Medline DYND11553Recalls & reported problems
Medline Industries, Inc. · CATHETER,FOLEY,100% SILICONE,8FR 3ML,LF · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10080196848836), received between Aug 12, 2020 and Sep 4, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Failure to Deflate” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 28, 2019Class IITerminated
Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Z-2043-2020 · Root cause: Manufacturing material removal · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Deflate2
- Appropriate Device Problem Term/Code Not Available1
- Material Rupture1
- Obstruction of Flow1
Patient problems
- Discomfort2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Clinical Signs, Symptoms or Conditions1
- Urinary Retention1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Failure to Deflate
It was reported that 3 out of 7 catheters would not deflate in the morning when removing.
Patient problems: Discomfort
MAUDE report 50619011 ↗MalfunctionSep 3, 2026Failure to Deflate
Patient problems: Discomfort
MAUDE report 50588579 ↗MalfunctionSep 2, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50566652 ↗InjurySep 13, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE FACILITY THE CATHETER WAS DISLODGED FROM THE PATIENT, 'AND THE BALLOON APPEARED TO HAVE BURST'. PER THE FACILITY AN ADDITIONAL CATHETER WAS PLACED AND NO SERIOUS INJURY OCCURRED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15411248 ↗MalfunctionAug 12, 2020Obstruction of Flow
IT WAS REPORTED THE FOLEY CATHETERS WERE CAUSING THE END USER "MORE PRESSURE THAN USUAL AND NOT DRAINING."
Patient problems: Urinary Retention
MAUDE report 10395831 ↗
FDA clearance and classification
Cleared through 510(k) K082815 on Oct 8, 2008 (decision: Substantially Equivalent). Submitted by Well Lead Medical Instruments as “Well Lead Silicone And Latex Foley Catheters”. The FDA review took 13 days. FDA record ↗
The FDA regulates product code EZL (catheter, retention type, balloon) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Medline DYND11553
Has the Medline DYND11553 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2043-2020 (Class II), began on May 28, 2019: Loose silicone particulate was found to be present on the shaft of the silicone catheters. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Medline DYND11553?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Aug 12, 2020 and Sep 4, 2026. In those reports, the most common specific device problems are “Failure to Deflate” (2), “Material Rupture” (1) and “Obstruction of Flow” (1); the most common patient problems are “Discomfort” (2) and “Urinary Retention” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Medline DYND11553?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Medline DYND11553 FDA approved?
It was cleared through 510(k) K082815 on Oct 8, 2008. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Medline DYND11553, and is it still sold?
The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand Medline, model DYND11553). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON487 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death381 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON235 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON168 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON152 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page