Skip to content

Medline Industries, Inc. DYND43030Recalls & reported problems

Medline Industries, Inc. · TUBE,ENDOTRACH,HVLP,CUFF,MURPHY,3.0MM · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10080196503773), received between Jul 1, 2026 and Jul 1, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

1 other recall cover devices cleared under the same FDA submission (K042683); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Material Integrity Problem1
  • Material Too Rigid or Stiff1

Patient problems

  • Swelling/ Edema1
  • Upper Respiratory Tract Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 1, 2026Difficult to Remove; Material Too Rigid or Stiff; Material Integrity Problem

    IT WAS REPORTED BY THE FACILITY THAT, "THE NEONATE WAS INTUBATED ON (B)(6) 2026 WITH A 3.0-CUFFED TUBE (MEDLINE BRAND). COULD NOT BE EXTUBATED IN NICU ON (B)(6) 2026 DUE TO THE FACT THAT THE ETT GOT STUCK IN THE LARYNX. MULTIPLE ATTEMPTS MADE BY RT AND PROVIDERS. THE CUFF WAS DEFLATED AND THE TAPE REMOVED BUT ETT SEEMED STUCK. DETERMINED AIRWAY WAS TOO SWOLLEN TO EXTUBATE AND ETT HAD TO BE RETAPED AND THIS ETT IS TO REMAIN IN PLACE UNTIL AIRWAY SWELLING SUBSIDES. PATIENT STARTED ON DECADRON. ENT CONSULTED. THIS NEONATE UNDERWENT AN ENT AIRWAY EVALUATION AND EXTUBATION IN THE OR AFTER REMOVAL OF ETT ON (B)(6) 2026. IT WAS FOUND THAT THE CUFF OF THIS ETT WAS EXCESSIVELY LARGE AND RIGID, WHICH CONTRIBUTED TO THE PROBLEM. IN ADDITION, AFTER FULL DEFLATION, THIS CUFF HAS LARGE, HARD, PROTRUDING EDGES THAT RESEMBLE THE MERCEDES-BENZ LOGO IN CROSS-SECTION".

    Patient problems: Swelling/ Edema; Upper Respiratory Tract Injury

    MAUDE report 25679414 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042683. One submission often covers many models, so check the product and lot details in each record.

  • Z-3172-2026 · Jul 31, 2026 · Open, Classified
    SunMed Tracheal Tube, Size 2.5; Model Number: 1-7333-25

Other Tube, tracheal (w/wo connector) devices

All Tube, tracheal (w/wo connector) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page