Skip to content

Bivona 67PFP55Recalls & reported problems

Icu Medical, Inc. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10351688518521), received between Aug 27, 2021 and Mar 21, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2611-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Inflation Problem3

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 21, 2024Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE BALLOON IS NOT BLOWING UP. THE EVENT WAS DISCOVERED PRIOR TO PATIENT USE INSIDE THE PATIENT'S HOME. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18951215 ↗
  • MalfunctionAug 27, 2021Inflation Problem; Inflation Problem

    INFORMATION RECEIVED A SMITHS MEDICAL TRACHEOSTOMY/SILICONE - BIVONA TUBES NEO/PED FLEXTEND PLUS CUFFS ARE STICKING ON ONE SIDE AND NOT INFLATING.

    Patient problems: Insufficient Information

    MAUDE report 12380168 ↗
  • MalfunctionAug 27, 2021Inflation Problem; Inflation Problem

    INFORMATION RECEIVED A SMITHS MEDICAL TRACHEOSTOMY/SILICONE - BIVONA TUBES NEO/PED FLEXTEND PLUS CUFFS ARE STICKING ON ONE SIDE AND NOT INFLATING.

    Patient problems: Insufficient Information

    MAUDE report 12380124 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2007 · Nov 17, 2006 · Terminated
    Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.