Icu Medical, Inc. · 98 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

23 open recalls name at least one Medex model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
774
Deaths reported
0
Injuries reported
57
Recalls
23

We found 98 models of Medex by Icu Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 774 adverse event reports in the FDA's MAUDE database, received between Sep 14, 2016 and Sep 30, 2026 — classified as 57 injuries, 717 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (248 reports).

23 FDA recall records name at least one of these models by its device identifier, 23 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 12 reportsDec 2023: 8 reportsJan 2024: 27 reportsFeb 2024: 10 reportsMar 2024: 10 reportsApr 2024: 34 reportsMay 2024: 19 reportsJun 2024: 19 reportsJul 2024: 5 reportsAug 2024: 9 reportsSep 2024: 15 reportsOct 2024: 14 reportsNov 2024: 23 reportsDec 2024: 35 reportsJan 2025: 13 reportsFeb 2025: 16 reportsMar 2025: 13 reportsApr 2025: 6 reportsMay 2025: 16 reportsJun 2025: 9 reportsJul 2025: 6 reportsAug 2025: 15 reportsSep 2025: 6 reportsOct 2025: 4 reportsNov 2025: 19 reportsDec 2025: 8 reportsJan 2026: 7 reportsFeb 2026: 20 reportsMar 2026: 32 reportsApr 2026: 10 reportsMay 2026: 8 reportsJun 2026: 3 reportsJul 2026: 14 reportsAug 2026: 2 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202635
Last 36 months. The most recent month may be incomplete.

Medex recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0204-2025 · Root cause: Process control · Recall details · FDA record ↗

  2. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0203-2025 · Root cause: Process control · Recall details · FDA record ↗

  3. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0201-2025 · Root cause: Process control · Recall details · FDA record ↗

  4. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0200-2025 · Root cause: Process control · Recall details · FDA record ↗

  5. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0199-2025 · Root cause: Process control · Recall details · FDA record ↗

  6. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0198-2025 · Root cause: Process control · Recall details · FDA record ↗

  7. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0197-2025 · Root cause: Process control · Recall details · FDA record ↗

  8. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0196-2025 · Root cause: Process control · Recall details · FDA record ↗

  9. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0195-2025 · Root cause: Process control · Recall details · FDA record ↗

  10. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0194-2025 · Root cause: Process control · Recall details · FDA record ↗

  11. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0193-2025 · Root cause: Process control · Recall details · FDA record ↗

  12. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0192-2025 · Root cause: Process control · Recall details · FDA record ↗

  13. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0191-2025 · Root cause: Process control · Recall details · FDA record ↗

  14. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0190-2025 · Root cause: Process control · Recall details · FDA record ↗

  15. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0189-2025 · Root cause: Process control · Recall details · FDA record ↗

  16. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0188-2025 · Root cause: Process control · Recall details · FDA record ↗

  17. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0187-2025 · Root cause: Process control · Recall details · FDA record ↗

  18. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0186-2025 · Root cause: Process control · Recall details · FDA record ↗

  19. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0185-2025 · Root cause: Process control · Recall details · FDA record ↗

  20. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0184-2025 · Root cause: Process control · Recall details · FDA record ↗

  21. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0183-2025 · Root cause: Process control · Recall details · FDA record ↗

  22. Sep 24, 2024Open, Classified

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Z-0182-2025 · Root cause: Process control · Recall details · FDA record ↗

  23. Nov 10, 2023Open, Classified

    There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

    Z-0739-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported Medex problems

Device problems

  • Fluid/Blood Leak248
  • Leak/Splash105
  • Disconnection51
  • Break50
  • Unable to Obtain Readings46
  • Gas/Air Leak44
  • Crack43
  • Incorrect, Inadequate or Imprecise Result or Readings32

Patient problems

  • No Clinical Signs, Symptoms or Conditions625
  • Insufficient Information73
  • No Consequences Or Impact To Patient31
  • Low Blood Pressure/ Hypotension23
  • Unspecified Infection10
  • Hemorrhage/Blood Loss/Bleeding7
  • Bradycardia4
  • High Blood Pressure/ Hypertension4

Medex models

98 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Medex adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.