CONTOUR® 7099CRecalls & reported problems
Ascensia Diabetes Care US Inc. · Blood Glucose Test strips. 50 test strips. · Class 2
All models of CONTOUR (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10301937099509), received between Jun 2, 2017 and Jun 2, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K062058); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Hypoglycemia1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 2, 2017Incorrect, Inadequate or Imprecise Result or Readings
EMT STATED THEY WERE CALLED BECAUSE A PERSON WAS FOUND LYING ON THE FLOOR. THEY TESTED THE INDIVIDUAL'S BLOOD WITH THE CONTOUR AND THE READING WAS 64MG/DL. EMT GAVE THE PATIENT 24G OF GLUCOSE GEL. HE WAS NOT SURE IF THE GEL WAS GIVEN BEFORE OR AFTER THE BLOOD TEST. THE PATIENT SHOWED LITTLE RESPONSE TO THE GEL, SO THE EMT CALLED THE FIRE DEPARTMENT. THEY RETESTED THE PATIENT WITH THEIR CONTOUR AND THE READING WAS 28MG/DL. THE PATIENT WAS GIVEN A DEXTROSE50 INJECTION AND THEN WAS FEELING BETTER. SHE WAS GIVEN SOMETHING TO EAT. THE DIFFERENCE BETWEEN READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. THE CUSTOMER RETURNED THE TEST STRIPS FOR INVESTIGATION. NEW METER AND STRIPS WERE SENT TO HIM.
Patient problems: Hypoglycemia
MAUDE report 6608600 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062058. One submission often covers many models, so check the product and lot details in each record.
- Z-3268-2011 · Aug 12, 2011 · TerminatedContour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with diabetes and healthcare professionals to monitor glucose concentrations in w
- Z-1001-2007 · Jun 6, 2007 · TerminatedAscensia Contour Blood Glucose Monitoring System ( Bayer )Diabetes Care System; Product 7152A.---NO USA Units are affected---.
Other Glucose Dehydrogenase, Glucose devices
- OneTouch Verio IQ 022-275LifeScan Europe GmbH · Glucose dehydrogenase, glucose6,035 reports
- OneTouch Verio 022-659LifeScan Europe GmbH · Glucose dehydrogenase, glucose4,219 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE TEST STRIPS - 50 Test StripsTrividia Health, Inc. · Glucose Dehydrogenase, Glucose2 deaths2,548 reports
- FreeStyle Lite 70805Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose2,366 reports
- FreeStyle Freedom Lite 70920Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose1,037 reports
- FreeStyle Freedom Lite 70914Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose760 reports
- Accu-Chek Guide 08256292001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose476 reports
- ACCU-CHEK AVIVA PLUS 06908268001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose472 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.