HI-TORQUE INFILTRAC 1030003JRecalls & reported problems
Abbott Vascular Inc. · HI-TORQUE INFILTRAC PLUS Guide Wire w/Hydrophilic Coating .014” J Tip 190 cm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08717648233715), received between Oct 21, 2020 and Sep 23, 2022 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Remove3
- Material Separation2
- Break1
- Difficult to Advance1
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2022Break; Break
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE LEFT ANTERIOR DESCENDING (LAD) ARTERY WITH 80% STENOSIS, MODERATE CALCIFICATION AND HEAVY TORTUOSITY. THE HI-TORQUE (HT) INFILTRAC GUIDE WIRE CROSSED THE LESION, HOWEVER, THE GUIDE WIRE HAD EXPOSED A THIN WIRE LIKE PIECE IN THE PROXIMAL END AND UNRAVELED. THE GUIDE WIRE STARTED TO COIL AROUND THE PROXIMAL END NOT ALLOWING ANY DEVICE TO CROSS OVER IT. THE CORE OF THE GUIDE WIRE IS IN ONE PIECE, HOWEVER, IT IS UNKNOWN IF THE COILS DETACHED FROM THE CORE OF THE GUIDE WIRE. A GUIDE LINER WAS NOT USED. ANOTHER GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15473187 ↗MalfunctionDec 1, 2021Material Separation; Difficult to Advance; Material Separation; Difficult to Advance
IT WAS REPORTED THIS WAS A PROCEDURE TO TREAT A MILDLY TORTUOUS RIGHT CORONARY ARTERY. AN INFILTRAC 3/190CM GUIDE WIRE WAS INSERTED INTO THE ANATOMY. IT WAS NOTED THAT RESISTANCE WAS FELT AS IT WAS AN IN-STENT RESTENOSIS OCCLUSION. DURING ADVANCEMENT, THE GUIDE WIRE TIP SEPARATION INTO TWO PIECES. ALTHOUGH THE TIP SEPARATED, THE ENTIRE GUIDE WIRE WAS ABLE TO BE REMOVED WITHOUT PERFORMING INTERVENTION. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12908859 ↗InjuryNov 2, 2021Difficult to Remove; Material Separation; Failure to Advance; Difficult to Remove; Material Separation; Failure to Advance
IT WAS REPORTED THAT AN INFILTRAC PLUS GUIDE WIRE FACED RESISTANCE WITH THE ANATOMY DURING ADVANCEMENT AND REMOVAL. THE GUIDE WIRE TIP SEPARATED, AND THE SEPARATED PORTION WAS EMBEDDED WITH AN UNSPECIFIED STENT. AN UNSPECIFIED GUIDE WIRE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THE PATIENT IS FINE. THERE WERE NO ADVERSE PATIENT SEQUELAE AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 12741107 ↗MalfunctionOct 26, 2020Difficult to Remove; Difficult to Remove
THIS IS BEING FILED ON THE FIRST INFILTRAC GUIDEWIRE USED IN THIS PATIENT. IT WAS REPORTED THAT THIS WAS A PERCUTANEOUS CORONARY INTERVENTION TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE MID LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY DURING A PHYSICIAN PREFERENCE TESTING. THE INFILTRAC GUIDE WIRE SUCCESSFULLY PENETRATED THE CTO LUMEN. DURING REMOVAL OF THE INFILTRAC GUIDE WIRE, IT WAS DIFFICULT TO REMOVE DUE TO THE PATIENT ANATOMY. AFTER REMOVAL OF THE WIRE FROM THE NON-ABBOTT BALLOON DILATATION CATHETER (BDC), CONTRAST WAS NOTED TO BE STAINING ON THE FLUOROSCOPY, SIGNALING A RUPTURED BDC. THE PHYSICIAN SUSPECTS THAT THE TIP OF THE INFILTRAC GUIDE WIRE CAUSED THE BALLOON RUPTURE. AFTER REMOVAL OF THE GUIDE WIRE, THERE WAS NO DAMAGE NOTED TO THE GUIDE WIRE. ANOTHER INFILTRAC WAS USED TO COMPLETE THE PROCEDURE. THE SECOND INFILTRAC WAS ALSO DIFFICULT TO RETRACT, BUT IT WAS REMOVED WITHOUT INCIDENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10736135 ↗MalfunctionOct 21, 2020Difficult to Remove; Difficult to Remove
THIS IS BEING FILED ON THE SECOND INFILTRAC GUIDEWIRE USED IN THIS PATIENT. IT WAS REPORTED THAT THIS WAS A PERCUTANEOUS CORONARY INTERVENTION TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE MID LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY DURING A PHYSICIAN PREFERENCE TESTING. THE INFILTRAC GUIDE WIRE SUCCESSFULLY PENETRATED THE CTO LUMEN. DURING REMOVAL OF THE INFILTRAC GUIDE WIRE, IT WAS DIFFICULT TO REMOVE DUE TO THE PATIENT ANATOMY. AFTER REMOVAL OF THE WIRE FROM THE NON-ABBOTT BALLOON DILATATION CATHETER (BDC), CONTRAST WAS NOTED TO BE STAINING ON THE FLUOROSCOPY, SIGNALING A RUPTURED BDC. THE PHYSICIAN SUSPECTS THAT THE TIP OF THE INFILTRAC GUIDE WIRE CAUSED THE BALLOON RUPTURE. AFTER REMOVAL OF THE GUIDE WIRE, THERE WAS NO DAMAGE NOTED TO THE GUIDE WIRE. ANOTHER INFILTRAC WAS USED TO COMPLETE THE PROCEDURE. THE SECOND INFILTRAC WAS ALSO DIFFICULT TO RETRACT, BUT IT WAS REMOVED WITHOUT INCIDENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10715515 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page