See RES for complete list. Medline Convenience Kits: 1)… recallZ-2099-2026 · Medline Industries, LP · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA root cause category: Process control
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/14/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-
Recall details
- Recall number
- Z-2099-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- May 7, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 5497 units
- Product code
- FTN
- Distribution
- Worldwide - US Nationwide distribution.
Product description
See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 4) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182;
Code information (models, lots, serial numbers)
Show the full code information (25k characters)
1) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21DDA661; 2) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21FDA888; 3) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21GDB740; 4) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21HDB434; 5) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21JDB530; 6) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23HDB786; 7) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23IDB182; 8) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23LDC086; 9) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 24IDA403; 10) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 24JDA650; 11) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25ADB705; 12) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25BDA228; 13) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25BDB702; 14) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25FDA013; 15) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25GDB671; 16) CDS984853J, UDI-DI: 10195327694180(each), 40195327694181(case), Lot Number: 24DDC169; 17) CDS984853J, UDI-DI: 10195327694180(each), 40195327694181(case), Lot Number: 24EDA131; 18) DYKMBNDL117C, UDI-DI: 10195327371104(each), 40195327371105(case), Lot Number: 23DLA834; 19) DYKMBNDL158, UDI-DI: 10193489478266(each), 40193489478267(case), Lot Number: 21ABI978; 20) DYKMBNDL158, UDI-DI: 10193489478266(each), 40193489478267(case), Lot Number: 21CBE548; 21) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 20LBY951; 22) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21ABJ202; 23) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21BBH637; 24) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21DBN049; 25) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21IBV640; 26) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22EBA143; 27) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22HBN456; 28) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22IBC515; 29) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 22JBM366; 30) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23GBR888; 31) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23HBB877; 32) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23HBG507; 33) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23IBB851; 34) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24FBD539; 35) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24FBL546; 36) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24IBB931; 37) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25BBA340; 38) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25BBN216; 39) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25DBK034; 40) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25DBS731; 41) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25EBM448; 42) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25GBA895; 43) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25HBP137; 44) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25KBR870; 45) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25LBM160; 46) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21ABI990; 47) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21ABX025; 48) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21DBC569; 49) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22BBN306; 50) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22EBO665; 51) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22FBC879; 52) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 23LBI220; 53) DYKMBNDL182A, UDI-DI: 10195327633806(each), 40195327633807(case), Lot Number: 24BBN507; 54) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24DBR913; 55) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24IBL678; 56) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24IBQ469; 57) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24KBS891; 58) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 25ABL533; 59) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 25FBO935; 60) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21ABI992; 61) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21CBG305; 62) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21JBJ252; 63) DYKMBNDL183A, UDI-DI: 10195327246594(each), 40195327246595(case), Lot Number: 23BBI487; 64) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24DBJ591; 65) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24DBT970; 66) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24GBC829; 67) DYKMBNDL34A, UDI-DI: 10193489872033(each), 40193489872034(case), Lot Number: 21DBQ512; 68) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21ABJ002; 69) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21CBD335; 70) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21GBR352; 71) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21LBJ742; 72) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22BBU074; 73) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22GBM861; 74) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22IBC434; 75) DYKMBNDL39A, UDI-DI: 10195327241308(each), 40195327241309(case), Lot Number: 23ABH677; 76) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 24JBO020; 77) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25GBW751; 78) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25HBK505; 79) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25HBQ024; 80) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21ABL308; 81) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21BBM839; 82) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21GBR353; 83) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 20LBY967; 84) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABJ589; 85) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABM089; 86) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABO607; 87) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21BBH629; 88) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21CBG791; 89) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21ABJ013; 90) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21ABL345; 91) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21DBC571; 92) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21GBR324; 93) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBK340; 94) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBM983; 95) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBO272; 96) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21IBV702; 97) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21JBS176; 98) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22HBN457; 99) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22HBQ100; 100) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22IBR235; 101) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22JBC409; 102) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBC618; 103) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBC626; 104) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBJ150; 105) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBQ852; 106) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22KBO876; 107) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23FBP246; 108) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23GBL960; 109) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23JBJ695; 110) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23LBA275; 111) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24EBR715; 112) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24GBF481; 113) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24HBM099; 114) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24KBT018; 115) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25FBH462; 116) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25GBL984; 117) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25KBP763; 118) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 20XBB257; 119) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21ABJ216; 120) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21ABX017; 121) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21BBK531; 122) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21DBC549; 123) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21FBF907; 124) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21FBR398; 125) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21HBK310; 126) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21IBB097; 127) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21IBV643; 128) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21JBJ751; 129) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21KBY757; 130) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21LBO572; 131) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21LBY278; 132) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22ABG018; 133) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22ABL251; 134) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22BBL848; 135) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22CBX537; 136) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22EBS816; 137) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBJ019; 138) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBJ143; 139) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBQ856; 140) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22KBO921; 141) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23GBO451; 142) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23GBP831; 143) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23JBN862; 144) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23LBE774; 145) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24CBF892; 146) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24CBR114; 147) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24EBE953; 148) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24EBU597; 149) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24HBD157; 150) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24LBL035; 151) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBE752; 152) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBO685; 153) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBU089; 154) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25CBG915; 155) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25EBE325; 156) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25EBM007; 157) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25FBP527; 158) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25HBS118; 159) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21ABM088; 160) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21ABU822; 161) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21CBM835; 162) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21EBH614; 163) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21GBF538; 164) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21GBR338; 165) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21IBT011; 166) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21JBH111; 167) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21LBD628; 168) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21LBH705; 169) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22FBF307; 170) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22FBT013; 171) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22HBO907; 172) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22IBC435; 173) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22OBH455; 174) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24IBE175; 175) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24JBE336; 176) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24LBI135; 177) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24LBS883; 178) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25DBO560; 179) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25FBB154; 180) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25GBF486; 181) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25HBG553; 182) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25LBP864; 183) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25LBQ326; 184) DYNJ0785310G, UDI-DI: 10193489268966(each), 40193489268967(case), Lot Number: 23KMH075; 185) DYNJ0854650J, UDI-DI: 10193489832426(each), 40193489832427(case), Lot Number: 24IMF274; 186) DYNJ36469I, UDI-DI: 10195327513931(each), 40195327513932(case), Lot Number: 24BBU341; 187) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21ADA533; 188) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21BDB779; 189) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21CDB475; 190) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22GMF684; 191) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22HMH034; 192) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22IMF713; 193) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23AME665; 194) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23HMA900; 195) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23IMG016; 196) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23KMA468; 197) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23LMB124; 198) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24AMI035; 199) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24CMD088; 200) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24DMJ046; 201) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24GMA471; 202) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24IME867; 203) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24KMD041; 204) DYNJ44852K, UDI-DI: 10195327127589(each), 40195327127580(case), Lot Number: 22CMI118; 205) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23AMC374; 206) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23AMC530; 207) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23CMF343; 208) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23DMD096; 209) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23FMH424; 210) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23HME689; 211) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24AMB315; 212) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24CMC758; 213) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24GMI186; 214) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24KME600; 215) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25CME690; 216) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25EME929; 217) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25HMH591; 218) DYNJ50957G, UDI-DI: 10193489448863(each), 40193489448864(case), Lot Number: 23GBV065; 219) DYNJ55358G, UDI-DI: 10195327173500(each), 40195327173501(case), Lot Number: 23FBN323; 220) DYNJ56737A, UDI-DI: 10193489395594(each), 40193489395595(case), Lot Number: 23CME388; 221) DYNJ58334D, UDI-DI: 10195327515225(each), 40195327515226(case), Lot Number: 24ABU081; 222) DYNJ60210A, UDI-DI: 10193489985108(each), 40193489985109(case), Lot Number: 21KBP457; 223) DYNJ64927F, UDI-DI: 10195327143091(each), 40195327143092(case), Lot Number: 23FBV335; 224) DYNJ68844A, UDI-DI: 10195327216504(each), 40195327216505(case), Lot Number: 23HMB858; 225) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 