Bivona 67NFPS35Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a Bivona tracheostomy tube with a cuff, made by ICU Medical, used to keep an airway open in patients of all ages. The 60 reports mostly describe cuff problems — cuffs that won't inflate, leak air or water, or tear — plus tubes found cracked or broken, including unused ones straight from the box. Nine reports involved injuries; no deaths were reported. One recall, opened in May 2024, says the securing flange on certain lots may tear due to a manufacturing defect; it remains open.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 60
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 60 adverse event reports that list this exact device (DI 10351688518798), received between Jul 18, 2017 and Jan 31, 2025 — classified as 9 injuries, 51 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (31 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2609-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Inflation Problem31
- Leak/Splash7
- Gas/Air Leak6
- Fluid/Blood Leak5
- Device Reprocessing Problem3
- Product Quality Problem3
- Air Leak2
- Defective Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions33
- Insufficient Information11
- No Consequences Or Impact To Patient5
- Hypoventilation4
- No Known Impact Or Consequence To Patient4
- No Information2
- No Patient Involvement2
- Increased Respiratory Rate1
Recent adverse event reports
The 25 most recent of 60 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 31, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT BROUGHT WITH THEM TO THE CLINIC THREE BROKEN TRACH TUBES. CONDITION OF USE WAS STRAIGHT OUT OF THE BOX, UNUSED, WITH BROKEN TRACH CUFFS- SLIT WHERE THE CUFF MEETS THE SHAFT. THE NEW TRACH REMOVED FROM BOX, INFLATED CUFF WITH STERILE WATER WITH IMMEDIATE LEAKAGE OF WATER FROM CUFF. THE EVENT OCCURRED AT PATIENT'S HOME. THERE WERE NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21283989 ↗MalfunctionJan 14, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED THAT ONE TRACH TUBE WOULD ONLY INFLATE HALFWAY, AND A SECOND ONE THE CUFF WOULDN'T INFLATE AT ALL. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21147539 ↗MalfunctionNov 14, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF WOULD NOT INFLATE. PILOT BALLOON WAS FILLING, BUT NO INFLATION ON CUFF SEEN AT ALL. ADDITIONAL 2 ML WERE INSERTED, PILOT BALLOON OVERFILLED, BUT CUFF WAS STILL NOT INFLATING. THERE WERE NO PATIENT INVOLVEMENT OR HUMAN HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20689244 ↗MalfunctionOct 18, 2024Material Separation; Material Separation
IT WAS REPORTED THAT CUFFED TRACHEOSTOMY WAS IN SITU IN A CHILD, EXTERNAL CUFF NOTED TO BE MISSING FROM END OF THE EXTENSION, PREVIOUSLY PRESENT. BREAKAGE OF TUBE, POTENTIAL RESULTING IN AN UNSAFE AIRWAY. TUBE IMMEDIATELY CHANGED. THERE WAS PATIENT INVOLVEMENT. IT WAS STATED THAT THIS WAS AN ISOLATED INCIDENT THAT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20481595 ↗MalfunctionJul 8, 2024Defective Component; Defective Component
IT WAS REPORTED THAT THE DEVICE HAD A DEFECTIVE CUFF. THERE WAS A PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19696105 ↗MalfunctionJun 7, 2024Contamination; Contamination
