Skip to content

FISHER & PAYKEL HEALTHCARE 900MR190Recalls & reported problems

Fisher & Paykel Healthcare Limited · WATER FEED SET · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 09420012429780), received between Feb 23, 2023 and Feb 23, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2023Break; Break

    A HEALTHCARE FACILITY IN CANADA REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE CLAMP OF A 900MR190 WATER FEEDSET WAS FOUND BROKEN DURING USE CAUSING THE MR250 MANUAL FEED ADULT HUMIDIFICATION CHAMBER TO OVERFILL. IT WAS ALSO REPORTED THAT THE "WATER FLOWED INTO THE PATIENT'S BREATHING CIRCUIT AND INTO THE PATIENT'S AIRWAY". THERE WERE NO ADVERSE PATIENT OUTCOMES REPORTED AS A RESULT OF THIS EVENT. HOWEVER, FURTHER INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16433244 ↗

Other Generator, Oxygen, Portable devices

All Generator, Oxygen, Portable devices →