i-STAT 03P77-25Recalls & reported problems
Abbott Point Of Care Inc. · EG6+ CARTRIDGE · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10054749000057), received between Sep 22, 2017 and Mar 11, 2024 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
18 other recalls cover devices cleared under the same FDA submission (K940918); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 1, 2026Open, Classified
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
Z-2000-2026 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings8
- False Positive Result2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Information2
- Anemia1
- Chest Pain1
- Heart Failure1
- Inflammation1
- Ischemic Heart Disease1
- Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON 04-MAR-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT RESULT ON A 72 YEAR OLD MALE PATIENT WITH MALIGNANT NEOPLASM OF LEFT KIDNEY. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED HGB(G/DL) HCT(%) I-STAT (B)(6) 2024 09:50 IMMEDIATELY 14.3 42 SYSMEX (B)(6) 2024 11:14 11:35 7.2 25.8 AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18877163 ↗MalfunctionMar 11, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT RESULT ON A 74 YEAR OLD FEMALE PATIENT WITH PERIPHERAL VASCULAR DISEASE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD DATE COLLECTED TESTED HGB(G/DL) HCT(%) I-STAT ON (B)(6) 2024 , 06:26 IMMEDIATELY , 12.2 36. SYSMEX ON (B)(6) 2024 , 11:34 12:15 , 6.5 20.9 .SYSMEX ON (B)(6) 2024 , 12:25 13:09 , 6.2 20.1. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Patient problems: Unspecified Vascular Problem
MAUDE report 18877162 ↗MalfunctionApr 5, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2022 ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT PO2 AND SO2 RESULTS ON A PATIENT . THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD: I-STAT; I-STAT. DATE : 30-MAR-2022 ; 30-MAR-2022. TEST/COLLECT TIME: 08:31; 08:33. PO2(MMHG): 37;80. SO2(%) : 66; 95. SAMPLE:A;B AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14013188 ↗MalfunctionJun 22, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON 11-JUN-2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT HEMOGLOBIN RESULT OF 16.0 G/DL ON A FEMALE PATIENT WITH ANEMIA. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD: I-STAT. DATE : (B)(6) 2021. TIME: 11:19. HGB: 16.0 G/DL. METHOD: LAB. DATE: NI. TIME: NI. HGB: LAB. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. THE CUSTOMER HAS ALSO OPTED TO REPLACE THE I-STAT ANALYZER SN (B)(4) USED DURING THE EVENT. THE ANALYZER WAS REPLACED AT NO CHARGE AND WILL INVESTIGATED SEPARATELY.
Patient problems: Anemia
MAUDE report 12045523 ↗MalfunctionJan 19, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT POTASSIUM RESULT OF 6.7 MMOL/L ON A FEMALE PATIENT SEEN FOR A PLANNED PROCEDURE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. THE CUSTOMER HAS NOT RESPONDED TO MULTIPLE ATTEMPTS FOR INFORMATION. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. (B)(6). AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11193093 ↗MalfunctionMar 20, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2019, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULT OF 6.0 ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. (B)(6). AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXIST. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE PATIENT WAS TESTED IN (B)(6) OF 2018 BUT THE ACTUAL DAY IS UNKNOWN. THE CUSTOMER PROVIDED TWO I-STAT RESULTS NO PATIENT INFORMATION, SAMPLE AND COLLECTION TIMES WITH MULTIPLE ATTEMPTS TO OBTAIN INFORMATION. CARTRIDGE LOT IS UNKNOWN AT THIS TIME AND COULD POTENTIALLY RENDER THE INVESTIGATION INCONCLUSIVE.
