Bivona 67PFSS55Recalls & reported problems
Icu Medical, Inc. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10351688518729), received between Apr 10, 2018 and Nov 11, 2025 — classified as 1 injury, 14 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2614-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Inflation Problem5
- Defective Device2
- Fluid/Blood Leak2
- Leak/Splash2
- Retraction Problem2
- Break1
- Crack1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Information2
- Discomfort1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2025Retraction Problem; Retraction Problem
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE CUFF WAS BROKEN. THERE WAS PATIENT INVOLVEMENT, BUT NO HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523458 ↗MalfunctionNov 10, 2025Retraction Problem; Retraction Problem
IT WAS REPORTED THAT THE ISSUE IS THAT THE PILOT PORT WAS RETRACTING INTO THE PILOT BALLOON WHEN INFLATED WITH WATER. DEVICE RETRIEVED FROM THE PATIENT AND SUPPLIED A NEW ONE. THE PATIENT HAS USED IT FOR 4 TIMES. PATIENT IS SAFE, DEVICE REMOVED FROM THE HOUSE. DATIX TO BE DONE AND TERTIARY HOSPITAL INFORMED ABOUT THE PROBLEM. DEFECTIVE DEVICE REMOVED FROM THE PATIENT AND NEW ONE PROVIDED. THE DEFECT WAS FOUND WHEN THE TUBE WAS IN SITU OF THE PATIENT. ONLY 1 ITEM WAS AFFECTED, WHICH WAS USED FOR 4 WEEKS ONLY. IT WAS FOUND BY PARENTS AND REPLACED IT WITH A NEW TUBE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE VENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23511069 ↗MalfunctionNov 20, 2024Inflation Problem; Inflation Problem
IT WAS STATED THAT THE CUFF WAS NOT INFLATING PROPERLY, UNEVENLY. THE PRODUCT FAULT OCCURRED WHILE IN USE WITH A PATIENT. THERE WAS MEDICAL INTERVENTION WHEN THE PATIENT EXPERIENCED DISCOMFORT. THE TRACH WAS REMOVED AND EXCHANGED FOR A DIFFERENT BRAND. OUTCOME OF THE EVENT WAS RESOLVED WITH ALTERNATIVE TRACH.
Patient problems: Discomfort
MAUDE report 20736713 ↗MalfunctionAug 14, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE TRACHEAL CANNULA WAS DEFECTIVE AND COULD NOT BE FULLY INFLATED. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19982992 ↗InjuryJan 19, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE CUSTOMER NOTICED A SPONTANEOUS LEAK EITHER IN THE CUFF OR IN THE BLOCKER LINE AFTER THREE MONTHS OF USE. THE CUSTOMER DID NOT KNOW THE NUMBER OF PREPARATIONS NOR THE TYPE OF PROCESSING. THERE WAS PATIENT INVOLVEMENT. THE PATIENT REQUIRED REPEATED SUCTION DUE TO INCREASED STIMULUS AND LEAKING WATER, BUT BESIDES THAT, NO PATIENT HARM/ADVERSE EVENT WAS REPORTED. THE OPERATOR OF THE DEVICE WAS THE PATIENT, AND THE EVENT TOOK PLACE IN THE PATIENTS¿ HOUSE AND THE PRODUCTS SHOWED DEFICIENCIES IMMEDIATELY UPON USE. THE CUSTOMER SAID THAT DEVICE PREPARATIONS WERE MADE ACCORDING TO THE INSTRUCTIONS. TWO DEVICES WERE INVOLVED REPRESENTING THE SAME PATIENT. THE SECOND REPORTED DEVICE WAS DOCUMENTED ON (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18546141 ↗MalfunctionOct 13, 2022Defective Device; Defective Device
IT WAS REPORTED THAT DURING PRE-TEST THE DEVICE CANNULA WAS FOUND TO BE DEFECTIVE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15595196 ↗MalfunctionOct 12, 2022Defective Device; Defective Device
IT WAS REPORTED THAT DURING PRE-TEST THE DEVICE CANNULA WAS FOUND TO BE DEFECTIVE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15591610 ↗MalfunctionSep 20, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT MULTIPLE TRACHEOSTOMY FAILED TO INFLATE DURING PRE-TEST. MULTIPLE LOTS AFFECTED WITH 11 TRACHEOSTOMY IN TOTAL. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15460021 ↗MalfunctionSep 20, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING PRE-TEST, USING 3ML OF WATER, THE CUFF WAS FOUND TO BE ASYMMETRICAL. WATER WAS TAKEN OUT OF THE CUFF AND REINSERTED WITH THE SAME ISSUE HAPPENING. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15460018 ↗MalfunctionSep 15, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED DURING PRE-TEST, USING 3ML OF WATER, THE CUFF WAS FOUND TO BE ASYMMETRICAL. WATER WAS TAKEN OUT OF THE CUFF AND REINSERTED WITH THE SAME ISSUE HAPPENING. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15433736 ↗MalfunctionSep 15, 2022Break; Tear, Rip or Hole in Device Packaging; Break; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE DEVICE WAS SENT TO CUSTOMER. WHEN IT WAS OPENED, IT WAS FOUND BROKEN. EYELET TEAR. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15427015 ↗MalfunctionSep 13, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING PRE-TEST, USING 3ML OF WATER, THE CUFF WAS FOUND TO BE ASYMMETRICAL. WATER WAS TAKEN OUT OF THE CUFF AND REINSERTED WITH THE SAME ISSUE HAPPENING. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15410871 ↗MalfunctionAug 18, 2019Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA FLEXTEND TTS PEDIATRIC STRAIGHT NECK FLANGE TRACHEOSTOMY TUBE WAS NOTED TO BE LEAKING. THE PATIENT'S MOM WAS PUT 3.5 ML OF WATER IN THE CUFF AND HEARD A LEAK. WHEN SHE CHECKED AGAIN, THERE WAS 2.4 ML OF WATER IN THE CUFF. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 8904575 ↗MalfunctionAug 18, 2019Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA FLEXTEND TTS PEDIATRIC STRAIGHT NECK FLANGE TRACHEOSTOMY TUBE WAS NOTED TO BE LEAKING. THE PATIENT'S MOM WAS PUT 3.5 ML OF WATER IN THE CUFF AND HEARD A LEAK. WHEN SHE CHECKED AGAIN, THERE WAS 2.4 ML OF WATER IN THE CUFF. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 8904572 ↗MalfunctionApr 10, 2018Crack; Crack
IT WAS REPORTED THAT THAT AFTER INSERTING A PORTEX® BIVONA® NEONATAL AND PEDIATRIC SILICONE TRACHEOSTOMY TUBE, THE CUFF CRACKED ON THE TIP OF THE TUBE. NO INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7419536 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.