Skip to content

NAMIC 70055009 (UDI 10193489065695)Recalls & reported problems

Medline Industries, Inc. · SC 3W HP RA OFF NB - PG · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10193489065695), received between Aug 18, 2023 and Sep 30, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem2
  • Gas/Air Leak1
  • Material Too Rigid or Stiff1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952285 ↗
  • MalfunctionSep 29, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936133 ↗
  • MalfunctionSep 2, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50566239 ↗
  • MalfunctionAug 12, 2026Material Too Rigid or Stiff

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50228846 ↗
  • MalfunctionAug 18, 2023

    ELDERLY MALE WITH HISTORY OF ATRIAL FIBRILLATION. PROCEDURE: ELECTROPHYSIOLOGY STUDY WITH ABLATION. 3-WAY STOPCOCK WAS LEAKING ON THE LUER LOCK SIDE, WHICH COULD HAVE INTRODUCED AIR. TRIED TO PROBLEM SOLVE SOURCE AND CHANGED THE IV TUBING BUT IT STILL WAS LEAKING AIR. THE STOPCOCK WAS CHANGED AND THERE WAS NO FURTHER LEAKING. NO KNOWN HARM TO THE PATIENT, MINOR DELAY IN PROCEDURE. MANUFACTURER RESPONSE FOR STOPCOCK, NAMIC (PER SITE REPORTER), WILL OBTAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17571530 ↗

Other Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass devices

All Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page