BOND-III processing moduleRecalls & reported problems

Leica Biosystems Melbourne Pty Ltd · The BOND system automates clinical protocols for immunostaining of pathology specimens mounted on microscope slides. Microscope slides subsequently undergo interpretation by a qualified healthcare professional to aid diagnosis. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 09349458000004), received between Oct 5, 2023 and May 22, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method2
  • Off-Label Use2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON 07 MAY 2025, LEICA BIOSYSTEMS RECEIVED THE FOLLOWING INFORMATION: ¿BOND-III, THERE ARE SCRATCHES ON THE DETECTION TISSUE AFTER THE INSTRUMENT GETS OFF THE PLANE, AND PART OF THE TISSUE IS LOST.¿ ON 09 MAY 2025, THE ASSIGNED LEICA FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER SITE AND DOCUMENTED THE FOLLOWING INFORMATION: ¿ACCORDING TO THE CUSTOMER'S DESCRIPTION, THE OCCURRENCE OF TISSUE SCRATCHES IS OCCASIONAL. ON-SITE INSPECTION REVEALED THAT THE PROBE, COVERTILE, SLIDE RACK, AND SSA WERE ALL NORMAL, AND NO ISSUES WERE FOUND WITH THE INSTRUMENT'S HARDWARE. ONLY IN THE DEWAXING SOLUTION REAGENT BOTTLE WERE SOME INCOMPATIBLE LIQUIDS FOUND, AND IN THE ALCOHOL REAGENT BOTTLE, SOME BLACK PARTICLE PRECIPITATES WERE DISCOVERED. CLEAN THE REAGENT BOTTLES AND REPLACE THE CONTAMINATED REAGENTS. TEST OK.¿ ON 10 MAY 2025, THE ASSIGNED LEICA FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER SITE AND DOCUMENTED THE FOLLOWING INFORMATION: ¿CLEANED DEWAX AND ALCOHOL BOTTLES AND REPLACED THE REAGENTS. THERE WERE NO SAME PROBLEM HAPPENING IN TWO DAYS, OBSERVING FOR A WHILE.¿ ON 14 MAY 2025, LEICA BIOSYSTEMS MELBOURNE RECEIVED INFORMATION FROM THE LOCAL LEICA QA SPECIALIST, QARA THAT ¿¿THIS C

    Patient problems: Insufficient Information

    MAUDE report 22075475 ↗
  • MalfunctionJun 13, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON (B)(6) 2024, THE COMPLAINANT CONTACTED LEICA BIOSYSTEMS WITH THE FOLLOWING DETAILS: ¿ERROR 126 AND COVERTILE ISSUE ERROR 126 ASPIRATION PROBE RECOVERED COVERTILE ARE MISALIGNED DURING RUN. FAS [NAME] IS AWARE OF THE CLIPPED CORNERS, AND IS REQUESTING SERVICE FOR ERROR 126¿. ON (B)(6) 2024, A LOCAL LEICA TECHNICAL SUPPORT ENGINEER DOCUMENTED: ¿...CUSTOMER IS USING CLIPPED CORNER SLIDES, ( A LIKELY CAUSE OF THE COVERTILE ISSUE )...¿ ON (B)(6) 2024, THE ASSIGNED LEICA FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER SITE AND RECORDED THE FOLLOWING INFORMATION: ¿THE CUSTOMER COMPLAINED OF RECIEVING [SIC] ERROR 126 SOFTWARE ERROR, XYZ_ERROR: ACTION_RUN_ERROR_XYZ_MOVE_FAIL. ERROR CODE 116 OCCURRED ONCE DURING THE RUN EVENTS. CHECKED ALL CALIBRATION, CONDUCTED EXTREMITY TEST , CHECKED Z AXIS HOME FLAG AND IT WAS NOT BENT. CHECKED Z AXIS INSULATOR BLOCK BRACKET AND LOOM WERE NOT FOULING ON THE ROBOT ARM. X AXIS BELT AND ALL CONNECTIONS WERE SECURE.¿ THE FSE ¿VERIFIED CALIBRATION ADVISED CUSTOMER TO MARK SLIDE TRAYS TO BETTER IDENTIFY POSSIBLE PROBLEMATIC TRAYS. CUSTOMER AT THIS TIME HAS 4 GEN 1 SLIDE TRAYS AND 1 GEN 3 TRAYS. CONDUCTED CALIBRATION VERIFICATION USING GEN 3 SLIDE TRAY WIT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19534399 ↗
  • InjuryOct 27, 2023Off-Label Use; Off-Label Use

