Skip to content

Bivona 67NFP40Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10351688518767), received between Jan 3, 2022 and Jun 15, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2608-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inflation Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2
  • Cardiac Arrest1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2026Inflation Problem

    IT WAS REPORTED THAT THE CUFF OF THE TRACHEOSTOMY TUBE WAS UNABLE TO BE INFLATED. THE EVENT OCCURRED DURING PATIENT USE. THERE WAS A PATIENT INVOLVED BUT NO PATIENT HARM OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503605 ↗
  • MalfunctionAug 13, 2025Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT A TRACH WAS DISLODGED DURING PATIENT TRANSFER FROM BED TO SCALE. FOLLOWING CODE EVENT, THE NURSE INSPECTED TRACH AND NOTED THAT THE CUFF HAD 1.5ML OF WATER AND THE TRACH DID NOT INFLATE. THE NURSE THEN FILLED IT AT ITS MAXIMUM CAPACITY OF 3ML, THUS TRACH STILL DID NOT INFLATE ALL THE WAY AROUND. AFTER THE PATIENT WAS STABILIZED, THEY TRIED TO USE AN ADDITIONAL PRODUCT TO RESOLVE THE ISSUE BUT TO NO AVAIL. THERE WAS PATIENT INVOLVEMENT AND OUTCOME OF THE INJURY IS STILL ONGOING. CPR(CARDIOPULMONARY RESUSCITATION) AND INCREASED PATIENT CARE NEEDED ARE LISTED AS MEDICAL INTERVENTION.

    Patient problems: Cardiac Arrest

    MAUDE report 22798346 ↗
  • MalfunctionSep 16, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THREE TRACH TUBE CUFFS ARE NOT INFLATING PROPERLY DESPITE CUFF MANIPULATION. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15437943 ↗
  • MalfunctionJan 3, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE CUFFS DID NOT INFLATE WHEN CHECKING PRIOR TO THE PATIENT TRACH CHANGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13142897 ↗
  • MalfunctionJan 3, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE CUFFS DID NOT INFLATE WHEN CHECKING PRIOR TO THE PATIENT TRACH CHANGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13142863 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2007 · Nov 17, 2006 · Terminated
    Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →