Skip to content

NervAlign Nerve Cuff NA-NC3040Recalls & reported problems

Renerve Limited · An absorbable collagen membrane implant that provides non-constricting protection for peripheral nerves. It is designed to be an interface between the nerve and the surrounding tissue and used for the repair of peripheral nerves in which there is no gap or where a gap closure is achieved by flexion of the extremity. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 09369998231904), received between Jan 28, 2026 and Jan 28, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection1

Patient problems

  • Abnormal scarring1
  • Implant Pain1
  • Sinusitis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 28, 2026Device Appears to Trigger Rejection

    THE SURGEON IMPLANTED A 30X40MM NERVALIGN NERVE CUFF ON (B)(6) 2025 AND THE PATIENT DID NOT IMMEDIATELY REPORT ANY ISSUES POST-OPERATIVELY. THE SURGEON DID NOT SEE THE PATIENT AGAIN UNTIL APPROXIMATELY ON (B)(6) 2025 (~5 MONTHS), THE PATIENT REPORTED SIGNIFICANT PAIN AND PRESENTED WITH HYPERTROPHIC SCARS. THE SURGEON REVISION PROCEDURE ON (B)(6) 2025 WAS TO ADDRESS NERVE ENTRAPMENT PRESUMED TO BE FROM A PREVIOUS SURGERY, WHICH HE DID NOT PERFORM. THE SURGEON CUT THE 30X40MM CUFF IN HALF AND USED THE MATERIAL IN BOTH PLACES WHERE CUTANEOUS SCARRING PRESENTS. NOTE: THE NERVE CUFF IS ALSO PROVIDED IN 10MMX10MM SIZE, USING MEMBRANES CUT INTO HALF IS NOT EXPECTED TO BE PROBLEMATIC. THE LINEAR INCISION (FULLY HEALED AND JUST BEHIND THE MALLEOLUS) FROM THE FIRST SURGERY IS OBSERVABLE AT THE PATIENT'S POSTEROLATERAL ANKLE. THE SURGEON SHARED THAT HIS PATIENT DOES HAVE ABNORMAL SENSITIVITIES / MEDICATION ALLERGIES TO ESTRADIOL AND NORELGESTROMIN. NO PREVIOUS KELOID FORMATION WAS OBSERVED FOLLOWING THE PRIMARY PROCEDURE, AND THE PATIENT DID NOT SHARE ANY KNOWN SENSITIVITY TO PORCINE MATERIALS. SINCE THE PATIENT CONTINUES TO EXPERIENCE INTERMITTENT SWELLING AND IS 5-MONTHS POST-OP, THE SURGEO

    Patient problems: Abnormal scarring; Implant Pain; Sinusitis

    MAUDE report 24197164 ↗

Other Cuff, Nerve devices

All Cuff, Nerve devices →