implanted, reducing MR to a grade of 1. It was noted that the patient tolerated the procedure well. There were no adverse patient effects and no clinically significant delay in the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934394 ↗MalfunctionSep 29, 2026Crack; Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934138 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that on (b)(6) 2026, a MitraClip procedure was performed to treat functional mitral regurgitation (MR) with a grade of 3. A MitraClip XTW (60312A2001) was inserted and deployed on the mitral valve. A second clip, a MitraClip XT (60212A1046) was then inserted and deployed on the mitral valve, reducing MR to trace.On (b)(6) 2026, the patient returned to the hospital with heart failure, which was treated per medical management.On (b)(6) 2026, the patient presented to the hospital. An echocardiogram was performed and revealed a hole in the anterior leaflet and increased MR with a grade of 4+. Both clips were noted to be attached to both leaflets.On (b)(6) 2026 The patient underwent surgery. The patient was reported to be stable and recovering.
Patient problems: Heart Failure/Congestive Heart Failure; Mitral Valve Insufficiency/ Regurgitation; Unspecified Tissue Injury
MAUDE report 50920386 ↗InjurySep 28, 2026Premature Activation; Difficult to Open or Close
It was reported that a MitraClip procedure was performed to treat functional mitral regurgitation (MR) with a grade of 4-5 on a patient with a small atrium. A steerable guide catheter (SGC) was prepared per the instructions for use (IFU) and successfully crossed the septum. A MitraClip XTW clip delivery system (CDS) was prepared and inserted. However, when the clip was inserted into the left atrium (LA), it was unable to open. Troubleshooting was performed, but the clip was still unable to open. Therefore, a decision was made to remove and replace the clip. However, as the CDS was retracted into the SGC, it appeared unstable; it appeared that it was not well attached to the mandrel, and every movement made, the clip would turn off the mandrel until the clip was perpendicular to the SGC. A snare was then inserted to retrieve the clip. However, the clip was unable to be retrieved, and remained inside the patient, attached to the septum. There was no clinically significant delay in the procedure. The patient was reported to be stable.
Patient problems: Embolism/Embolus; Foreign Body In Patient
MAUDE report 50919991 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Thrombosis/Thrombus
MAUDE report 50917831 ↗InjurySep 28, 2026Positioning Failure; Difficult to Remove; Retraction Problem; Failure to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917027 ↗InjurySep 25, 2026Incomplete Coaptation
Patient problems: Dyspnea; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 50899936 ↗MalfunctionSep 25, 2026Unintended Movement
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899619 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered, combining the brand names used for this product line.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page