AirLife™ 001350Recalls & reported problems

Vyaire Medical, Inc. · AirLife™ U/Connect-it Oxygen Tubing 7 feet (2.1 m) Crush-Resistant Oxygen Tubing, Vinyl Tipped · Class 1

All models of AirLife (25)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10190752116252), received between Oct 24, 2022 and Nov 7, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem4
  • Infusion or Flow Problem1

Patient problems

  • Low Oxygen Saturation3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT A MEDICAL DEVICE INCIDENT (MDI) WITH HARM OR WITH THE POTENTIAL FOR SERIOUS HARM/DEATH HAS BEEN REPORTED. - CARDINAL HEALTH CANADA PRODUCT COMPLAINT (B)(4). UNKNOWN PATIENT ARRIVED VIA EMS IN CRITICAL CONDITION FOLLOWING A SEVERE BURN INJURY. INTERMITTENT SPO2 CHALLENGES INCLUDING DURING INTUBATION WITH MANUAL VENTILATION VIA MASK, AND MANUAL VENTILATION VIA ETT. SPO2 54% AT THE LOWEST. PATIENT REQUIRING CONTINUOUS HIGH FLOW O2 THERAPY (15LPM VIA NON-REBREATHER, FOLLOWED BY CONTINUOUS MANUAL VENTILATION VIA MASK THEN ETT FOR 45 MINUTES, THEN EVENTUALLY 100% ON VENTILATOR, DUE TO PATIENT'S INJURIES. DURING THIS TIME ON 3 OCCASIONS, RRTS NOTICED BAGGER WAS DISCONNECTED FROM O2 TUBING ON THE BAGGER SIDE.

    Patient problems: Low Oxygen Saturation

    MAUDE report 23501763 ↗
  • InjuryNov 7, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT A MEDICAL DEVICE INCIDENT (MDI) WITH HARM OR WITH THE POTENTIAL FOR SERIOUS HARM/DEATH HAS BEEN REPORTED. - CARDINAL HEALTH CANADA PRODUCT COMPLAINT PCS251008IOELIY . UNKNOWN PATIENT ARRIVED VIA EMS IN CRITICAL CONDITION FOLLOWING A SEVERE BURN INJURY. INTERMITTENT SPO2 CHALLENGES INCLUDING DURING INTUBATION WITH MANUAL VENTILATION VIA MASK, AND MANUAL VENTILATION VIA ETT. SPO2 54% AT THE LOWEST. PATIENT REQUIRING CONTINUOUS HIGH FLOW O2 THERAPY (15LPM VIA NON-REBREATHER, FOLLOWED BY CONTINUOUS MANUAL VENTILATION VIA MASK THEN ETT FOR 45 MINUTES, THEN EVENTUALLY 100% ON VENTILATOR, DUE TO PATIENT'S INJURIES. DURING THIS TIME ON 3 OCCASIONS, RRTS NOTICED BAGGER WAS DISCONNECTED FROM O2 TUBING ON THE BAGGER SIDE. .

    Patient problems: Low Oxygen Saturation

    MAUDE report 23501762 ↗
  • InjuryNov 7, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT A MEDICAL DEVICE INCIDENT (MDI) WITH HARM OR WITH THE POTENTIAL FOR SERIOUS HARM/DEATH HAS BEEN REPORTED. - CARDINAL HEALTH CANADA PRODUCT COMPLAINT (B)(4). UNKNOWN PATIENT ARRIVED VIA EMS IN CRITICAL CONDITION FOLLOWING A SEVERE BURN INJURY. INTERMITTENT SPO2 CHALLENGES INCLUDING DURING INTUBATION WITH MANUAL VENTILATION VIA MASK, AND MANUAL VENTILATION VIA ETT. SPO2 54% AT THE LOWEST. PATIENT REQUIRING CONTINUOUS HIGH FLOW O2 THERAPY (15LPM VIA NON-REBREATHER, FOLLOWED BY CONTINUOUS MANUAL VENTILATION VIA MASK THEN ETT FOR 45 MINUTES, THEN EVENTUALLY 100% ON VENTILATOR, DUE TO PATIENT'S INJURIES. DURING THIS TIME ON 3 OCCASIONS, RRTS NOTICED BAGGER WAS DISCONNECTED FROM O2 TUBING ON THE BAGGER SIDE.

    Patient problems: Low Oxygen Saturation

    MAUDE report 23501759 ↗
  • MalfunctionJul 18, 2025Connection Problem; Connection Problem

    IT WAS REPORTED: LOOSE CONNECTION BETWEEN THE TUBING AND THE EQUIPMENT; IT LITERALLY FALLS OFF WITH THE SLIGHTEST BIT OF MOTION THE MATERIAL FEELS STIFF. NO REPORTED INJURY. WHILE NO HARM OR INJURY WAS REPORTED, ANY COMPONENT THAT IS PART OF THE VENT CIRCUIT (FLEX CONNECTOR, DOUBLE SWIVEL ELBOW, Y-ADAPTER, T-PIECE, ETT ADAPTER, ETC.) BREAKS, CRACKS, OR LEAKS AFTER THE VENT CIRCUIT HAS BEEN ESTABLISHED ON THE PATIENT COULD CAUSE A DELAY IN CARE AND OR LEAD TO A SERIOUS INJURY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22549423 ↗
  • MalfunctionOct 24, 2022Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE TUBING OXYGEN 7FT CRUSH RESISTANT 50/CS HAD AN ISSUE WITH THE NEBULIZERS THAT DO NOT PRODUCE MIST AND SOME JUST BUBBLES. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15664094 ↗

FDA clearance and classification

The FDA regulates product code CAT (cannula, nasal, oxygen) under 21 CFR 868.5340 as a Class I device. 21 CFR 868.5340 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AirLife™ 001350

Has the AirLife™ 001350 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10190752116252), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AirLife™ 001350?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 24, 2022 and Nov 7, 2025. In those reports, the most common specific device problems are “Connection Problem” (4) and “Infusion or Flow Problem” (1); the most common patient problem is “Low Oxygen Saturation” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AirLife™ 001350?

None of the reports are classified as deaths, 3 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AirLife™ 001350 FDA approved?

Devices of this type (product code CAT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Vyaire Medical, Inc.

Who makes the AirLife™ 001350, and is it still sold?

The FDA's device database lists Vyaire Medical, Inc. as the company responsible for it (brand AirLife™, model 001350). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.