Fisher & Paykel Healthcare BC161-10Recalls & reported problems
Fisher & Paykel Healthcare Limited · BC161 Bubble CPAP System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 09420012412430), received between Nov 8, 2020 and Nov 20, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Connection Problem4
- Unintended Movement4
- Device Dislodged or Dislocated2
- Leak/Splash1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Unintended Movement; Unintended Movement
A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE CPAP PROBE OF THE BUBBLE CPAP GENERATOR SLIPPED OUT OF POSITION. HOWEVER, THE UNIT WAS FURTHER INSPECTED BY THE HEALTHCARE FACILITY AND NO ISSUES WERE IDENTIFIED WITH THE DEVICE ITSELF. THE BUBBLE CPAP GENERATOR IS PART OF THE BC161-10, BUBBLE CPAP SYSTEM KIT (SUBJECT DEVICE). THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23613135 ↗MalfunctionNov 6, 2025Unintended Movement; Unintended Movement
A HEALTHCARE FACILITY IN TEXAS REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE CPAP PROBE OF THE THREE BUBBLE CPAP GENERATORS WAS SLIPPING DOWN OUT OF POSITION. THE BUBBLE CPAP GENERATOR IS PART OF THE BC161-10, BUBBLE CPAP SYSTEM KIT (SUBJECT DEVICE). THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23489973 ↗MalfunctionOct 5, 2025Connection Problem; Connection Problem
A HEALTHCARE FACILITY IN GEORGIA REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A TEMPERATURE/FLOW PROBE WAS LOOSE AND "FALLING OUT" FROM THE TEMPERATURE PROBE PORT OF A BC161-10 BUBBLE CPAP SYSTEM BEFORE PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23218949 ↗MalfunctionJun 8, 2025Connection Problem; Connection Problem
A HEALTHCARE FACILITY IN ALASKA REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT TEMPERATURE/FLOW PROBES WERE "FALLING OFF" FROM A TEMPERATURE PROBE PORT OF A BC161-10 BUBBLE CPAP SYSTEM. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22164756 ↗MalfunctionMar 19, 2025Connection Problem; Device Dislodged or Dislocated; Connection Problem; Device Dislodged or Dislocated
A HEALTHCARE FACILITY REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A TEMPERATURE PROBE WAS OBSERVED TO BE LOOSE AND DISLODGING AT THE PATIENT-END TEMPERATURE PROBE PORT OF AN INFANT BREATHING CIRCUIT DURING PATIENT USE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21649597 ↗MalfunctionMar 19, 2025Connection Problem; Device Dislodged or Dislocated; Connection Problem; Device Dislodged or Dislocated
A HEALTHCARE FACILITY REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT THE TEMPERATURE PROBE WAS OBSERVED TO BE LOOSE AND DISLODGING AT THE PATIENT-END TEMPERATURE PROBE PORT OF TWO INFANT BREATHING CIRCUITS DURING SET UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21649596 ↗MalfunctionSep 26, 2024Unintended Movement; Unintended Movement
A HEALTHCARE FACILITY IN TEXAS REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE PEEP GAUGE OF THE BC161 BUBBLE CPAP SYSTEM WAS NOT STAYING IN PLACE DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20315753 ↗MalfunctionJun 26, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE BC161 BUBBLE CPAP SYSTEM HAD CUTS IN THE CIRCUITS. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19620688 ↗MalfunctionJun 19, 2023Unintended Movement; Unintended Movement
A DISTRIBUTOR IN INDIA REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE CPAP PROBE OF A BUBBLE CPAP GENERATOR WAS LOOSE. THE BUBBLE CPAP GENERATOR IS PART OF THE BC161 BUBBLE CPAP SYSTEM KIT. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17159781 ↗MalfunctionNov 8, 2020Leak/Splash; Leak/Splash
A DISTRIBUTOR IN (B)(4) REPORTED ON BEHALF OF A HEALTHCARE FACILITY VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER BASE WAS LEAKING WATER DURING USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10806672 ↗
Other Ventilator, Non-Continuous (Respirator) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page