IT WAS REPORTED THAT MULTIPLE TRACHEOSTOMY TUBES ONCE PROCESSED BY CSR ARE BEING RETURNED TO UNIT WITH "FUZZY" SUBSTANCE ON OBTURATOR. THERE IS NO PATIENT HARM. NO MEDICAL INTERVENTION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19486058 ↗InjuryMay 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT BABY WITH TRACH AND VENT WAS AGITATED, FLAILING ARMS. THE TRACH TUBE HUB DISCONNECTED FROM THE BODY OF THE TRACH TUBE. THEY BELIEVED THAT IT IS A SWIVEL ADAPTER. AN EMERGENCY TRACH TUBE CHANGE WAS PERFORMED, AND THE OUTCOME WAS RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19359666 ↗MalfunctionApr 23, 2024Defective Device; Insufficient Device Problem Information; Defective Device; Insufficient Device Problem Information
IT WAS REPORTED THAT THE BALLOON WAS FOUND DEFECTIVE ON PREVIOUSLY STERILIZED TRACH. THE NATURE OF THE "DEFECT" WAS NOT PROVIDED. THE INCIDENT HAPPENED UPON TRACH CHANGED AND DID NOT REACH PATIENT. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19163466 ↗MalfunctionNov 22, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT UPON INSPECTION OF NEW TRACH, THE BALLOON FAILED TO INFLATE PRIOR TO INSERTION. TRACH WAS NOT USED ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18195019 ↗MalfunctionMar 27, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE DEVICE HAD AN INVERTED BELL-SHAPED CUFF THAT WOULD NOT INFLATE WELL TOWARDS THE BOTTOM, WITH A MORE ROUNDED TOP. THE REPORTED LOT NUMBER COULD NOT BE VERIFIED, THERE IS A MISSING DIGIT. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16621864 ↗MalfunctionMar 27, 2023Inflation Problem; Gas/Air Leak; Inflation Problem; Gas/Air Leak
IT WAS REPORTED THAT AFTER THE DEVICE WAS REPROCESSED IT HAD A LOPSIDED CUFF; LOOKED "LIKE A BULGE IN A TIRE". THIS CREATED ISSUES WITH THE SEAL AGAINST THE TRACHEAL WALL; CUSTOMER CANNOT EASILY VENTILATE WITH A LEAK. IT IS UNKNOWN IF THIS WAS HOW THE TRACH WAS BEFORE REPROCESSING. THE DEVICE HAD BEEN REPROCESSED LESS THAN 5 TIMES. THE REPORTED LOT NUMBER COULD NOT BE VERIFIED, THERE IS A MISSING DIGIT. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16621645 ↗MalfunctionMar 27, 2023Inflation Problem; Gas/Air Leak; Inflation Problem; Gas/Air Leak
IT WAS REPORTED THAT AFTER THE DEVICE WAS REPROCESSED IT HAD A LOPSIDED CUFF; LOOKED "LIKE A BULGE IN A TIRE". THIS CREATED ISSUES WITH THE SEAL AGAINST THE TRACHEAL WALL; CUSTOMER CANNOT EASILY VENTILATE WITH A LEAK. IT IS UNKNOWN IF THIS WAS HOW THE TRACH WAS BEFORE REPROCESSING. THE DEVICE HAD BEEN REPROCESSED LESS THAN 5 TIMES. THE REPORTED LOT NUMBER COULD NOT BE VERIFIED, THERE IS A MISSING DIGIT. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16621588 ↗MalfunctionMar 27, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE DEVICE CUFF WOULD NOT INFLATE AT ALL PRIOR TO USE. THE DEVICE HAD NOT BEEN REPROCESSED. UNFORTUNATELY, THAT PRODUCT WAS THROWN AWAY BY THE PROVIDER. THE REPORTED LOT NUMBER COULD NOT BE VERIFIED, THERE IS A MISSING DIGIT. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16621227 ↗MalfunctionMar 9, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF OF THE TRACH TUBE WOULD NOT INFLATE. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16517981 ↗MalfunctionMar 9, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF OF THE TRACH TUBE WOULD NOT INFLATE. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16513156 ↗MalfunctionFeb 21, 2023Device Reprocessing Problem; Gas/Air Leak; Device Reprocessing Problem; Gas/Air Leak
