Skip to content

Supera™ S-70-100-120-P6Recalls & reported problems

Abbott Vascular Inc. · Supera™ Peripheral Stent System 7.0 mm x 100 mm x 120 cm 6 F · Class 3 · Implant

All models of SUPERA (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648350627), received between Apr 3, 2024 and Jul 1, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Deformation Due to Compressive Stress1
  • Difficult or Delayed Activation1
  • Improper or Incorrect Procedure or Method1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 1, 2024Material Separation; Improper or Incorrect Procedure or Method; Material Separation; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT THE SUPERFICIAL FEMORAL ARTERY. THE VESSEL DIAMETER WAS 7.0MM AND WAS PREDILATED WITH A 7.0MM BALLOON DILATATION CATHETER. A 6FR X 45CM SHEATH WAS USED DURING THE PROCEDURE. THE 7.0X100MM SUPERA STENT WAS DEPLOYED AT THE TARGET VESSEL, BUT THE SUPERA TIP GOT CAUGHT ON THE STENT AND SHEARED OFF. THE THUMB SLIDE WAS NOT FULLY RETRACTED TO THE START POSITION AND BOTH THE SYSTEM WITH DEPLOYMENT LEVERS LOCKED PRIOR TO REMOVAL. THE SEPARATED TIP WAS SNARED. THERE WAS NO DAMAGE TO THE IMPLANTED SUPERA STENT IN THE TARGET VESSEL. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY DURING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643597 ↗
  • MalfunctionApr 3, 2024Material Separation; Stretched; Difficult or Delayed Activation; Deformation Due to Compressive Stress; Material Separation; Stretched; Difficult or Delayed Activation; Deformation Due to Compressive Stress

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A CHRONIC TOTAL OCCLUSION OF THE POPLITEAL AND DISTAL SUPERFICIAL FEMORAL ARTERY. THE 7.0X100MM SUPERA SELF EXPANDING STENT SYSTEM (SES) WAS ADVANCED TO THE TARGET LESION. IT WAS THOUGHT THE STENT HAD FULLY DEPLOYED HOWEVER DURING REMOVAL IT WAS NOTED THE STENT WAS STILL PARTIALLY IN THE SHEATH AND HAD ELONGATED. DURING RETRACTION THE STENT WAS ABLE TO BE FULLY DEPLOYED; HOWEVER SINCE IT ELONGATED IT WAS DEPLOYED IN THE TARGET LESION AND PARTIALLY IN HEALTHY TISSUE. AFTER THE SES WAS REMOVED FROM THE PATIENT IT WAS NOTED THE NOSE CONE WAS MISSING AND THE SHEATH WAS KINKED. ANGIOGRAM WAS PERFORMED HOWEVER THE MISSING NOSE CONE WAS NOT FOUND IN THE PATIENT. AS A PRECAUTION, DOPPLER WAS PERFORMED TO ENSURE THE LEG HAD PROPER FLOW, AND THE FLOW SOUNDED GREAT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19031397 ↗

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates