CARDINAL HEALTH SU130-401DRecalls & reported problems

Cardinal Health 200, LLC · JP 3-SPRING RES W/7FR PVC DRAIN/TR · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10192253016413), received between Jun 12, 2024 and Jun 12, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Microbial Contamination of Device” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2024Open, Classified

    Cardinal Health is issuing a lot number specific voluntary removal of Jackson-Pratt 3-Spring Reservoir kits due to specific products being shipped to users before undergoing sterilization.

    Z-1735-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Microbial Contamination of Device2

Patient problems

  • Post Operative Wound Infection2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2024Microbial Contamination of Device; Microbial Contamination of Device

    MW5154606: CUSTOMER REPORTED A PATIENT HAD A RIGHT ANTERIOR TOTAL HIP ARTHROPLASTY ON (B)(6) 2024, IN WHICH A JACKSON-PRATT 3-SPRING RESERVOIR KIT WITH PVC ROUND DRAIN, 7FR DISTRIBUTED BY CARDINAL HEALTH WAS PLACED. ON (B)(6) 2024 WE WERE NOTIFIED BY CARDINAL HEALTH OF A PRODUCT RECALL DUE TO PRODUCT NOT BEING STERILIZED PRIOR TO DISTRIBUTION. THIS PATIENT OBTAINED A POST OPERATIVE INFECTION DIAGNOSED AND WAS STARTED ON ORAL ANTIBIOTICS. PATIENT RECEIVED MULTIPLE DEBRIDEMENT¿S IN OFFICE AND EVENTUALLY REQUIRED NEGATIVE PRESSURE WOUND THERAPY. THIS PATIENT IS BELIEVED TO HAVE RECEIVED A NON-STERILE DRAIN RESULTING IN SURGICAL COMPLICATIONS. A WOUND CULTURE OBTAINED FROM A SURGICAL WOUND SHOULD HAVE NO GROWTH.

    Patient problems: Post Operative Wound Infection

    MAUDE report 19525214 ↗
  • InjuryJun 12, 2024Microbial Contamination of Device; Microbial Contamination of Device

    MW5154605: CUSTOMER REPORTED A PATIENT HAD A RIGHT ANTERIOR TOTAL HIP ARTHROPLASTY ON (B)(6) 2024, IN WHICH A JACKSON-PRATT 3-SPRING RESERVOIR KIT WITH PVC ROUND DRAIN, 7FR DISTRIBUTED BY CARDINAL HEALTH WAS PLACED. ON 3/28/24 WE WERE NOTIFIED BY CARDINAL HEALTH OF A PRODUCT RECALL DUE TO PRODUCT NOT BEING STERILIZED PRIOR TO DISTRIBUTION. THIS PATIENT OBTAINED A POST OPERATIVE INFECTION DIAGNOSED AND WAS STARTED ON ORAL ANTIBIOTICS. PATIENT RECEIVED MULTIPLE DEBRIDEMENT¿S IN OFFICE AND EVENTUALLY REQUIRED NEGATIVE PRESSURE WOUND THERAPY. THIS PATIENT IS BELIEVED TO HAVE RECEIVED A NON-STERILE DRAIN RESULTING IN SURGICAL COMPLICATIONS. A WOUND CULTURE OBTAINED FROM A SURGICAL WOUND SHOULD HAVE NO GROWTH.

    Patient problems: Post Operative Wound Infection

    MAUDE report 19525206 ↗