MedexRecalls & reported problems across all models
Icu Medical, Inc. · 98 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
23 open recalls name at least one Medex model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 774
- Deaths reported
- 0
- Injuries reported
- 57
- Recalls
- 23
We found 98 models of Medex by Icu Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 774 adverse event reports in the FDA's MAUDE database, received between Sep 14, 2016 and Sep 30, 2026 — classified as 57 injuries, 717 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (248 reports).
23 FDA recall records name at least one of these models by its device identifier, 23 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Medex recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0204-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0203-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0201-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0200-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0199-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0198-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0197-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0196-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0195-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0194-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0193-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0192-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0191-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0190-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0189-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0188-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0187-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0186-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0185-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0184-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0183-2025 · Root cause: Process control · Recall details · FDA record ↗
- Sep 24, 2024Open, Classified
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0182-2025 · Root cause: Process control · Recall details · FDA record ↗
- Nov 10, 2023Open, Classified
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
Z-0739-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported Medex problems
Device problems
- Fluid/Blood Leak248
- Leak/Splash105
- Disconnection51
- Break50
- Unable to Obtain Readings46
- Gas/Air Leak44
- Crack43
- Incorrect, Inadequate or Imprecise Result or Readings32
Patient problems
- No Clinical Signs, Symptoms or Conditions625
- Insufficient Information73
- No Consequences Or Impact To Patient31
- Low Blood Pressure/ Hypotension23
- Unspecified Infection10
- Hemorrhage/Blood Loss/Bleeding7
- Bradycardia4
- High Blood Pressure/ Hypertension4
Medex models
98 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Medex MX9505TPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall273 reports
- Medex MX9341LStopcock, i.V. Set63 reports
- Medex MX9543Arterial blood sampling kit1 recall4 reports
- Medex MX7827Set, administration, intravascular1 recall1 report
- Medex MX9548PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9547Arterial blood sampling kit1 recall0 reports
- Medex MX9545Arterial blood sampling kit1 recall0 reports
- Medex MX9544PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9530TPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX950X2SCPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX950SCNTPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX950SCFTPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9504TPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9502TPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX9501TPROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX950PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX20897DEVICE, MONITORING, INTRACRANIAL PRESSURE1 recall0 reports
- Medex MX20846R2Filter, infusion line1 recall0 reports
- Medex MX20740TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR1 recall0 reports
- Medex MX20086PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex M20754TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR1 recall0 reports
- Medex MXG300243R1PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MXG300260Arterial blood sampling kit1 recall0 reports
- Medex MX20103R1PROBE, BLOOD-FLOW, EXTRAVASCULAR1 recall0 reports
- Medex MX4705INFUSOR, PRESSURE, FOR I.V. BAGS36 reports
- Medex MX5341LStopcock, i.V. Set30 reports
- Medex FS116Filter, infusion line30 reports
- Medex MX448HFBFilter, infusion line27 reports
- Medex MX43660Tubing, fluid delivery27 reports
- Medex MX4341LStopcock, i.V. Set23 reports
- Medex MX960TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR21 reports
- Medex MX9604ATRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR16 reports
- Medex MX4810INFUSOR, PRESSURE, FOR I.V. BAGS15 reports
- Medex MX43860Tubing, fluid delivery13 reports
- Medex MX5311LStopcock, i.V. Set12 reports
- Medex MX2341LStopcock, i.V. Set11 reports
- Medex MBF27Set, administration, intravascular11 reports
- Medex MX4805INFUSOR, PRESSURE, FOR I.V. BAGS9 reports
- Medex MX448F78MGSet, administration, intravascular8 reports
- Medex MX9605ATRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR6 reports
- Medex MX9342LStopcock, i.V. Set6 reports
- Medex MX9311LStopcock, i.V. Set6 reports
- Medex MX452FLTubing, fluid delivery5 reports
- Medex MX451FLTubing, fluid delivery5 reports
- Medex 537240PSSet, administration, intravascular5 reports
- Medex 536040Set, administration, intravascular5 reports