Patient problems: No Information
MAUDE report 8435836 ↗MalfunctionOct 24, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2018, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT HEMOGLOBIN RESULT ON A (B)(6) MALE PATIENT WITH EPIGASTRIC PAIN X 3 DAYS, AND VOMITING X3 EPISODES. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD, DATE, TIME, HCT, HGB. I-STAT, (B)(6) 2018, UNK, 43%, 14.6 G/DL; LAB, (B)(6) 2018, 0538, 26.6%, 8.6 G/DL. COLLECTION TIMES ARE UNKNOWN. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Patient problems: Inflammation; Pain; Septic Shock
MAUDE report 8001395 ↗MalfunctionOct 9, 2017Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT POTASSIUM (K) RESULT ON A PATIENT. THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, TEST TIME: UNK, K: >9.0 . RESULTS LAB: NORMAL (A PRECISE RESULT IS NOT AVAILABLE). FOLLOW UP TESTING WAS DONE IN THE LAB. RESULTS LAB: NORMAL (A PRECISE RESULT IS NOT AVAILABLE). AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY APOC, HOWEVER THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED MAY 8TH TO 12TH, 2017 AT ABBOTT POINT OF CARE (B)(4).
Patient problems: No Information
MAUDE report 6931087 ↗MalfunctionSep 22, 2017False Positive Result; False Positive Result
MalfunctionSep 22, 2017False Positive Result; False Positive Result
ON (B)(6) 2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG6+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT ON A (B)(6) MALE PATIENT WITH ISCHEMIC HEART DISEASE AND CHEST PAIN. THE PATIENT WAS ADMITTED, WAS STILL AN IN-PATIENT AS OF (B)(6) 2017 AND IN STABLE CONDITION. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, TYPE: ARTERIAL, DATE: (B)(6) 2017, COLLECTED: 13:08, TESTED: 13:08, NA: 116 K:2.3, HB: 10.9, HCT:32.0; METHOD: I-STAT, TYPE: VENOUS, DATE: (B)(6) 2017, COLLECTED: 13:40, TESTED: 14:05, NA: 138, K: 4.0 HB: 11.6, HCT:34.4. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.
Patient problems: Chest Pain; Ischemic Heart Disease
MAUDE report 6888207 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K940918. One submission often covers many models, so check the product and lot details in each record.
- Z-1999-2026 · Apr 1, 2026 · Open, Classifiedi-STAT EG7+ cartridge; List Number: 03P76-25;
- Z-2582-2025 · Aug 21, 2025 · Open, Classifiedi-STAT CG4+ cartridge (white). List Number: 03P85-25.
- Z-0881-2021 · Dec 22, 2020 · TerminatedAbbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary wh
- Z-0880-2021 · Dec 22, 2020 · TerminatedAbbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capill
- Z-0615-2019 · Sep 13, 2018 · TerminatedAbbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capillary whole blood. List Number/UDI# : 03P88-25 ,(01)10054749
- Z-1306-2014 · Jan 1, 2014 · Terminatedi-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based d
- Z-2133-2012 · Jan 1, 2011 · TerminatedI-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, c
- Z-0595-2011 · Sep 20, 2010 · Terminatedi-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and
- Z-0593-2011 · Sep 20, 2010 · Terminatedi-STAT Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. the 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and the i-st
- Z-0596-2011 · Sep 20, 2010 · TerminatedMilitary Kit i-STAT1; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and the
- Z-0594-2011 · Sep 20, 2010 · TerminatedFUSO Analyzer 300F Abbott Point of Care, Princeton, NJ 08540 Model 200. the 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and the i
- Z-0929-2011 · Aug 3, 2010 · Terminatedi-STAT CG8+ Cartridge; pH, pCO2, PO2, NA, K, iCa, Glu, Hct; IVD Abbott Point of Care The test for PCO2, as part of the i-STAT System, is intended for use in the vitro quantificati
- Z-0930-2011 · Aug 3, 2010 · Terminatedi-STAT G3+ Cartridge; pH, pCO2, PO2; IVD Abbott Point of Care The test for PCO2, as part of the i-STAT System, is intended for use in the vitro quantification of carbon dioxide pa
- Z-1794-2010 · Feb 5, 2010 · Terminatedi-STAT EG6+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood
- Z-1793-2010 · Feb 5, 2010 · Terminatedi-STAT EG7+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood
…and 3 more.
Other ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH devices
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
- ABL800 FLEX 393-800Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls81 reports
- ABL90 FLEX Analyzer 393-090Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall112 reports
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- GEM Premier 5000 PAK BG/COOX 75 TEST 00055407504Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 600 TEST 00055360010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.