Medex MX4705Recalls & reported problems

Icu Medical, Inc. · Class 1

All models of Medex (98)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This ICU Medical Medex MX4705 pressure infusor squeezes IV bags to help IV fluids flow in faster. The 36 reports mostly describe air or gas leaking from the pressure bag or inflation nozzle, sometimes with the nozzle coming loose, plus cracked or broken parts and pressure that would not build. No deaths and no injuries were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10351688507037), received between Jan 4, 2022 and Dec 1, 2025 — classified as 36 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 6 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 4 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak20
  • Leak/Splash8
  • Crack4
  • Detachment of Device or Device Component4
  • Unintended Deflation3
  • Decrease in Pressure2
  • Fluid/Blood Leak2
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Insufficient Information10

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 1, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE PRESSURE BAG WAS LEAKING, THE DEVICE WAS NEWLY PURCHASED AND STILL UNUSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23680570 ↗
  • MalfunctionNov 12, 2025Leak/Splash; Detachment of Device or Device Component; Gas/Air Leak; Leak/Splash; Detachment of Device or Device Component; Gas/Air Leak

    IT WAS REPORTED THAT THE DEVICE EXHIBITED AIR LEAKAGE WITH THE INFLATION NOZZLE DETACHED. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23536334 ↗
  • MalfunctionNov 12, 2025Leak/Splash; Detachment of Device or Device Component; Gas/Air Leak; Leak/Splash; Detachment of Device or Device Component; Gas/Air Leak

    IT WAS REPORTED THAT THE DEVICE EXHIBITED AIR LEAKAGE WITH THE INFLATION NOZZLE DETACHED. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23536294 ↗
  • MalfunctionNov 12, 2025Leak/Splash; Detachment of Device or Device Component; Gas/Air Leak; Leak/Splash; Detachment of Device or Device Component; Gas/Air Leak

    IT WAS REPORTED THAT THE DEVICE EXHIBITED AIR LEAKAGE WITH THE INFLATION NOZZLE DETACHED. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23536232 ↗
  • MalfunctionAug 11, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE PRESSURE INFUSOR EXHIBITED AIR LEAKAGE AND DEBONDING. THERE WAS NO PATIENT INVOLVEMENT, NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22770296 ↗
  • MalfunctionAug 11, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE PRESSURE INFUSOR EXHIBITED AIR LEAKAGE AND DEBONDING. THERE WAS NO PATIENT INVOLVEMENT, NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22770131 ↗
  • MalfunctionFeb 12, 2025Fracture; Product Quality Problem; Fracture; Product Quality Problem

    IT WAS REPORTED THAT THE CONNECTING PIPE WAS BROKEN. THE EVENT OCCURRED ON 12-JAN-2025. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21368251 ↗
  • MalfunctionOct 7, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE NURSE FOUND THAT THE PRESSURIZED BAG BEGAN TO LEAK AND COULD NOT REACH 300MMHG FOR CLINICAL USE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20391376 ↗
  • MalfunctionSep 27, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE CLINICIANS REPORTED THAT UPON CHECKING, THERE WAS LEAKAGE AND AIR LEAKAGE AT THE GAPS BETWEEN THE EDGES OF THE PRESS. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20322474 ↗
  • MalfunctionSep 27, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT UPON CHECKING, THERE WAS LEAKAGE AND AIR LEAKAGE AT THE GAPS BETWEEN THE EDGES OF THE PRESS. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20322269 ↗
  • MalfunctionSep 27, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT UPON CHECKING, THERE WAS LEAKAGE AND AIR LEAKAGE AT THE GAPS BETWEEN THE EDGES OF THE PRESS. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20322120 ↗
  • MalfunctionJun 18, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THERE WAS LEAKAGE ON PRESSURE INFUSOR AND WAS RESOLVED BY REPLACING WITH A NEW PRODUCT. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS A PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19562936 ↗
  • MalfunctionJun 10, 2024Decrease in Pressure; Gas/Air Leak; Decrease in Pressure; Gas/Air Leak

    IT WAS REPORTED THAT THE THIRD DAY OF THE PATIENT'S CARDIAC SURGERY, THE NURSE FOUND THAT THE THREE-WAY CONNECTION OF THE PRESSURIZED BAG WAS LEAKING AND RESULTED IN INSUFFICIENT PRESSURE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19498423 ↗
  • MalfunctionJun 7, 2024Crack; Gas/Air Leak; Crack; Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS AN AIR LEAK AND COULD NOT BE PRESSURIZED NORMALLY. THE AIR LEAK WAS ALLEGED TO BE IN ONE OF THE CRACKS UNDERNEATH. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19488655 ↗
  • MalfunctionMay 30, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICE WAS DEFECTIVE. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19428708 ↗
  • MalfunctionMay 21, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE SIDE OF THE PRESSURE BAG SPLIT. THE PRODUCT FAULT OCCURRED FOUND WHILE IN USE WITH PATIENT. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS A PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19359117 ↗
  • MalfunctionMay 21, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS LEAKAGE ON PRESSURE THE INFUSOR. THE PRODUCT FAULT OCCURRED FOUND WHILE IN USE WITH PATIENT. THERE WAS NO MEDICAL INTERVENTION REQUIRED AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19358854 ↗
  • MalfunctionMay 1, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE HOSPITAL EXPERIENCED AN AIR LEAK DURING CLINICAL USE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19220261 ↗
  • MalfunctionDec 6, 2023Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS LEAKAGE DURING USE. ADVERSE EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18280020 ↗
  • MalfunctionDec 6, 2023Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS LEAKAGE DURING USE. ADVERSE EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18280019 ↗
  • MalfunctionDec 6, 2023Disconnection; Disconnection

    IT WAS REPORTED THAT DURING USE, "THE BLUE BALLOON CONNECTING TUBE WAS DISCONNECTED DURING USE". ADVERSE EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18280018 ↗
  • MalfunctionDec 6, 2023Crack; Gas/Air Leak; Crack; Gas/Air Leak

    IT WAS REPORTED THAT THE BOTTOM OF THE PRESSURIZED BAG CRACKED AND LEAKED DURING USE. ADVERSE EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18280017 ↗
  • MalfunctionDec 6, 2023Crack; Gas/Air Leak; Crack; Gas/Air Leak

    IT WAS REPORTED THAT THE BOTTOM OF THE PRESSURIZED BAG CRACKED AND LEAKED DURING USE. ADVERSE EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18280016 ↗
  • MalfunctionDec 6, 2023Crack; Gas/Air Leak; Crack; Gas/Air Leak

    IT WAS REPORTED THAT THE BOTTOM OF THE PRESSURIZED BAG CRACKED AND LEAKED DURING USE. ADVERSE EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18280015 ↗
  • MalfunctionJul 11, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE BAG WAS LEAKING, NOT KEEPING FULL. ADVERSE PATIENT EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 17302311 ↗

Other INFUSOR, PRESSURE, FOR I.V. BAGS devices