Fisher & Paykel Healthcare BC161-10Recalls & reported problems

Fisher & Paykel Healthcare Limited · BC161 Bubble CPAP System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 09420012412430), received between Nov 8, 2020 and Nov 20, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem4
  • Unintended Movement4
  • Device Dislodged or Dislocated2
  • Leak/Splash1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2025Unintended Movement; Unintended Movement

    A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE CPAP PROBE OF THE BUBBLE CPAP GENERATOR SLIPPED OUT OF POSITION. HOWEVER, THE UNIT WAS FURTHER INSPECTED BY THE HEALTHCARE FACILITY AND NO ISSUES WERE IDENTIFIED WITH THE DEVICE ITSELF. THE BUBBLE CPAP GENERATOR IS PART OF THE BC161-10, BUBBLE CPAP SYSTEM KIT (SUBJECT DEVICE). THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23613135 ↗
  • MalfunctionNov 6, 2025Unintended Movement; Unintended Movement

    A HEALTHCARE FACILITY IN TEXAS REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE CPAP PROBE OF THE THREE BUBBLE CPAP GENERATORS WAS SLIPPING DOWN OUT OF POSITION. THE BUBBLE CPAP GENERATOR IS PART OF THE BC161-10, BUBBLE CPAP SYSTEM KIT (SUBJECT DEVICE). THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23489973 ↗
  • MalfunctionOct 5, 2025Connection Problem; Connection Problem

    A HEALTHCARE FACILITY IN GEORGIA REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A TEMPERATURE/FLOW PROBE WAS LOOSE AND "FALLING OUT" FROM THE TEMPERATURE PROBE PORT OF A BC161-10 BUBBLE CPAP SYSTEM BEFORE PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23218949 ↗
  • MalfunctionJun 8, 2025Connection Problem; Connection Problem

    A HEALTHCARE FACILITY IN ALASKA REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT TEMPERATURE/FLOW PROBES WERE "FALLING OFF" FROM A TEMPERATURE PROBE PORT OF A BC161-10 BUBBLE CPAP SYSTEM. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22164756 ↗
  • MalfunctionMar 19, 2025Connection Problem; Device Dislodged or Dislocated; Connection Problem; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A TEMPERATURE PROBE WAS OBSERVED TO BE LOOSE AND DISLODGING AT THE PATIENT-END TEMPERATURE PROBE PORT OF AN INFANT BREATHING CIRCUIT DURING PATIENT USE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21649597 ↗
  • MalfunctionMar 19, 2025Connection Problem; Device Dislodged or Dislocated; Connection Problem; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT THE TEMPERATURE PROBE WAS OBSERVED TO BE LOOSE AND DISLODGING AT THE PATIENT-END TEMPERATURE PROBE PORT OF TWO INFANT BREATHING CIRCUITS DURING SET UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21649596 ↗
  • MalfunctionSep 26, 2024Unintended Movement; Unintended Movement

    A HEALTHCARE FACILITY IN TEXAS REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE PEEP GAUGE OF THE BC161 BUBBLE CPAP SYSTEM WAS NOT STAYING IN PLACE DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20315753 ↗
  • MalfunctionJun 26, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE BC161 BUBBLE CPAP SYSTEM HAD CUTS IN THE CIRCUITS. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19620688 ↗
  • MalfunctionJun 19, 2023Unintended Movement; Unintended Movement

    A DISTRIBUTOR IN INDIA REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE CPAP PROBE OF A BUBBLE CPAP GENERATOR WAS LOOSE. THE BUBBLE CPAP GENERATOR IS PART OF THE BC161 BUBBLE CPAP SYSTEM KIT. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17159781 ↗
  • MalfunctionNov 8, 2020Leak/Splash; Leak/Splash

    A DISTRIBUTOR IN (B)(4) REPORTED ON BEHALF OF A HEALTHCARE FACILITY VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER BASE WAS LEAKING WATER DURING USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10806672 ↗

Other Ventilator, Non-Continuous (Respirator) devices

All Ventilator, Non-Continuous (Respirator) devices →