21LBP376; 226) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23BBD661; 227) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23BBP453; 228) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23CBG820; 229) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23HBY228; 230) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23JBU455; 231) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23KBN973; 232) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 24BBL537; 233) DYNJ82998B, UDI-DI: 10195327377861(each), 40195327377862(case), Lot Number: 24AMB126; 234) DYNJ83788, UDI-DI: 10195327305833(each), 40195327305834(case), Lot Number: 23CBO212; 235) DYNJ83884B, UDI-DI: 10198459234620(each), 40198459234621(case), Lot Number: 25GBO120; 236) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24EBR736; 237) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24FBL440; 238) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24JBV233; 239) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 25ABF915; 240) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25ABT952; 241) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25CBF253; 242) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25DBT195; 243) DYNJ83950C, UDI-DI: 10198459363948(each), 40198459363949(case), Lot Number: 25HBM610; 244) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 24CBQ297; 245) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25ABJ747; 246) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25BBD646; 247) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25DBG170; 248) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25GBD859; 249) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25HBS096; 250) DYNJ86816, UDI-DI: 10195327679903(each), 40195327679904(case), Lot Number: 24FMB609; 251) DYNJ87444, UDI-DI: 10198459040122(each), 40198459040123(case), Lot Number: 24GBQ988; 252) DYNJ902005I, UDI-DI: 10195327347710(each), 40195327347711(case), Lot Number: 24DLA844; 253) DYNJ902558G, UDI-DI: 10193489728859(each), 40193489728850(case), Lot Number: 23IMC207; 254) DYNJ902558G, UDI-DI: 10193489728859(each), 40193489728850(case), Lot Number: 23HMG989; 255) DYNJ903555D, UDI-DI: 10195327084615(each), 40195327084616(case), Lot Number: 22GBT713; 256) DYNJ903555D, UDI-DI: 10195327084615(each), 40195327084616(case), Lot Number: 22HBF672; 257) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23FBN475; 258) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBG876; 259) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBP518; 260) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBP519; 261) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 23DBS608; 262) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 23HBQ462; 263) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24HBK601; 264) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24IBQ524; 265) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24JBQ211; 266) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24JBU528; 267) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24KBB289; 268) DYNJ907847A, UDI-DI: 10193489937589(each), 40193489937580(case), Lot Number: 23IMF989; 269) DYNJ907874A, UDI-DI: 10195327241247(each), 40195327241248(case), Lot Number: 23FLA249; 270) DYNJ910457A, UDI-DI: 10198459262272(each), 40198459262273(case), Lot Number: 25GMJ928; 271) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24GME383; 272) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24IMF741; 273) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24JMD608; 274) DYNJT3283, UDI-DI: 10198459323645(each), 40198459323646(case), Lot Number: 25JBI499; 275) DYNJT3417, UDI-DI: 10198459324888(each), 40198459324889(case), Lot Number: 25EBN203; 276) DYNJVB91096B, UDI-DI: 10195327257156(each), 40195327257157(case), Lot Number: 24IDA577; 277) PHS806973G, UDI-DI: 10193489926996(each), 40193489926997(case), Lot Number: 23EBQ702
Devices named in this recall
64 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ68844AMedline Industries, Inc. · Plastic surgery and accessories kit4 recalls0 reports
- Medline Industries, Inc. DYNJ44852MMedline Industries, Inc. · Plastic surgery and accessories kit4 recalls0 reports
- Medline Industries, Inc. DYNJ0854650JMedline Industries, Inc. · Plastic surgery and accessories kit3 recalls0 reports
- Medline Industries, Inc. DYKMBNDL117CMedline Industries, Inc. · Plastic surgery and accessories kit3 recalls0 reports
- Medline Industries, Inc. DYNJ910462Medline Industries, Inc. · Plastic surgery and accessories kit3 recalls0 reports
- Medline Industries, Inc. DYNJ0785310GMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ38988JMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ56737AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ50957GMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL170Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL69Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL39Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL84Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL78Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL158Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL183Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL46Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL54Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL182Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ902558GMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. ACC010499Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL34AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. PHS806973GMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ69351Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907847AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ60210AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ903555DMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ44852KMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ64927FMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ55358GMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ39831FMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ903555FMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL170AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL69AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL78AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907874AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL39AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL183AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJVB91096BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ83788Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ903849CMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ902005IMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ36469IMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ86165Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL69BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL78BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL170BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL182AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL183BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ86816Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL182BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ83950AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ87444Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL84CMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL39CMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ83884BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ83950BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJT3283Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJT3417Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ82998BMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ58334DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. CDS984853JMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ910457AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ83950CMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98642).
