Skip to content

Bivona 67PFP60Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10351688518538), received between May 12, 2017 and Feb 11, 2025 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2611-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inflation Problem3
  • Defective Device2
  • Defective Component1
  • Gas/Air Leak1
  • Leak/Splash1
  • Material Deformation1
  • Material Split, Cut or Torn1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Pain1
  • Unspecified Respiratory Problem1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2025Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THERE ARE PROBLEMS WITH THE BALLOON: IT DOES NOT SWELL EVENLY, WHICH CREATES DIFFICULTIES IN VENTILATION. TROUBLESHOOTING WAS DONE BUT THE PROBLEM PERSISTS AND THE CANNULAS CANNOT BE USED. WHEN LOOKING AT THE CANNULA COMPARED TO OTHERS THERE IS A GAP BETWEEN THE STEM AND THE END OF THE BALLOON. THERE WAS PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21358102 ↗
  • MalfunctionFeb 11, 2025Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE DEVICE BALLOON WOULD NOT INFLATE EVENLY. THE USER PERFORMED THE CUFF MANIPULATIONS BUT THE PROBLEM PERSISTS. THE USER TRIED TO USE IT ON HIS SON TWICE BUT HE WAS "DESATURATING". THE CANNULA WAS COMPARED TO OTHERS, AND THERE APPEARED TO BE A GAP BETWEEN THE STEM AND THE END OF THE BALLOON. THE EVENT DATE IS UNKNOWN. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21355139 ↗
  • MalfunctionDec 27, 2022Defective Device; Defective Device

    IT WAS REPORTED THAT THE DEVICE WAS DEFECTIVE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16051805 ↗
  • MalfunctionApr 29, 2022Defective Device; Defective Device

    IT WAS REPORTED THAT TEN DAYS AFTER INSERTION, WHEN THE TUBE WAS REMOVED BY THE PATIENT'S FATHER FOR CLEANING, IT WAS OBSERVED THAT THE TUBE LENGTH APPEARED TO BE LONGER THAN USUAL. THE CUFF DID NOT LAY FLAT AGAINST SHAFT ONCE DEFLATED. NO CONSEQUENCE FOR PATIENT BUT FELT UNCOMFORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14249909 ↗
  • MalfunctionMar 31, 2022Inflation Problem; Gas/Air Leak; Inflation Problem; Gas/Air Leak

    IT WAS REPORTED THAT WHEN CHANGING THE TRACHEOSTOMY CANNULA AND INFLATING THE CUFF IT WAS IMPOSSIBLE TO HAVE A SEAL: LEAKAGE ON VENTILATION AND PAIN FOR THE PATIENT. WHEN REMOVING THE CANNULA, THEY NOTICED THAT THE CUFF REMAINED STUCK TO ONE WALL OF THE CANNULA. THE PATIENT WAS BADLY VENTILATED (FOR 20 MINUTES). THE CANNULA WAS CHANGED TO RESOLVE THE REPORTED EVENT.

    Patient problems: Pain; Unspecified Respiratory Problem

    MAUDE report 13969107 ↗
  • InjuryJan 17, 2019Material Split, Cut or Torn; Material Split, Cut or Torn

    INFORMATION WAS RECEIVED INDICATING THAT A PATIENT WITH A SMITHS MEDICAL PORTEX BIVONA FLEXTEND¿ TTS¿ NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBE WAS REPORTED TO BE EXPERIENCING SECRETIONS AND AIR COMING FROM THE STOMA WHEN THE CUFF IS NOT FULLY INFLATED. SUBSEQUENTLY, THE TRACH TUBE WAS REMOVED AND THE NURSE FILLED THE TRACH WITH WATER. A PINHOLE TEAR IN THE CUFF WAS OBSERVED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS.

    Patient problems: No Code Available

    MAUDE report 8259005 ↗
  • InjuryDec 28, 2018Separation Failure; Separation Failure

    INFORMATION WAS RECEIVED INDICATING THAT THE CONNECTOR CANNOT BE RELEASED FROM A SMITHS MEDICAL PORTEX BIVONA FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE. SUBSEQUENTLY, THE TUBE CHANGED OUT WITH NO REPORTED ADVERSE PATIENT EFFECTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8202396 ↗
  • MalfunctionMay 12, 2017Leak/Splash; Defective Component; Material Deformation; Leak/Splash; Defective Component; Material Deformation

    IT WAS REPORTED THAT THE CUFF OF A PORTEX® BIVONA FLEXTEND¿ TTS¿ PEDIATRIC TRACHEOSTOMY TUBE WAS ASYMMETRICAL, CAUSING A LEAK WHEN USED WITH A PATIENT'S VENTILATOR; MAKING IT DIFFICULT TO VENTILATE THE PATIENT. NO INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6563670 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2007 · Nov 17, 2006 · Terminated
    Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates