CLARITY II COMPL_BODY_ICDRecalls & reported problems
Lutronic Corporation · COMPL_BODY_ICD · Class 2
All models of CLARITY II (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08809447652542), received between Sep 23, 2022 and Aug 8, 2025 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Use of Incorrect Control/Treatment Settings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Use of Incorrect Control/Treatment Settings2
- Continuous Firing1
Patient problems
- Full thickness (Third Degree) Burn3
- Blister1
- No Clinical Signs, Symptoms or Conditions1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SITE PERFORMED A VESSEL/ROSACEA TREATMENT ON A PATIENT'S NOSE USING CLARITY II. DURING TREATMENT, A LOUD "POP" SOUND WAS HEARD, AND THEN A WHITE BLISTER APPEARED ON PATIENT'S NOSE. SITE STATES, "PATIENT RECEIVED A FULL THICKNESS BURN TO NOSE." CYNOSURE LUTRONIC FIELD SERVICE ENGINEER (FSE) ARRIVED ON SITE TO PERFORM DEVICE EVALUATION. DURING THIS TIME CYNOSURE LUTRONIC FSE CONFIRMED DEVICE HAD "ERROR 63 FAULTY POWER MODULE." CYNOSURE LUTRONIC FSE WAS UNABLE TO COMPLETE REPAIR DURING THIS VISIT. ON THE SECOND FSE VISIT, THE DEVICE WAS BROUGHT BACK TO THE CYNOSURE LUTRONIC (B)(6) LOCATION FOR FURTHER REPAIR. THE DEVICE HAD THE PUMPING BOARD ASSEMBLY REPLACED AND THE ERROR CODE WAS SUCCESSFULLY CLEARED. THE DEVICE WAS SUBJECTED TO FINAL TESTING AT THE WESTFORD LOCATION WITH PASSING RESULTS. IT IS UNKNOWN IF THE ERROR CODE OBSERVED CONTRIBUTED TO THE REPORTED EVENT. CYNOSURE LUTRONIC CLINICAL REACHED OUT TO THE SITE TO COLLECT ADDITIONAL INFORMATION. CYNOSURE LUTRONIC CLINICAL DETERMINED THE EVENT TO BE INCONCLUSIVE. REVIEW OF THE TREATMENT PARAMETERS USED WERE FOUND TO BE WITHIN THE RECOMMENDED SETTINGS WITHIN THE QUICK START GUIDE (QSG). THE PATIENT WAS PRESCRIBE
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 22750107 ↗InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SITE PERFORMED A VESSEL/ROSACEA TREATMENT ON A PATIENT'S NOSE USING CLARITY II. DURING TREATMENT, A LOUD "POP" SOUND WAS HEARD, AND THEN A WHITE BLISTER APPEARED ON PATIENT'S NOSE. SITE STATES, "PATIENT RECEIVED A FULL THICKNESS BURN TO NOSE."
Patient problems: Blister
MAUDE report 22738588 ↗MalfunctionJun 12, 2025Continuous Firing; Continuous Firing
DURING A PREVENTATIVE MAINTENANCE VISIT ONSITE, CYNOSURE FIELD SERVICE ENGINEER OBSERVED CONTINUOUS FIRING WITH CLARITY II. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22195251 ↗InjuryMay 21, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings
PATIENT'S TATTOO WAS ACCIDENTALLY HIT DURING LASER HAIR REDUCTION TREATMENT USING CLARITY II. CYNOSURE'S MEDICAL DIRECTOR WAS CONSULTED ON THIS EVENT AND STATED, "THIS IS A THIRD DEGREE. I WOULD EXPECT A SCAR AND THE WOUND HEALS IN BY SECONDARY INTENTION. ONCE FULLY CLOSED, I WOULD RECOMMEND SILICON GEL, MASSAGE AND EITHER MICRO NEEDLING OR NON ABLATIVE LASER RESURFACING." PATIENT WAS GIVEN TOPICAL ANTIBIOTICS AS PREVENTATIVE MEDICATION. CYNOSURE CLINICAL DETERMINED THE EVENT WAS DUE TO USER ERROR. DURING LASER HAIR REMOVAL TREATMENTS THE TATTOO AREA MUST BE STRICTLY AVOIDED. THE CLARITY II QSG INSTRUCTS: "DO NOT TREAT OVER OR CLOSE TO TATTOOS OR PERMANENT MAKE UP." DUE TO SCARRING EXPECTED, THIS IS CONSIDERED A PERMANENT INJURY AND THEREFORE REPORTABLE.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 22060447 ↗InjuryMay 20, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings
A PATIENTS TATTOO WAS ACCIDENTLY HIT DURING LASER HAIR REDUCTION TREATMENT WITH CLARITY II DEVICE.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 22054763 ↗MalfunctionSep 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS BEING TREATED FOR LASER HAIR REMOVAL ON HER CHIN WITH 1064NM NDYAG LASER. WHILE LASER WAS FIRING, CRYOGEN FUNCTION STOPPED FIRING. THIS RESULTED IN MINOR 1ST DEGREE THERMAL BURN AND WAS TREATED WITH SILVADENE CREAM..
Patient problems: Superficial (First Degree) Burn
MAUDE report 15472808 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page