CLARITY II COMPL_BODY_ICDRecalls & reported problems

Lutronic Corporation · COMPL_BODY_ICD · Class 2

All models of CLARITY II (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08809447652542), received between Sep 23, 2022 and Aug 8, 2025 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Use of Incorrect Control/Treatment Settings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Use of Incorrect Control/Treatment Settings2
  • Continuous Firing1

Patient problems

  • Full thickness (Third Degree) Burn3
  • Blister1
  • No Clinical Signs, Symptoms or Conditions1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SITE PERFORMED A VESSEL/ROSACEA TREATMENT ON A PATIENT'S NOSE USING CLARITY II. DURING TREATMENT, A LOUD "POP" SOUND WAS HEARD, AND THEN A WHITE BLISTER APPEARED ON PATIENT'S NOSE. SITE STATES, "PATIENT RECEIVED A FULL THICKNESS BURN TO NOSE." CYNOSURE LUTRONIC FIELD SERVICE ENGINEER (FSE) ARRIVED ON SITE TO PERFORM DEVICE EVALUATION. DURING THIS TIME CYNOSURE LUTRONIC FSE CONFIRMED DEVICE HAD "ERROR 63 FAULTY POWER MODULE." CYNOSURE LUTRONIC FSE WAS UNABLE TO COMPLETE REPAIR DURING THIS VISIT. ON THE SECOND FSE VISIT, THE DEVICE WAS BROUGHT BACK TO THE CYNOSURE LUTRONIC (B)(6) LOCATION FOR FURTHER REPAIR. THE DEVICE HAD THE PUMPING BOARD ASSEMBLY REPLACED AND THE ERROR CODE WAS SUCCESSFULLY CLEARED. THE DEVICE WAS SUBJECTED TO FINAL TESTING AT THE WESTFORD LOCATION WITH PASSING RESULTS. IT IS UNKNOWN IF THE ERROR CODE OBSERVED CONTRIBUTED TO THE REPORTED EVENT. CYNOSURE LUTRONIC CLINICAL REACHED OUT TO THE SITE TO COLLECT ADDITIONAL INFORMATION. CYNOSURE LUTRONIC CLINICAL DETERMINED THE EVENT TO BE INCONCLUSIVE. REVIEW OF THE TREATMENT PARAMETERS USED WERE FOUND TO BE WITHIN THE RECOMMENDED SETTINGS WITHIN THE QUICK START GUIDE (QSG). THE PATIENT WAS PRESCRIBE

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 22750107 ↗
  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SITE PERFORMED A VESSEL/ROSACEA TREATMENT ON A PATIENT'S NOSE USING CLARITY II. DURING TREATMENT, A LOUD "POP" SOUND WAS HEARD, AND THEN A WHITE BLISTER APPEARED ON PATIENT'S NOSE. SITE STATES, "PATIENT RECEIVED A FULL THICKNESS BURN TO NOSE."

    Patient problems: Blister

    MAUDE report 22738588 ↗
  • MalfunctionJun 12, 2025Continuous Firing; Continuous Firing

    DURING A PREVENTATIVE MAINTENANCE VISIT ONSITE, CYNOSURE FIELD SERVICE ENGINEER OBSERVED CONTINUOUS FIRING WITH CLARITY II. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22195251 ↗
  • InjuryMay 21, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings

    PATIENT'S TATTOO WAS ACCIDENTALLY HIT DURING LASER HAIR REDUCTION TREATMENT USING CLARITY II. CYNOSURE'S MEDICAL DIRECTOR WAS CONSULTED ON THIS EVENT AND STATED, "THIS IS A THIRD DEGREE. I WOULD EXPECT A SCAR AND THE WOUND HEALS IN BY SECONDARY INTENTION. ONCE FULLY CLOSED, I WOULD RECOMMEND SILICON GEL, MASSAGE AND EITHER MICRO NEEDLING OR NON ABLATIVE LASER RESURFACING." PATIENT WAS GIVEN TOPICAL ANTIBIOTICS AS PREVENTATIVE MEDICATION. CYNOSURE CLINICAL DETERMINED THE EVENT WAS DUE TO USER ERROR. DURING LASER HAIR REMOVAL TREATMENTS THE TATTOO AREA MUST BE STRICTLY AVOIDED. THE CLARITY II QSG INSTRUCTS: "DO NOT TREAT OVER OR CLOSE TO TATTOOS OR PERMANENT MAKE UP." DUE TO SCARRING EXPECTED, THIS IS CONSIDERED A PERMANENT INJURY AND THEREFORE REPORTABLE.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 22060447 ↗
  • InjuryMay 20, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings

    A PATIENTS TATTOO WAS ACCIDENTLY HIT DURING LASER HAIR REDUCTION TREATMENT WITH CLARITY II DEVICE.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 22054763 ↗
  • MalfunctionSep 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS BEING TREATED FOR LASER HAIR REMOVAL ON HER CHIN WITH 1064NM NDYAG LASER. WHILE LASER WAS FIRING, CRYOGEN FUNCTION STOPPED FIRING. THIS RESULTED IN MINOR 1ST DEGREE THERMAL BURN AND WAS TREATED WITH SILVADENE CREAM..

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 15472808 ↗