Skip to content

Supera™ S-70-030-120-P6Recalls & reported problems

Abbott Vascular Inc. · Supera™ Peripheral Stent System 7.0 mm x 30 mm x 120 cm 6 F · Class 3 · Implant

All models of SUPERA (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648344244), received between Apr 22, 2025 and Apr 22, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Difficult to Remove1
  • Improper or Incorrect Procedure or Method1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2025Difficult to Remove; Material Separation; Improper or Incorrect Procedure or Method; Activation Failure; Difficult to Remove; Material Separation; Improper or Incorrect Procedure or Method; Activation Failure

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MODERATELY CALCIFIED LESION IN THE RIGHT COMMON FEMORAL ARTERY WITH A 7MM VESSEL DIAMETER. AFTER THE LESION WAS PRE-DILATED WITH A 6.0X40MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) AN ATHERECTOMY WAS PERFORMED WITH A NON-ABBOTT DEVICE. A 7.0X30MM SUPERA SELF-EXPANDING STENT (SES) WAS ADVANCED THROUGH A 6F NON-ABBOTT SHEATH; AND DEPLOYED 4 TO 5CM FROM THE DISTAL END OF THE SHEATH, AND UNDER FLUOROSCOPY THE STENT APPEARED TO BE FULLY DEPLOYED. FOLLOWING THIS, THE THUMBSLIDE WAS FULLY RETRACTED AND, WITH THE DEPLOYMENT LEVER STILL IN THE OPEN POSITION, THE DELIVERY SYSTEM WAS ATTEMPTED TO BE REMOVED; HOWEVER, RESISTANCE WITH THE GUIDEWIRE WAS FELT, AND THE TIP OF THE SES WAS NOTED TO SEPARATE. AN ATTEMPT TO CAPTURE THE SEPARATED TIP WITH THE SHEATH WAS MADE, BUT THE TIP WAS NOTED TO DISLODGE FROM INSIDE THE SHEATH BUT WAS STILL OVER THE GUIDEWIRE IN USE. THEREFORE, A HEMOSTAT WAS USED TO RETRIEVE THE SEPARATED TIP, AND AT THIS POINT, THE STENT WAS NOTED TO BE REMOVED AS WELL AND LOCATED INSIDE THE INTRODUCER SHEATH. FINALLY, AS A REMEDIAL MEASURE ANOTHER SUPERA SES WAS DEPLOYED TO TREAT THE LESION AND THE PROCEDURE WAS COM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21882043 ↗