Henry Schein 5700734Recalls & reported problems
Henry Schein, Inc. · 23G Safety Butterfly Blood Collection Needle · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10304040063634), received between May 16, 2025 and Jun 17, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Defective Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component4
- Infusion or Flow Problem1
Patient problems
- Needle Stick/Puncture3
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2025Infusion or Flow Problem; Infusion or Flow Problem
CUSTOMER REPORTED THAT THE NEEDLE IS SLIDING OFF AND WONT STAY IN PLACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22226603 ↗MalfunctionMay 16, 2025Defective Component; Defective Component
THE CUSTOMER REPORTS THAT THE SAFETY MECHANISM MALFUNCTIONED CAUSING A NEEDLESTICK INJURY. AS OF NOW, NO UDI#. A 62 YEAR OLD FEMALE WAS PERFORMING A HAND DRAW PROCEDURE WHEN THEY HAD RECEIVED A NEEDLESTICK INJURY. IT WAS A DIRTY STICK AS PATIENT HAD BEEN DRAWN. THE SITE AREA IS WASHED AND FOLLOWED UP WITH THE ER. FOLLOW SERIAL BLOOD DRAWS AT 6 WEEKS, 3 MONTHS, 6 MONTHS POST EXPOSURE. LNI FILED. ALL FOLLOW UP THROUGH OCCUPATIONAL HEALTH CLINICS. ONLY ONE REPORTED DEVICE INJURY AT THIS CLINIC FOR THIS DEVICE.
Patient problems: Needle Stick/Puncture
MAUDE report 22035791 ↗MalfunctionMay 16, 2025Defective Component; Defective Component
CUSTOMER THE CUSTOMER REPORTS THAT THE SAFETY MECHANISM MALFUNCTIONED CAUSING A NEEDLESTICK INJURY. AS OF NOW, NO UDI#. A 29 YEAR OLD FEMALE WAS PERFORMING A RIGHT AC PROCEDURE. GLOVES WERE BEING WORN AT THE TIME. IT WAS A DIRTY NEEDLESTICK AS PATIENT HAD BEEN DRAWN. THE SITE OF THE NEEDLESTICK WAS WASHED AND FOLLOWED UP WITH THE ER. FOLLOW SERIAL BLOOD DRAWS AT 6 WEEKS, 3 MONTHS, 6 MONTHS POST EXPOSURE. LNI FILED. ALL FOLLOW UP THROUGH OCCUPATIONAL HEALTH CLINICS. ONLY ONE REPORTED DEVICE INJURY AT THIS CLINIC FOR THIS DEVICE.
Patient problems: Needle Stick/Puncture
MAUDE report 22035758 ↗MalfunctionMay 16, 2025Defective Component; Defective Component
THE CUSTOMER REPORTS THAT THE SAFETY MECHANISM MALFUNCTIONED CAUSING A NEEDLESTICK INJURY. AS OF NOW, NO UDI#. A 37 YEAR OLD FEMALE WAS PERFORMING A RIGHT HAND PROCEDURE WHEN THEY HAD RECEIVED A NEEDLESTICK INJURY. IT WAS A DIRTY STICK. THE SITE AREA WAS WASHED AND FOLLOWED UP WITH THE ER. FOLLOW SERIAL BLOOD DRAWS AT 6 WEEKS, 3 MONTHS, 6 MONTHS POST EXPOSURE. LNI FILED. ALL FOLLOW UP THROUGH (B)(6). ONLY ONE REPORTED DEVICE INJURY AT THIS CLINIC FOR THIS DEVICE. GLOVES WERE BEING WORN AT THE TIME.
Patient problems: Needle Stick/Puncture
MAUDE report 22035754 ↗MalfunctionMay 16, 2025Defective Component; Defective Component
"CUSTOMER REPORTED THE PLASTIC PROTECTIVE CASING AROUND THE NEEDLES EASILY SLIDE OFF WITHOUT HAVING TO TOUCH THEM. THE SAFETY LOCKING MECHANISM IS OFTEN LOOSE AND CLOSES UPON ITSELF WHILE THE NEEDLE IS INSIDE THE PATIENT'S ARM. SOME NEEDLES HAVE HAD MISSING ATTACHMENTS LIKE CONNECTORS. NEEDLE HOLDER ATTACHMENT DETACHED WHILE CHANGING TUBES. IT SEEMS TO BE MORE PAINFUL FOR THE PATIENTS."
Patient problems: Pain
MAUDE report 22035524 ↗
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.