BOND-PRIME automated stainerRecalls & reported problems
Leica Biosystems Melbourne Pty Ltd · The BOND system is designed for immunostaining histological specimens mounted on microscope slides. · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 09349458004828), received between Nov 26, 2025 and Nov 26, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Calibration Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Calibration Problem2
- Off-Label Use2
Patient problems
- Insufficient Information2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 26, 2025Off-Label Use; Calibration Problem; Off-Label Use; Calibration Problem
ON (B)(6) 2025, THE CUSTOMER CONTACTED LEICA BIOSYSTEMS AND REPORTED "FALSE POSITIVE ON HER 2 CORRELATING WITH ASI IMAGER." THE INFORMATION AVAILABLE DOES NOT IDENTIFY THE SPECIFIC STAINING RUN(S) WHERE "FALSE POSITIVE ON HER 2" WAS REPORTED BY THE CUSTOMER. PER THE CUSTOMER THE INFORMATION DETAILS OF THIS EVENT RESULTED IN A 23-24% INCREASE IN POSITIVE CASES FROM "LAST YEAR." THE CUSTOMER DOCUMENTED THEY WERE CONTACTED BY THE CLINICIAN WITH CONCERNS AS COMPARED TO LAST YEAR "...THEIR HER2 POSITIVE PERCENTAGES WERE HIGH...COMPARING WHAT THEIR ASI IMAGING SCANNER SCORED THEM AS VS WHAT THEIR SENT-OUT TEST RESULTS WERE FROM [LABORATORY]. [LABORATORY] STATING THAT THEY WERE EQUIVOCAL, FISH NEGATIVE. CUSTOMER SAID THAT THEIR HER2 BREAST CASES THAT WERE SCORED 2+ REALLY STATED AS LOW 3+ ON THEIR IMAGING SYSTEM. PATIENTS UNDERWENT TREATMENT AS SUCH." TWENTY (20) PATIENT CASES HAVE BEEN AFFECTED, WITH "PATIENT'S GIVEN TREATMENT, TREATMENT MORE TOXIC ON PATIENT'S ALREADY UNDERGOING TREATMENTS." INFORMATION PROVIDED BY THE LEICA SENIOR APPLICATIONS SPECIALIST DETAILED THE "CUSTOMER WAS MADE AWARE THAT THIS WAS A CLASS I QUALITATIVE ANTIBODY USED ON THEIR BOND PRIME INSTRUMENT. I MADE SURE T
Patient problems: Insufficient Information
MAUDE report 23661348 ↗InjuryNov 26, 2025Off-Label Use; Calibration Problem; Off-Label Use; Calibration Problem
ON (B)(6) 2025, THE CUSTOMER CONTACTED LEICA BIOSYSTEMS AND REPORTED ¿FALSE POSITIVE ON HER 2 CORRELATING WITH ASI IMAGER¿. THE INFORMATION AVAILABLE DOES NOT IDENTIFY THE SPECIFIC STAINING RUN(S) WHERE ¿FALSE POSITIVE ON HER 2¿ WAS REPORTED BY THE CUSTOMER. PER THE CUSTOMER THE INFORMATION DETAILS OF THIS EVENT RESULTED IN A 23-24% INCREASE IN POSITIVE CASES FROM ¿LAST YEAR¿. THE CUSTOMER DOCUMENTED THEY WERE CONTACTED BY THE CLINICIAN WITH CONCERNS AS COMPARED TO LAST YEAR ¿¿THEIR HER2 POSITIVE PERCENTAGES WERE HIGH...COMPARING WHAT THEIR ASI IMAGING SCANNER SCORED THEM AS VS WHAT THEIR SENT OUT TEST RESULTS WERE FROM [LABORATORY]. [LABORATORY] STATING THAT THEY WERE EQUIVOCAL, FISH NEGATIVE. CUSTOMER SAID THAT THEIR HER2 BREAST CASES THAT WERE SCORED 2+ REALLY STATED AS LOW 3+ ON THEIR IMAGING SYSTEM. PATIENTS UNDERWENT TREATMENT AS SUCH.¿ TWENTY (20) PATIENT CASES HAVE BEEN AFFECTED, WITH ¿PATIENT¿S GIVEN TREATMENT, TREATMENT MORE TOXIC ON PATIENT¿S ALREADY UNDERGOING TREATMENTS.¿ INFORMATION PROVIDED BY THE LEICA SENIOR APPLICATIONS SPECIALIST DETAILED THE ¿CUSTOMER WAS MADE AWARE THAT THIS WAS A CLASS I QUALITATIVE ANTIBODY USED ON THEIR BOND PRIME INSTRUMENT. I MADE SURE TO
Patient problems: Insufficient Information
MAUDE report 23659211 ↗
Other Slide Stainer, Automated devices
- Autostainer Link 48 AS480Agilent Technologies Singapore Pte Ltd · Slide stainer, automated1 recall723 reports
- Dako CoverStainer CS100Agilent Technologies Singapore Pte Ltd · Slide stainer, automated135 reports
- Artisan Link Pro AR310Agilent Technologies Singapore Pte Ltd · Slide stainer, automated90 reports
- VENTANA HE 600 System 06917259001Ventana Medical Systems, Inc. · Slide stainer, automated1 recall3 reports
- BenchMark ULTRA 110vVentana Medical Systems, Inc. · SLIDE STAINER, AUTOMATED37 reports
- Dako Omnis GI100Agilent Technologies Singapore Pte Ltd · Slide stainer, automated22 reports
- PREVI® COLOR GRAM V2 414292Biomerieux SA · Slide stainer, automated20 reports
- BOND-III processing moduleLeica Biosystems Melbourne Pty Ltd · Slide Stainer, Automated5 reports