Esprit™ (REF 1212350-38)Recalls & reported problems

Abbott Vascular Inc. · Esprit™ BTK Everolimus Eluting Resorbable Scaffold System 3.50mm x 38mm Over-The-Wire · Class 3 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648353703), received between Mar 6, 2026 and Jun 4, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Ambient Temperature Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Ambient Temperature Problem1
  • Difficult or Delayed Activation1
  • Difficult to Advance1
  • Improper or Incorrect Procedure or Method1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Separation; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN ANTERIOR TIBIAL ARTERY LESION. A 3.5X38 MM ESPIRIT BTK OTW BIORESORBABLE VASCULAR SCAFFOLD (BVS) WAS UNABLE TO BE ADVANCED ALONG THE GUIDE WIRE. AN ADDITIONAL ESPRIT BTK BVS WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. RETURNED DEVICE ANALYSIS INDICATED THAT THE OUTER MEMBER WAS SEPARATED AT THE MID LAP SEAL. THE INNER MEMBER WAS SEPARATED 21 CM PROXIMAL TO THE MID LAP SEAL. THE INNER MEMBER WAS ALSO STRETCHED OUT DISTALLY TO THE NOTED SEPARATION FOR A LENGTH OF 14.4 CM. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25394123 ↗
  • MalfunctionApr 2, 2026Improper or Incorrect Procedure or Method; Ambient Temperature Problem

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A CALCIFIED LESION IN THE PROXIMAL ANTERIOR TIBIAL ARTERY. IT WAS NOTED DURING PREPARATION THAT THE TEMPERATURE TAGS ON THE 3.50X38MM ESPRIT BTK AND 3.75X38MM ESPRIT BTK SYSTEMS WERE TRIGGERED. THE DEVICES WERE NOT CORRECTLY STORED IN A TEMPERATURE MONITORED ROOM. THE SCAFFOLDS WERE IMPLANTED IN THE PATIENT WITHOUT ISSUE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24766806 ↗
  • MalfunctionMar 6, 2026Difficult or Delayed Activation

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A CALCIFIED LESION IN THE RIGHT ANTERIOR TIBIAL (AT) ARTERY. THE LESION WAS PREPARED WITH A DIAMONDBACK ATHERECTOMY DEVICE AND BALLOON INFLATIONS. A NON-ABBOTT GUIDE WIRE WAS ADVANCED TO THE LESION WITH RESISTANCE NOTED DUE TO THE CALCIUM. THE 3.50X38MM OTW ESPRIT BTK SYSTEM WAS ADVANCED TO THE TARGET LESION AND THE BALLOON INFLATED TO 16 ATMOSPHERES (ATM) HOWEVER UNDER ANGIO IT WAS NOTED THERE WAS A PORTION OF THE MIDDLE OF THE BALLOON THAT DID NOT INFLATE. INFLATION WAS INCREASED TO 20 ATM HOWEVER THE MIDDLE STILL DID NOT INFLATE. THE SCAFFOLD WAS DEPLOYED THEREFORE THE ESPRIT BTK SYSTEM WAS RETRACTED. THE PHYSICIAN WANTED TO TEST THE BALLOON THEREFORE IT WAS INFLATED IN ANOTHER PART OF THE LESION AND THE BALLOON WAS ABLE TO FULLY INFLATE WITH NO WAIST. THE ESPRIT BTK SYSTEM WAS REMOVED AND THE LESION POST DILATED WITH A NON-COMPLIANT BALLOON. IN THE PHYSICIANS OPINION, THE WAIST WAS DUE TO CALCIUM IN THE LESION AND NOT A BALLOON ISSUE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531836 ↗

Other Scaffold, infrapopliteal, absorbable devices

All Scaffold, infrapopliteal, absorbable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.