Skip to content

Medline Industries, Inc. DYNJ0785310GRecalls & reported problems

Medline Industries, Inc. · MJR PLASTICS PK MV SURG CT-LF · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jan 7, 2026Class IIOpen, Classified

    Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

    Z-2099-2026 · Root cause: Process control · Recall details · FDA record ↗

  2. Jan 7, 2026Class IIOpen, Classified

    Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

    Z-1397-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code FTN (plastic surgery and accessories kit) under 21 CFR 878.3925 as a Class I device. 21 CFR 878.3925 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Medline Industries, Inc. DYNJ0785310G

Has the Medline Industries, Inc. DYNJ0785310G been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2099-2026 (Class II), began on Jan 7, 2026: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Medline Industries, Inc. DYNJ0785310G, and is it still sold?

The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand Medline Industries, Inc., model DYNJ0785310G). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Plastic surgery and accessories kit devices

All Plastic surgery and accessories kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page