BOND-PRIME automated stainerRecalls & reported problems

Leica Biosystems Melbourne Pty Ltd · The BOND system is designed for immunostaining histological specimens mounted on microscope slides. · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 09349458004828), received between Nov 26, 2025 and Nov 26, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Calibration Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Calibration Problem2
  • Off-Label Use2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2025Off-Label Use; Calibration Problem; Off-Label Use; Calibration Problem

    ON (B)(6) 2025, THE CUSTOMER CONTACTED LEICA BIOSYSTEMS AND REPORTED "FALSE POSITIVE ON HER 2 CORRELATING WITH ASI IMAGER." THE INFORMATION AVAILABLE DOES NOT IDENTIFY THE SPECIFIC STAINING RUN(S) WHERE "FALSE POSITIVE ON HER 2" WAS REPORTED BY THE CUSTOMER. PER THE CUSTOMER THE INFORMATION DETAILS OF THIS EVENT RESULTED IN A 23-24% INCREASE IN POSITIVE CASES FROM "LAST YEAR." THE CUSTOMER DOCUMENTED THEY WERE CONTACTED BY THE CLINICIAN WITH CONCERNS AS COMPARED TO LAST YEAR "...THEIR HER2 POSITIVE PERCENTAGES WERE HIGH...COMPARING WHAT THEIR ASI IMAGING SCANNER SCORED THEM AS VS WHAT THEIR SENT-OUT TEST RESULTS WERE FROM [LABORATORY]. [LABORATORY] STATING THAT THEY WERE EQUIVOCAL, FISH NEGATIVE. CUSTOMER SAID THAT THEIR HER2 BREAST CASES THAT WERE SCORED 2+ REALLY STATED AS LOW 3+ ON THEIR IMAGING SYSTEM. PATIENTS UNDERWENT TREATMENT AS SUCH." TWENTY (20) PATIENT CASES HAVE BEEN AFFECTED, WITH "PATIENT'S GIVEN TREATMENT, TREATMENT MORE TOXIC ON PATIENT'S ALREADY UNDERGOING TREATMENTS." INFORMATION PROVIDED BY THE LEICA SENIOR APPLICATIONS SPECIALIST DETAILED THE "CUSTOMER WAS MADE AWARE THAT THIS WAS A CLASS I QUALITATIVE ANTIBODY USED ON THEIR BOND PRIME INSTRUMENT. I MADE SURE T

    Patient problems: Insufficient Information

    MAUDE report 23661348 ↗
  • InjuryNov 26, 2025Off-Label Use; Calibration Problem; Off-Label Use; Calibration Problem

    ON (B)(6) 2025, THE CUSTOMER CONTACTED LEICA BIOSYSTEMS AND REPORTED ¿FALSE POSITIVE ON HER 2 CORRELATING WITH ASI IMAGER¿. THE INFORMATION AVAILABLE DOES NOT IDENTIFY THE SPECIFIC STAINING RUN(S) WHERE ¿FALSE POSITIVE ON HER 2¿ WAS REPORTED BY THE CUSTOMER. PER THE CUSTOMER THE INFORMATION DETAILS OF THIS EVENT RESULTED IN A 23-24% INCREASE IN POSITIVE CASES FROM ¿LAST YEAR¿. THE CUSTOMER DOCUMENTED THEY WERE CONTACTED BY THE CLINICIAN WITH CONCERNS AS COMPARED TO LAST YEAR ¿¿THEIR HER2 POSITIVE PERCENTAGES WERE HIGH...COMPARING WHAT THEIR ASI IMAGING SCANNER SCORED THEM AS VS WHAT THEIR SENT OUT TEST RESULTS WERE FROM [LABORATORY]. [LABORATORY] STATING THAT THEY WERE EQUIVOCAL, FISH NEGATIVE. CUSTOMER SAID THAT THEIR HER2 BREAST CASES THAT WERE SCORED 2+ REALLY STATED AS LOW 3+ ON THEIR IMAGING SYSTEM. PATIENTS UNDERWENT TREATMENT AS SUCH.¿ TWENTY (20) PATIENT CASES HAVE BEEN AFFECTED, WITH ¿PATIENT¿S GIVEN TREATMENT, TREATMENT MORE TOXIC ON PATIENT¿S ALREADY UNDERGOING TREATMENTS.¿ INFORMATION PROVIDED BY THE LEICA SENIOR APPLICATIONS SPECIALIST DETAILED THE ¿CUSTOMER WAS MADE AWARE THAT THIS WAS A CLASS I QUALITATIVE ANTIBODY USED ON THEIR BOND PRIME INSTRUMENT. I MADE SURE TO

    Patient problems: Insufficient Information

    MAUDE report 23659211 ↗

Other Slide Stainer, Automated devices

All Slide Stainer, Automated devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.