- Medex MX934Z3LMStopcock, i.V. Set4 reports
- Medex MX560Set, administration, intravascular4 reports
- Medex MX493Set, administration, intravascular4 reports
- Medex MX4343LMStopcock, i.V. Set4 reports
- Medex MX413SLSet, administration, intravascular4 reports
- Medex MX390Syringe, balloon inflation4 reports
- Medex MX1390Syringe, balloon inflation4 reports
- Medex MX458LTubing, fluid delivery3 reports
- Medex MX453Set, administration, intravascular3 reports
- Medex MX450FLTubing, fluid delivery3 reports
- Medex MX4345LMStopcock, i.V. Set3 reports
- Medex MX412SLSet, administration, intravascular3 reports
- Medex MX252LStopcock, i.V. Set3 reports
- Medex MTF37Set, administration, intravascular3 reports
- Medex M20178R2TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR3 reports
- Medex FS113Filter, infusion line3 reports
- Medex 292004Cuff, tracheal tube, inflatable3 reports
- Medex MX934Z4LMStopcock, i.V. Set2 reports
- Medex MX694RSet, administration, intravascular2 reports
- Medex MX693RSet, administration, intravascular2 reports
- Medex MX6812Set, administration, intravascular2 reports
- Medex MX613HSet, administration, intravascular2 reports
- Medex MX5341LNStopcock, i.V. Set2 reports
- Medex MX522LSet, administration, intravascular2 reports
- Medex MX4710INFUSOR, PRESSURE, FOR I.V. BAGS2 reports
- Medex MX4342LMStopcock, i.V. Set2 reports
- Medex MX4315LMStopcock, i.V. Set2 reports
- Medex MX4311LStopcock, i.V. Set2 reports
- Medex MX2311LStopcock, i.V. Set2 reports
- Medex BF100Set, administration, intravascular2 reports
- Medex TF55Set, administration, intravascular1 report
- Medex SM5009Filter, infusion line1 report
- Medex MX960XYTRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR1 report
- Medex MX7961Set, administration, intravascular1 report
- Medex MX7624Set, administration, intravascular1 report
- Medex MX5342LStopcock, i.V. Set1 report
- Medex MX522SLSet, administration, intravascular1 report
- Medex MX5096020MGSet, administration, intravascular1 report
- Medex MX497HPSet, administration, intravascular1 report
- Medex MX494Set, administration, intravascular1 report
- Medex MX453SLSet, administration, intravascular1 report
- Medex MX448L60MGSet, administration, intravascular1 report
- Medex MX448HLMGSet, administration, intravascular1 report
- Medex MX448HL60Tubing, fluid delivery1 report
- Medex MX448HLTubing, fluid delivery1 report
- Medex MX4313FMStopcock, i.V. Set1 report
- Medex MX4312LStopcock, i.V. Set1 report
- Medex MX20119Set, administration, intravascular1 report
- Medex MX20108Set, administration, intravascular1 report
- Medex BTF36Set, administration, intravascular1 report
- Medex BBF26Set, administration, intravascular1 report
- Medex 53035Set, administration, intravascular1 report
Recent Medex adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949831 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937298 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931964 ↗MalfunctionSep 28, 2026Backflow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923282 ↗MalfunctionSep 25, 2026Air/Gas in Device
It was reported that air was observed entering the line of the ULTRA¿ 4-Way Stopcock with Swivel Male Luer Lock, despite the connections being reported as secure. There was patient involvement; however, no patient harm or adverse event was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903500 ↗MalfunctionSep 25, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50902236 ↗MalfunctionSep 24, 2026Break; Fluid/Blood Leak
It was reported that, upon evaluation of the additionally received Medex Stopcocks, cracking was observed along the A-port shafts of the returned device. The reported defect was confirmed. There was patient involvement and moderate patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883606 ↗MalfunctionSep 23, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50863041 ↗InjurySep 22, 2026Separation Problem
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50861391 ↗MalfunctionSep 22, 2026Separation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859868 ↗MalfunctionSep 22, 2026Separation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859789 ↗MalfunctionSep 21, 2026Fluid/Blood Leak; Reflux within Device
It was reported that blood refluxed into the tubing after connecting the transducer to the radial artery catheter. After replacing the transducer, blood return and leakage stopped. There was a patient involvement and there were no additional adverse consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50845233 ↗MalfunctionSep 18, 2026Fluid/Blood Leak; Reflux within Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829696 ↗MalfunctionSep 18, 2026Fluid/Blood Leak
It was reported that blood was drawn via the right radial artery catheter. During flushing, leakage was observed at the sealed connection between the saline line and sensor. The consumable was replaced. There was a patient involvement and there were no additional adverse consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50818553 ↗MalfunctionSep 13, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50747125 ↗
Other Icu Medical, Inc. products
- CADD123 models · 53,012 reports · 21 recalls
- Medfusion16 models · 13,114 reports · 5 recalls
- Bivona469 models · 2,811 reports · 29 recalls
- Level 152 models · 3,733 reports · 1 recall
- Portex409 models · 3,479 reports · 5 recalls
- Cleo6 models · 2,782 reports
- Pneupac23 models · 1,972 reports
- Plum26 models · 1,358 reports · 6 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page