Esprit™ 1203250-28Recalls & reported problems
Abbott Vascular Inc. · Esprit™ BTK Everolimus Eluting Resorbable Scaffold System 2.50mm x 28mm Rapid-Exchange · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648357237), received between Aug 1, 2024 and Jul 16, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance2
- Improper or Incorrect Procedure or Method2
- Ambient Temperature Problem1
- Break1
- Deformation Due to Compressive Stress1
- Failure to Advance1
- Material Too Rigid or Stiff1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2026Improper or Incorrect Procedure or Method; Difficult to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25846461 ↗MalfunctionMar 2, 2026Improper or Incorrect Procedure or Method; Ambient Temperature Problem
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE RIGHT ANTERIOR TIBIAL ARTERY. THE 2.50 X 28 MM ESPRIT BTK SCAFFOLD WAS SUCCESSFULLY IMPLANTED. AFTER DEPLOYMENT, IT WAS OBSERVED THAT THE THERMOALERT TAG WAS NOTED TO BE TRIPPED. IT IS UNKNOWN WHEN THE TAG ALERT SENSOR WAS TRIGGERED AS THE DEVICE WAS NOT STORED IN A TEMPERATURE MONITORED ROOM. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24491646 ↗MalfunctionAug 1, 2024Break; Material Too Rigid or Stiff; Failure to Advance; Deformation Due to Compressive Stress; Difficult to Advance; Break; Material Too Rigid or Stiff; Failure to Advance; Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE PERONEAL ARTERY WITH MODERATE CALCIFICATION AND MILD TORTUOSITY. A 6FR X 45 CM SHEATH AND A 0.014¿ NON-ABBOTT GUIDEWIRE WAS USED IN THE CASE. THE PHYSICIAN ACCESSED THE ANATOMY THROUGH THE RIGHT GROIN AND WENT UP AND OVER THE ILIAC ARCH TO THE LEFT SIDE. IVUS (INTRAVASCULAR ULTRASOUND) WAS USED TO VISUALIZE THE SUPERFICIAL FEMORAL ARTERY (SFA) AND A DRUG COATED BALLOON (DCB) WAS ADVANCED TO THE SFA. ANOTHER IVUS RUN WAS COMPLETED AFTER THE INFLATION OF THE DCB AND EVERYTHING LOOKED GOOD. THE TARGET LESION WAS IN THE PERONEAL ARTERY AND WAS PRE-DILATED WITH A BALLOON DILATATION CATHETER. THE ESPRIT BTK EVEROLIMUS ELUTING BIORESORBABLE SCAFFOLD SYSTEM WAS PREPPED AND INSERTED. THE PHYSICIAN COMMENTED THAT THE DEVICE FELT STIFF AND THERE WAS SOME RESISTANCE EXPERIENCED WHEN ADVANCING. THE DEVICE WAS ADVANCED UP AND OVER THE ILIAC ARCH, BUT ONCE ON THE OTHER SIDE, THERE WAS RESISTANCE AND THE DELIVERY SYSTEM WAS PUSHING THE SHEATH BACK TOWARDS THE AORTA. THE SDS WOULD NOT GO OVER THE ILIAC ARTERY. AFTER MULTIPLE ATTEMPTS, THE SDS WAS REMOVED FROM THE PATIENT WITHOUT ANY ISSUE. THE PROCEDURE WAS ENDED. WHEN THE SD
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19886947 ↗
Other Scaffold, Infrapopliteal, Absorbable devices
- Esprit™ 1203300-38Abbott Vascular Inc. · Scaffold, Infrapopliteal, Absorbable1 death47 reports
- Esprit™ 1203250-38Abbott Vascular Inc. · Scaffold, Infrapopliteal, Absorbable37 reports
- Esprit™ 1203350-38Abbott Vascular Inc. · Scaffold, Infrapopliteal, Absorbable1 death30 reports
- Esprit™ 1203375-38Abbott Vascular Inc. · Scaffold, Infrapopliteal, Absorbable16 reports
- Esprit™ 1203300-28Abbott Vascular Inc. · Scaffold, Infrapopliteal, Absorbable10 reports
- Esprit™Abbott Vascular Inc. · Scaffold, infrapopliteal, absorbable4 reports
- Esprit™ 1203350-18Abbott Vascular Inc. · Scaffold, Infrapopliteal, Absorbable3 reports
- Esprit™Abbott Vascular Inc. · Scaffold, infrapopliteal, absorbable3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page