    ON (B)(6) 2023, LEICA BIOSYSTEMS MELBOURNE RECEIVED THE FOLLOWING INFORMATION FROM THE LOCAL LEICA SENIOR APPLICATION CONSULTANT: "1 PATIENT: PROBABLE HARM, DID NOT HAVE NEOADJ [SIC] CHEMO, STOPPED HORMONAL THERAPY, 5 PATIENTS: DOUBT ABOUT CLINICAL CONSEQUENCE (PARTIAL TTT RESPONSES TO NEOADJ [SIC] CHEMO OR NO NEOADJ [SIC] CHEMO), 2 PATIENTS: HORMONAL THERAPY DISCONTINUATION". AN IDENTIFIER, AGE OR DATE OF BIRTH AND GENDER FOR THE EIGHT (8) PATIENT CASES FOR WHICH THERE IS "POSSIBLE HARM." HAS NOT BEEN PROVIDED TO LEICA BIOSYSTEMS.

    Patient problems: Insufficient Information

    MAUDE report 18018060 ↗
  • InjuryOct 27, 2023Off-Label Use; Off-Label Use

    ON 05 DECEMBER 2023, LEICA BIOSYSTEMS MELBOURNE RECEIVED THE FOLLOWING INFORMATION FROM THE LOCAL LEICA SENIOR APPLICATION CONSULTANT: "1 PATIENT: PROBABLE HARM, DID NOT HAVE NEOADJ [SIC] CHEMO, STOPPED HORMONAL THERAPY, 5 PATIENTS: DOUBT ABOUT CLINICAL CONSEQUENCE (PARTIAL TTT RESPONSES TO NEOADJ [SIC] CHEMO OR NO NEOADJ [SIC] CHEMO), 2 PATIENTS: HORMONAL THERAPY DISCONTINUATION". AN IDENTIFIER, AGE OR DATE OF BIRTH AND GENDER FOR THE EIGHT (8) PATIENT CASES FOR WHICH THERE IS "...POSSIBLE HARM." HAS NOT BEEN PROVIDED TO LEICA BIOSYSTEMS.

    Patient problems: Insufficient Information

    MAUDE report 18018058 ↗
  • MalfunctionOct 5, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON (B)(6) 2023, THE COMPLAINANT CONTACTED LEICA BIOSYSTEMS AND THE FOLLOWING DETAILS WERE DOCUMENTED: "FAS [NAME] CALLED IN THE ISSUE. MAIN ROBOT CRITICAL ERR, RUN ABANDONED ON SSA 2 ABLE TO REPEAT SLIDES. HESITANT TO USE THE OTHER SSA AND NO REPORTED ISSUE ON THE OTHER TRAYS". ON (B)(6) 2023, THE ASSIGNED LOCAL LEICA FIELD APPLICATIONS SPECIALIST IMMUNOHISTOCHEMISTRY (FAS) VISITED THE CUSTOMER SITE AND RECORDED THE FOLLOWING INFORMATION: "CUSTOMER REPORTED STAINING ISSUES ON RECENT BOND III PROCESSING MODULE AND REQUESTED ASSISTANCE". THE FAS NOTED "THE PROCESSING MODULE IS WELL-MAINTAINED, PH APPROPRIATE BULK FLUIDS, AND SYRINGE PRIMES APPROPRIATELY...PULLED RUN EVENTS FOR THE AFFECTED SLIDES AND FOUND THAT A MAIN ROBOT CRITICAL ERROR HAD OCCURRED, AND THE RUN WAS ABANDONED. REQUESTED AN FSE ON SITE TO FURTHER EVALUATE THE ISSUE." THE LEICA FAS ALSO CONFIRMED THAT THERE WAS NO IMPACT TO PATIENT DIAGNOSIS DUE TO THIS EVENT. THE FAS FURTHER NOTED "CUSTOMER STATES THAT 9 SLIDES DID NOT STAIN/INCONSISTENT STAINING AND SOMEHOW COVERTILES ARE NOT PLACED BACK ON THE SLIDES DURING PROCESSING." THE FAS DOCUMENTED THE AFFECTED PATIENT SLIDES WERE DERIVED FROM ONE (1) "[NAME] DAB" TISSUE ST

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17883021 ↗

Other Slide Stainer, Automated devices

All Slide Stainer, Automated devices →