Infant Flow™ 7772000LPRecalls & reported problems
Vyaire Medical, Inc. · Infant Flow LP nCPAP Generator without Prongs · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10190752152687), received between Aug 25, 2023 and Sep 25, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2024
SITUATION/PRIMARY CONCERN: DEFECT - VYAIRE INFANT FLOW GENERATOR NCPAP. BACKGROUND: RT WAS CALLED TO PT BESIDE BY RN TO LOOK AT EQUIPMENT. PT IS RECEIVING NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE (NCPAP) THROUGH THE VYAIRE SIPAP MACHINE. RN NOTICED A HOLE IN THE EXPIRATORY LIMB OF THE INFANT FLOW TUBING WHICH WAS LEAKING WATER ONTO PT ISOLETTE BEDDING. RT GRABBED NEW INFANT FLOW GENERATOR KIT FROM CLEAN SUPPLY. WHEN EXCHANGING THE FLOW GENERATOR, RT NOTICED TWO HOLES ON THE NEW FLOW GENERATOR EXPIRATORY LIMB AS WELL. (LOT: 0004262729) RT GRABBED SECOND FLOW GENERATOR KIT WITH DIFFERENT LOT NUMBER, AND IT HAD A HOLE IN IT AS WELL. (LOT: 0004240543). RT CONFIRMED THAT THIS IS A MANUFACTURING ERROR AND HOLES SHOULD NOT BE IN EXPIRATORY LIMB. THIS ERROR DOES NOT EFFECT PT VENTILATION OR OXYGENATION. HOWEVER, DEPENDING ON WHERE THE HOLES ARE LOCATED, THE WATER CAN LEAK OUT CAUSING BEDDING TO GET WET AND BABY POTENTIALLY GETTING COLD. RECOMMENDATION: RT WILL PLACE TAPE AROUND EXPIRATORY LIMB HOLE. IF LEAKING CONTINUES TO BE A PROBLEM, RT WILL EXCHANGE VYAIRE SIPAP FOR SERVO-U VENTILATOR WITH FLEXI TRUNK NCPAP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20300704 ↗InjuryAug 25, 2023Break; Break
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 7772000LP PACKAGE ASSEMBLY,GENERATOR WITHOUT PRONGS BROKE IN HALF WHILE USE ON A BABY. FURTHERMORE, THERE WAS NO HARM DONE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17627949 ↗MalfunctionAug 25, 2023Break; Break
PLASTIC INTERFACE CONNECTING TO GENERATOR BROKE IN HALF WHILE USED ON BABY NOT DURING MANIPULATION OF INTERFACE. NOTED AS BABY WAS CRYING IN ISOLETTE. THE GENERATOR WAS CHANGED, NEW GENERATOR INTEGRITY WAS EXAMINED AND DETERMINED TO BE SAFE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17625153 ↗
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page