1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10351688518576), received between Nov 17, 2023 and Nov 20, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 4 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2612-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Gas/Air Leak4
  • Material Puncture/Hole3
  • Air/Gas in Device1
  • Defective Device1
  • Deflation Problem1
  • Inflation Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2025Gas/Air Leak; Air/Gas in Device; Gas/Air Leak; Air/Gas in Device

    IT WAS REPORTED THAT THE PARENTS OF THE PATIENT INFORMED THE CUSTOMER THAT AN EMERGENCY TRACHEOSTOMY CHANGE HAD TO BE PERFORMED DUE TO THE BALLOON FAILURE ON THE CHILD'S TUBE. THE TUBE HAS ONLY BEEN CHANGED TWICE PREVIOUSLY, ON (B)(6) 2025 AND (B)(6) 2025. THE EVENT OCCURRED AT THE SPACE CENTRE FOR CHILDREN WITH COMPLEX NEEDS. THE PROVIDED PATIENT DETAILS ARE AS FOLLOWS: INITIALS - MSA, AGE - 10 YEARS 7 MONTHS, GENDER - MALE, WEIGHT ¿ 39,4KGS AND ETHNICITY/RACE ¿ BRITISH ASIAN. THERE WAS PATIENT INVOLVEMENT AND NO HARM. A SIMILAR COMPLAINT FROM THE CUSTOMER HAS BEEN DOCUMENTED UNDER CN-364723.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23604797 ↗
  • MalfunctionNov 12, 2025Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak

    IT WAS REPORTED THAT THE CUFF WAS NOT FULLY DEFLATED, ONLY ONE SIDE WAS INFLATED, AND THE PATIENT WAS HARMED DUE TO AIR LEAKAGE AROUND THE TRACHEA. REPORTER STATED THE OUTCOME WAS STILL ONGOING, AND INDICATED DAMAGE WAS OBSERVED ON THE CUFF AND RESULTING IN THE LEAKAGE. ANY ADVERSE EVENTS OF THE PEDIATRIC PATIENT AS A RESULT OF THE AIR LEAKAGE WERE NOT INDICATED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533462 ↗
  • MalfunctionNov 12, 2025Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak

    IT WAS REPORTED THAT THE CUFF WAS NOT FULLY DEFLATED, ONLY ONE SIDE WAS INFLATED, AND THE PATIENT WAS HARMED DUE TO AIR LEAKAGE AROUND THE TRACHEA. REPORTER STATED THE OUTCOME WAS STILL ONGOING, AND INDICATED DAMAGE WAS OBSERVED ON THE CUFF AND RESULTING IN THE LEAKAGE. ANY ADVERSE EVENTS OF THE PEDIATRIC PATIENT AS A RESULT OF THE AIR LEAKAGE WERE NOT INDICATED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533456 ↗
  • MalfunctionNov 12, 2025Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak

    IT WAS REPORTED THAT THE CUFF WAS NOT FULLY DEFLATED, ONLY ONE SIDE WAS INFLATED, AND THE PATIENT WAS HARMED DUE TO AIR LEAKAGE AROUND THE TRACHEA. REPORTER STATED THE OUTCOME WAS STILL ONGOING, AND INDICATED DAMAGE WAS OBSERVED ON THE CUFF AND RESULTING IN THE LEAKAGE. ANY ADVERSE EVENTS OF THE PEDIATRIC PATIENT AS A RESULT OF THE AIR LEAKAGE WERE NOT INDICATED BY THE CUSTOMER.

    Patient problems: Insufficient Information

    MAUDE report 23533368 ↗
  • MalfunctionMar 7, 2025Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem

    IT WAS REPORTED THAT DURING THE ROUTINE 3-HOURLY CUFF RELIEF, WHICH INVOLVES DEFLATING AND REFLATING BALLOON WITH 3.5ML OF WATER PATIENT'S MOTHER ENCOUNTERED DIFFICULTIES WITH THE VALVE. SHE STATED THAT THE VALVE WAS A VERY TIGHT FIT WHEN SHE INSERTED THE SYRINGE, REQUIRING BOTH HER AND CAREGIVERS TO APPLY SIGNIFICANT PRESSURE TO THE TUBE. THE TUBE WAS REPORTED TO BE USED ONCE OVER A 14-DAY PERIOD. THE EVENT HAPPENED AT THE PATIENT'S HOME. THERE WERE UNKNOWN ADVERSE EVENTS REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21546400 ↗
  • MalfunctionNov 17, 2023Defective Device; Insufficient Device Problem Information; Defective Device; Insufficient Device Problem Information

    IT WAS REPORTED THAT IMMEDIATELY ON USE, THE DEVICE WAS FOUND TO HAVE A "DEFECTIVE CUFF". NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18160333 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2007 · Nov 17, 2006 · Terminated
    Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page