Medex MX9505TRecalls & reported problems
Icu Medical, Inc. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medex pressure transducer appears in 265 adverse event reports, with no deaths and 2 injuries. The reports mostly describe blood or fluid leaking, unreliable pressure readings, and sensors showing as detached. Reports fell to 27 in the last 12 months from 75 in the year before. One open recall, from September 2024, concerns manufacturing defects that can cause inaccurate pressure monitoring or an inability to zero the device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 273
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 273 adverse event reports that list this exact device (DI 10351688503992), received between Sep 16, 2020 and Sep 21, 2026 — classified as 2 injuries, 271 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (56 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K942377); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0192-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak56
- Unable to Obtain Readings42
- Leak/Splash40
- Incorrect, Inadequate or Imprecise Result or Readings27
- Failure to Calibrate19
- Display or Visual Feedback Problem15
- Gas/Air Leak15
- Data Problem13
Patient problems
- No Clinical Signs, Symptoms or Conditions261
- Insufficient Information9
- No Consequences Or Impact To Patient2
- Heartburn/Indigestion1
- Ischemic Heart Disease1
- Valvular Insufficiency/ Regurgitation1
Recent adverse event reports
The 25 most recent of 273 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Fluid/Blood Leak; Reflux within Device
It was reported that blood refluxed into the tubing after connecting the transducer to the radial artery catheter. After replacing the transducer, blood return and leakage stopped. There was a patient involvement and there were no additional adverse consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50845233 ↗MalfunctionSep 18, 2026Fluid/Blood Leak; Reflux within Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829696 ↗MalfunctionSep 18, 2026Fluid/Blood Leak
It was reported that blood was drawn via the right radial artery catheter. During flushing, leakage was observed at the sealed connection between the saline line and sensor. The consumable was replaced. There was a patient involvement and there were no additional adverse consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50818553 ↗MalfunctionSep 1, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50550658 ↗MalfunctionAug 31, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50519688 ↗MalfunctionJul 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074927 ↗MalfunctionJul 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074926 ↗MalfunctionJul 29, 2026Break; Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074903 ↗MalfunctionJul 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074898 ↗MalfunctionMar 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE TRANSDUCER WAS DISPLAYING INTRACRANIAL PRESSURE (ICP) READINGS OF 25-45MMGH. THE PATIENT RECEIVED TREATMENT FOR THE INCREASED ICP READINGS INCLUDING MEDICATIONS AND A HEAD COMPUTED TOMOGRAPHY (CT). THE ELEVATED READINGS WERE NOT CORRELATING WITH THE CLINICAL PRESENTATION OF THE PATIENT. IT WAS UNSURE IF AN ACTUAL HARM WAS DONE TO THE PATIENT. IT WAS ALSO NOTED AFTER THE QUESTIONABLE TRANSDUCER WAS CHANGED OUT THAT THE WIRING HARNESS WHICH RUNS FROM THE PLUG TO THE TRANSDUCER WAS TWISTED. THERE WAS NO PATIENT HARM/ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24535294 ↗MalfunctionMar 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE PRODUCT HAD EXPERIENCED BLOOD BACKFLOW. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24512702 ↗MalfunctionMar 4, 2026Incorrect, Inadequate or Imprecise Result or Readings; Failure to Sense
IT WAS REPORTED THAT AN 83-YEAR-OLD MALE PATIENT BORN ON 20-FEB-1942 HAD A RADIAL ARTERY CATHETER SUCCESSFULLY PLACED. AFTER CONNECTING THE TRANSDUCER, THE MONITOR DISPLAYED 'SENSOR DETACHED' AND DID NOT SHOW THE CONNECTED SENSOR. UPON CHECKING, THE ARTERIAL CATHETER PATHWAY WAS PATENT, THE PRESSURE TRANSDUCER AND PRESSURE CONNECTING LINE WERE SECURELY CONNECTED, AND THE PRESSURE TUBING AND PRESSURE MODULE WERE IN GOOD CONDITION. AFTER REPLACING THE TRANSDUCER, ALL DISPLAYS RETURNED TO NORMAL. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24507917 ↗MalfunctionMar 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT AFTER USE, IT WAS FOUND THAT THE LIQUID INPUT VOLUME FAR EXCEEDED 3 ML/H, REACHING 10 ML/H. AFTER THE DOCTOR REPLACED THE PRESSURE SENSOR, IT RETURNED TO NORMAL. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24506162 ↗MalfunctionFeb 18, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE ANESTHESIOLOGY DEPARTMENT CONNECTED THE DEVICE FOR USE AND THE DOCTOR NOTICED BLOOD LEAKAGE AT THE JOINT. THERE WAS PATIENT INVOLVEMENT. NO PATIENT HARM/ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24380510 ↗MalfunctionFeb 16, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE ANESTHESIOLOGY DEPARTMENT CONNECTED THE DEVICE FOR USE AND THE DOCTOR NOTICED BLOOD LEAKAGE AT THE JOINT. THERE WAS PATIENT INVOLVEMENT. NO PATIENT HARM/ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24345269 ↗MalfunctionFeb 3, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE PRODUCT WAS LEAKING DURING USE. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24246518 ↗MalfunctionJan 19, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS LEAKAGE AT THE CONNECTION BETWEEN THE PRODUCT'S FRONT END AND THE ARTERIAL NEEDLE DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24108519 ↗MalfunctionDec 29, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED A PATIENT UNDERWENT CORONARY ANGIOGRAPHY IN THE CATHETERIZATION LAB, DURING WHICH THE DOCTOR FOUND THAT THE PRESSURE SENSOR MEASUREMENT WAS INACCURATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23916970 ↗MalfunctionDec 23, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT AFTER A DOCTOR REPLACED THE PRESSURE SENSOR, LIQUID WAS FOUND LEAKING AT THE CONNECTION. UPON INSPECTION, A CRACK WAS FOUND AT THE PRESSURE SENSOR JOINT. WHEN PRESSURE WAS APPLIED, LIQUID LEAKED FROM THE CRACKED END OF THE TUBING, AND THE PRESSURE SENSOR WAS IMMEDIATELY REPLACED. NO ADVERSE EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23879357 ↗MalfunctionDec 16, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING USE LEAKAGE WAS FOUND AT THE INTERFACE OF THE PRESSURE SENSOR AND ITS ACCESSORIES AND WAS THEREFORE REPLACED WITH NO FURTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23818817 ↗MalfunctionDec 4, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY UNDER COMBINED SPINAL-EPIDURAL ANESTHESIA DUE TO AVASCULAR NECROSIS OF THE FEMORAL HEAD. A CATHETER WAS INSERTED INTO THE LEFT RADIAL ARTERY FOR PRESSURE MEASUREMENT. BLOOD LEAKAGE WAS DISCOVERED IN THE LEAD-AHEAD TUBING OF THE PRESSURE SENSOR AND ITS ACCESSORIES. UPON INSPECTION, LEAKAGE WAS FOUND AT THE CONNECTOR. THE PRESSURE SENSOR AND ITS ACCESSORIES WERE IMMEDIATELY REPLACED, AND THE PATIENT FUNCTIONED NORMALLY. THE SURGERY WAS SUCCESSFULLY COMPLETED. THE PROCEDURE DID NOT INCREASE THE SURGICAL RISK. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23720743 ↗MalfunctionDec 1, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE REPORT THAT CLINICAL FEEDBACK INDICATED THAT THE MEASUREMENT DATA WAS INACCURATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23682430 ↗MalfunctionNov 26, 2025Pressure Problem; Pressure Problem
IT WAS REPORTED THAT DURING WARD ROUNDS, IT WAS DISCOVERED THAT THERE WAS WATER ACCUMULATION IN THE SUCTION TUBE. AFTER RULING OUT OTHER FACTORS, IT WAS FOUND THAT THE PRESSURE SENSOR AND ITS ACCESSORY VALVE WERE NOT PROPERLY SEALED, CAUSING FLUID TO FLOW INTO THE AIRWAY CONNECTOR. THE PROBLEM WAS DETECTED PROMPTLY; THUS, THE PATIENT SUFFERED NO HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23657091 ↗MalfunctionNov 25, 2025Pressure Problem; Pressure Problem
IT WAS REPORT THAT IN THE INTENSIVE CARE UNIT, THE PRESSURE SENSOR WAS CONNECTED. WHEN FLUSHING THE ARTERIAL BACKFLOW, A LEAK IN THE PRESSURE SENSOR WAS FOUND. CAUSING NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23646553 ↗MalfunctionNov 24, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE PATIENT NEEDED INVASIVE ARTERIAL BLOOD PRESSURE MONITORING DURING SURGERY. BUT THE BLOOD PRESSURE COULD NOT BE MEASURED, AND AFTER REPLACING IT WITH A NEW ONE, IT COULD BE MEASURED NORMALLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638796 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K942377. One submission often covers many models, so check the product and lot details in each record.
- Z-0185-2025 · Sep 24, 2024 · Open, Classifiedmedex TRANSTAR MONITORING KIT 1/EA, Product Code REF MX20103R1
- Z-0186-2025 · Sep 24, 2024 · Open, Classifiedmedex TranStar Kit w/ Stopcock1/EA, Product Code REF MX20740
- Z-0184-2025 · Sep 24, 2024 · Open, Classifiedmedex NICU Monitoring Kit Set 1/EA, Product Code REF MX20086
Other PROBE, BLOOD-FLOW, EXTRAVASCULAR devices
- Medex MX9544Icu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX950X2SCIcu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9548Icu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX20086Icu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9502TIcu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX950SCFTIcu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX20103R1Icu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9504TIcu Medical, Inc. · PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.