- Z-2089-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A
- Z-2090-2026 · Open, ClassifiedMedline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP…
- Z-2091-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AA; 2) ENDO KIT W/SYRINGE, Model Number:DYKE1743; 3) BGMC ANGIOGRAPHY PACK-LF, Model Number:DYNJ0800934D;…
- Z-2092-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT…
- Z-2093-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) OPEN HEART CDS, Model Number:CDS840015AQ; 2) OPEN HEART CDS, Model Number:CDS840015AQ; 3) OPEN HEART CDS, Model Number:CDS84…
- Z-2094-2026 · Open, ClassifiedMedline Convenience Kits: 1) PICC LINE TRAY, Model Number:00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number:ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Number:…
- Z-2095-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:DY…
- Z-2096-2026 · Open, ClassifiedMedline Convenience Kits: 1) DYNDA3003, Model Number: DENTAL PACK; 2) DYNJ87157, Model Number: GULLO SURGICAL PACK
- Z-2097-2026 · Open, ClassifiedSee RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK…
- Z-2098-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, …
- Z-2100-2026 · Open, ClassifiedMedline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; 3) LAP BASIN, Model Number: DYNJVB91109D
- Z-2101-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER 6-IV F…
- Z-2102-2026 · Open, ClassifiedMedline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A; 2) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 3) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 4…
- Z-2103-2026 · Open, ClassifiedMedline Convenience Kits: 1) CENTRAL LINE UNIVERSAL INSERTI, Model Number: CVI3780; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) KIT SURG ONC INSERT CENT.VENOU, Model Number:…
- Z-2105-2026 · Open, ClassifiedMedline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A
- Z-2107-2026 · Open, ClassifiedMedline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYK…
- Z-2108-2026 · Open, ClassifiedMedline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL…
- Z-2109-2026 · Open, ClassifiedSee full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, …
- Z-2110-2026 · Open, ClassifiedMedline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Nu…
- Z-2111-2026 · Open, ClassifiedMedline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ037380…
- Z-2112-2026 · Open, ClassifiedMedline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 3) TOT/ GYN LITHOTOMY CDS, Model…
- Z-2113-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104; 2) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104A; 3) KIT OPHTHALMOLOGY CORNEAL TR…
- Z-2114-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS98083…
- Z-2115-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55A; 3) TRACH TOTE, Model Number:DYNJ85691
- Z-2116-2026 · Open, ClassifiedMedline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number:PAIN0150C
- Z-2117-2026 · Open, ClassifiedMedline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721
- Z-2118-2026 · Open, ClassifiedMedline Convenience Kits: 1) PICC CATHETER INSERTION TRAY, Model Number: CVI4300; 2) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 3) KIT ACES INSERT PEG, Model Number:…
- Z-2119-2026 · Open, ClassifiedMedline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L
- Z-2120-2026 · Open, ClassifiedMedline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
- Z-2121-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page