Medex MX43860Recalls & reported problems
Icu Medical, Inc. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10351688507532), received between Feb 23, 2024 and Mar 3, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Disconnection” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Disconnection7
- Break2
- Detachment of Device or Device Component2
- Leak/Splash2
- Fluid/Blood Leak1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Disconnection
IT WAS REPORTED THAT THE TUBING GOT DISCONNECTED FROM THE PATIENT ARTERIAL LINE WHILE WORKING WITH THE PATIENT. THE EVENT OCCURRED WHILE IN USE WITH PATIENT AT HOSPITAL. IT DID NOT LEAD TO HARM, BUT THE ARTERIAL LINE HAD TO BE REMOVED, AND PATIENT HAD MODERATE AMOUNT OF BLEEDING NOTED FROM THE DISLODGEMENT. THERE WAS AN ONGOING MONITORING OF THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24503009 ↗MalfunctionOct 30, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THEY ARE CONTINUING TO HAVE ISSUES WITH ARTERIAL LINE EXTENSION TUBING DISCONNECTIONS. THEY WERE EASILY ABLE TO DISCONNECT THE TWO PIECES, AND THEY ARE USED ON PRESSURIZED ARTERIAL LINES WHERE PATIENTS CAN LOSE LARGE VOLUMES OF BLOOD IF NOT CAUGHT IN TIME. THERE IS PATIENT INVOLVEMENT AS IT WAS USED ON PATIENTS WITH PRESSURIZED ARTERIAL LINES. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23430070 ↗MalfunctionSep 18, 2025Separation Problem; Separation Problem
IT WAS REPORTED THAT WHILE THE PATIENT WAS INFUSING CRITICAL MEDICATIONS INTO LINE, THE PATIENT TURNED, AND THE LINE DISCONNECTED FROM STOP COCK. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23084770 ↗MalfunctionMay 22, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE PLASTIC TUBING THAT GOES INTO THE STOPCOCK EASILY COMES OFF. REPORTER STATED THAT THERE WERE INSTANCES THAT THEY HAD CHANGED OUT 3 DEVICES AND SOME INSTANCES WHEN THEY THREW THE J-LOOP AWAY. THERE WAS PATIENT INVOLVEMENT BUT IT'S UNKNOWN IF PATIENT WAS INJURED/HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22070641 ↗MalfunctionMay 22, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE PLASTIC TUBING THAT GOES INTO THE STOPCOCK EASILY COMES OFF. REPORTER STATED THAT THERE WERE INSTANCES THAT THEY HAD CHANGED OUT 3 DEVICES AND SOME INSTANCES WHEN THEY THREW THE J-LOOP AWAY. THERE WAS PATIENT INVOLVEMENT BUT IT'S UNKNOWN IF PATIENT WAS INJURED/HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22070631 ↗MalfunctionMay 21, 2025Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED THAT EXTENSION TUBING WAS ATTACHED TO A NEEDLELESS CONNECTOR ON ONE END AND TO A MANIFOLD WITH TWO ONGOING MEDICATIONS, HEPARIN AND PRECEDEX, ON THE OTHER END. THE NURSE TURNED THE PATIENT AND SAW THAT THE EXTENSION TUBING, CONNECTED TO THE STOPCOCK, HAD DISCONNECTED OR "POPPED OFF." A LARGE PUDDLE OF LIQUID WAS FOUND ON THE PATIENT'S BED, AND IT IS UNCLEAR HOW LONG THE PATIENT HAD BEEN WITHOUT MEDICATION. THERE WAS NO MEDICAL INTERVENTION AND PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22066240 ↗MalfunctionMar 6, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THEY HAD MULTIPLE INSTANCES OF THE EXTENSION WITH STOPCOCK GETTING DISCONNECTED AT THE CONNECTION WHERE THE STOPCOCK FUSED WITH THE EXTENSION PIECE. THERE WERE UNKNOWN PATIENT HARMS OR ADVERSE EVENTS REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21536808 ↗MalfunctionDec 27, 2024Disconnection; Disconnection
IT WAS REPORTED THAT THE ARTERIAL LINE DISCONNECTED AT 4-WAY STOPCOCK. THE EVENT OCCURRED DURING USE. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21029175 ↗MalfunctionOct 25, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DEVICE HAD BONDING ISSUES; BONDING COMING OFF. THE PATIENTS' ARTERIAL LINES WERE LEAKING BLOOD. THERE WAS PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20533156 ↗MalfunctionOct 22, 2024Break; Break
IT WAS REPORTED THAT THE EXTENSION TUBING BROKE OFF AT THE STOPCOCK CONNECTION. THERE WAS PATIENT INVOLVEMENT, AND NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20507898 ↗MalfunctionMay 13, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE FOUR-WAY STOPCOCK EXTENSION UTILIZED WITH ARTERIAL LINES WERE SPONTANEOUSLY SEPARATING WHERE THE TUBING MEETS THE STOPCOCK. THERE WAS A PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19297773 ↗MalfunctionApr 17, 2024Break; Disconnection; Break; Disconnection
IT WAS REPORTED THAT THE PRODUCT WAS CONNECTED TO AN ARTERIAL LINE TUBING AND WAS BEING USED ON A PATIENT. THE EXTENSION TUBING BROKE OFF AT THE STOPCOCK CONNECTION. THERE WAS PATIENT INVOLVEMENT. IT IS UNKNOWN IF PATIENT HAD HARM OR AN ADVERSE EVENT OCCURRED.
Patient problems: Insufficient Information
MAUDE report 19126218 ↗MalfunctionFeb 23, 2024Disconnection; Disconnection
IT WAS REPORTED THAT THE STOPCOCK SEPARATED AT THE TUBING CONNECTION DURING USE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18773031 ↗
Other Tubing, fluid delivery devices
- High Pressure Tubing CMerit Medical Systems, Inc. · Tubing, fluid delivery1 recall0 reports
- High Pressure Tubing CMerit Medical Systems, Inc. · Tubing, fluid delivery1 recall0 reports
- Medex MX43660Icu Medical, Inc. · Tubing, fluid delivery27 reports
- Medex MX451FLIcu Medical, Inc. · Tubing, fluid delivery5 reports
- Medex MX452FLIcu Medical, Inc. · Tubing, fluid delivery5 reports
- 10 in. (25 cm) Q2 Checkmate extension set 9545Quest Medical Inc · Tubing, Fluid Delivery3 reports
- Medex MX450FLIcu Medical, Inc. · Tubing, fluid delivery3 reports
- Medex MX458LIcu Medical, Inc. · Tubing, fluid delivery3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page