Skip to content

MitraClipRecalls & reported problems across all models

3 companies · 19 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one MitraClip model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
9,056
Deaths reported
224
Injuries reported
4,992
Recalls
2

We found 19 models of MitraClip in the FDA's device registry (GUDID). Together they are named in 9,056 adverse event reports in the FDA's MAUDE database, received between Aug 4, 2016 and Sep 30, 2026 — classified as 224 deaths, 4,992 injuries, 3,840 malfunctions.

The most frequently reported specific device problem is “Incomplete Coaptation” (2,334 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 100 reportsDec 2023: 85 reportsJan 2024: 87 reportsFeb 2024: 79 reportsMar 2024: 92 reportsApr 2024: 86 reportsMay 2024: 83 reportsJun 2024: 118 reportsJul 2024: 61 reportsAug 2024: 107 reportsSep 2024: 79 reportsOct 2024: 94 reportsNov 2024: 80 reportsDec 2024: 75 reportsJan 2025: 84 reportsFeb 2025: 76 reportsMar 2025: 76 reportsApr 2025: 82 reportsMay 2025: 80 reportsJun 2025: 95 reportsJul 2025: 82 reportsAug 2025: 64 reportsSep 2025: 40 reportsOct 2025: 85 reportsNov 2025: 99 reportsDec 2025: 99 reportsJan 2026: 70 reportsFeb 2026: 92 reportsMar 2026: 114 reportsApr 2026: 85 reportsMay 2026: 65 reportsJun 2026: 50 reportsJul 2026: 34 reportsAug 2026: 47 reportsSep 2026: 44 reportsOct 2026: 0 reports202420252026118
Last 36 months. The most recent month may be incomplete.

MitraClip recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 8, 2022Open, Classified

    An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

    Z-0116-2023 · Root cause: Process control · Recall details · FDA record ↗

  2. Sep 8, 2022Open, Classified

    An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

    Z-0115-2023 · Root cause: Process control · names 5 of these models · Recall details · FDA record ↗

Most reported MitraClip problems

Device problems

  • Incomplete Coaptation2,334
  • Unintended Movement1,546
  • Adverse Event Without Identified Device or Use Problem1,297
  • Difficult to Open or Close1,207
  • Leak/Splash737
  • Difficult to Remove670
  • Difficult or Delayed Positioning579
  • Positioning Failure511

Patient problems

  • No Clinical Signs, Symptoms or Conditions3,873
  • Mitral Valve Insufficiency/ Regurgitation1,299
  • Unspecified Tissue Injury1,099
  • No Consequences Or Impact To Patient1,087
  • Mitral Regurgitation483
  • Tissue Damage417
  • Dyspnea321
  • Foreign Body In Patient273

MitraClip models

19 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent MitraClip adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page