Skip to content

CLARITY II COMPL_BODY_AIRRecalls & reported problems

Lutronic Corporation · COMPL_BODY_AIR Package DI Number : 08809447652634 · Class 2

All models of CLARITY II (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08809447652610), received between Feb 7, 2025 and Feb 7, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2
  • Unexpected/Unintended Radiation Output2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 7, 2025Electrical /Electronic Property Problem; Unexpected/Unintended Radiation Output; Electrical /Electronic Property Problem; Unexpected/Unintended Radiation Output

    SITE REPORTED CLARITY II'S FOOT PEDAL IS FIRING UNINTENTIONALLY AND "THE ONLY WAY THEY CAN STOP IT IS TO TURN THE MACHINE OFF." NO INJURIES WERE REPORTED. CYNOSURE FIELD SERVICE ENGINEER (FSE) ARRIVED AT SITE AND EVALUATED THE DEVICE. FSE WAS ABLE TO DUPLICATE THE CONTINUOUS MISFIRING. FSE STATES, "THE UNIT HAS OUT OF DATE FIRMWARE AND DOES NOT HAVE A FERRITE COIL OR GROUNDED HANDPIECE ASSEMBLY." TO RESOLVE THE REPORTED ISSUE, FSE REPLACED THE HANDPIECE WITH PROPER GROUND VERSION, ADDED A FERRITE COIL AND PERFORMED THE NECESSARY FIRMWARE UPDATES. THIS EVENT IS REPORTABLE SINCE THIS IS AN ARO (ACCIDENTAL RADIATION OCCURRENCE) DUE TO DEVICE FIRING UNINTENTIONALLY ON ITS OWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21334192 ↗
  • MalfunctionFeb 7, 2025Electrical /Electronic Property Problem; Unexpected/Unintended Radiation Output; Electrical /Electronic Property Problem; Unexpected/Unintended Radiation Output

    THE CUSTOMER REPORTED THAT WHEN USING CLARITY II DEVICE FOOT PEDAL, THE DEVICE BEGAN FIRING CONTINUOUSLY FROM THE HAND PIECE. THE DEVICE HAD TO BE SHUT DOWN TO STOP THE HAND PIECE FROM FIRING. IT WAS REPORTED THAT THERE WERE NO INJURIES THAT OCCURRED DURING THIS INCIDENT. THERE WAS NO INJURY REPORTED ASSOCIATED WITH THIS INCIDENT. THE CLINICAL TEAM MADE MULTIPLE CONTACT ATTEMPTS TO OBTAIN ADDITIONAL INFO ABOUT THE REPORTED EVENT, INCLUDING PATIENT DETAILS AS WELL AS TREATMENT AND PARAMETER DETAILS, HOWEVER, THESE INFORMATION REQUESTS WENT UNANSWERED. A TRAINED FIELD SERVICE ENGINEER WENT TO THE CUSTOMER SITE FOR EVALUATION. DURING THIS TIME THE REPORTED ISSUE OF CONTINUOUS FIRING WAS DUPLICATED. TO CORRECT THIS ISSUE, THE DEVICE FIRMWARE WAS UPDATED AND THE HAND PIECE HAD THE FERRITE GROUNDING COIL INSTALLED. THE FIELD SERVICE ENGINEER RECOMMENDED THE CUSTOMER REPLACE THIS HAND PIECE WITH AN UPDATED AND GROUND HAND PIECE VERSION. WHEN THE REPAIRS WERE COMPLETE, THE DEVICE WAS TESTED AND PASSED ALL ENERGY AND FUNCTIONAL TESTS AS EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21333929 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page