IT WAS REPORTED THAT THE CUFF WAS USED AND REPROCESSED BUT NOW HAS A LOPSIDED APPEARANCE WITH A BULGE, WHICH AFFECTS THE SEAL AGAINST THE TRACHEAL WALL AND MAKES VENTILATION DIFFICULT. THERE HAS BEEN NO REPORT OF PATIENT INJURY, NO OBSERVABLE CLINICAL SYMPTOMS, OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16414604 ↗MalfunctionFeb 21, 2023Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT THE TRACH TUBE AFTER RE-PROCESSING HAS AN INVERTED BELL-SHAPED CUFF WITH POOR INFLATION TOWARDS THE BOTTOM AND A MORE ROUNDED TOP. THERE HAS BEEN NO REPORT OF PATIENT INJURY, NO OBSERVABLE CLINICAL SYMPTOMS, OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16414603 ↗MalfunctionFeb 21, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF WOULD NOT INFLATE BEFORE USE (DISCARDED BY THE PROVIDER). THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16414599 ↗MalfunctionFeb 21, 2023Device Reprocessing Problem; Gas/Air Leak; Device Reprocessing Problem; Gas/Air Leak
IT WAS REPORTED THAT THE CUFF WAS USED AND REPROCESSED, BUT NOW HAS A LOPSIDED APPEARANCE WITH A BULGE, WHICH AFFECTS THE SEAL AGAINST THE TRACHEAL WALL AND MAKES VENTILATION DIFFICULT. THERE HAS BEEN NO REPORT OF PATIENT INJURY, NO OBSERVABLE CLINICAL SYMPTOMS, OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16414597 ↗MalfunctionFeb 7, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF WAS NOT INFLATING FULLY DURING PRE-TEST. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16316461 ↗InjuryNov 14, 2022Inflation Problem; Gas/Air Leak; Inflation Problem; Gas/Air Leak
IT WAS REPORTED THAT THE PATIENT WAS DUE FOR WEEKLY TRACH CHANGE. PATIENT HAS A 3.5 TRACH TUBE CUFFED WITH 2ML OF STERILE WATER. AN HOUR AFTER TRACH CHANGE, PATIENT WAS EVENTING WITH INCREASED C02S, RESTRICTED VOLUMES ON VENT AND GIANT AIR LEAK WITH HEAD BOBBING AND RETRACTIONS. RESPIRATORY THERAPIST (RT) AT BEDSIDE WITH REGISTERED NURSE (RN) TROUBLESHOOTING CAUSE OF EVENTS AND INCREASING C02. CHANGED TRACH OUT EMERGENTLY WITH RT; PREVIOUS TRACH CUFF WAS FOUND TO NOT INFLATE. (UNKNOWN LOT NUMBER).
Patient problems: Hypoventilation; Respiratory Distress Syndrome of Newborns
MAUDE report 15793632 ↗MalfunctionOct 19, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING PRE-TEST, THE CUFFS FAILED TO INFLATE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15635773 ↗MalfunctionOct 13, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING A HOSPITAL PRE-TEST OF THE TRACH TUBE, THE CUFF WOULD NOT INFLATE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15595620 ↗MalfunctionSep 24, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING PRE-TEST, PATIENT MOTHER WAS TRYING TO INFLATE THE CUFF AND IT DID NOT WORK DESPITE TRYING 4-5 TIMES. SHE THEN TRIED AGAIN AND HEARD A POP AND THE CUFF THEN INFLATED. MOTHER ATTEMPTED TO FILL WITH WATER BUT CUFF LEAKED. SHE OPENED TWO OTHERS AND THE CUFF WOULD NOT INFLATED ON THOSE EITHER. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15476009 ↗MalfunctionSep 22, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING PRE-TEST, PATIENT MOTHER WAS TRYING TO INFLATE THE CUFF AND IT DID NOT WORK DESPITE TRYING 4-5 TIMES. SHE THEN TRIED AGAIN AND HEARD A POP AND THE CUFF THEN INFLATED. MOTHER ATTEMPTED TO FILL WITH WATER BUT CUFF LEAKED. SHE OPENED TWO OTHERS AND THE CUFF WOULD NOT INFLATED ON THOSE EITHER. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15467644 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page