Skip to content

DuraGen® Plus DP1045Recalls & reported problems

Integra Lifesciences Corporation · DuraGen® Plus Dural Regeneration Matrix , 4 in x 5 in · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10381780023470), received between Mar 13, 2024 and Dec 27, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).

1 other recall cover devices cleared under the same FDA submission (K032693); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Split, Cut or Torn1

Patient problems

  • Meningitis1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPORTED THE FOLLOWING INCIDENT DESCRIPTION: "48-YEAR-OLD FEMALE PATIENT ADMITTED TO THE HOSPITAL, IN THE INTERNAL MEDICINE DEPARTMENT FOR MENINGITIS WITH 1280 ELEMENTS/MM3 PREDOMINANTLY LYMPHOCYTIC. 2 MONTHS AFTER A CRANIECTOMY FOR A LEFT FRONTO-TEMPORAL CRANIOMENINGIOMA, TREATED WITH KEPPRA AS PRIMARY PROPHYLAXIS. AFTER REVIEW OF THE LITERATURE, THE POSSIBILITY OF KEPPRA-INDUCED MENINGITIS (IMMUNO-ALLERGIC MECHANISM) PROMPTED THE DISCONTINUATION OF KEPPRA AFTER NEUROLOGICAL ADVICE, BUT THIS HYPOTHESIS SEEMS IMPLAUSIBLE AS THERE IS A PERSISTENT PERSISTENCE OF A LYMPHOCYTIC REACTION DESPITE DISCONTINUATION. BACTERIOLOGICAL AND MYCOLOGICAL ANALYSES OF LUMBAR PUNCTURES DO NOT SUGGEST AN INFECTIOUS ORIGIN. THE MAIN HYPOTHESIS IS THEREFORE THAT OF A POSTOPERATIVE CHEMICAL MENINGITIS. GIVEN THE NEGATIVITY OF THE INFECTIOUS WORKUP AND THE ABSENCE OF A BIOPSIABLE SITE ON THE CT SCAN, CORTICOSTEROID THERAPY WITH TAPERING WAS INITIATED, LEADING TO MARKED IMPROVEMENT AND COMPLETE WITHDRAWAL SYMPTOMATIC TREATMENTS. PATIENT INFORMATION: PATIENT'S CURRENT CONDITION: CORTICOSTEROID THERAPY WITH TAPERING IS INITIATED, LEADING TO MARKED IMPROVEMENT AND COMPLETE WITHDRAWAL OF SYMPTOMATIC

    Patient problems: Meningitis

    MAUDE report 21031016 ↗
  • MalfunctionMar 13, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A FACILITY REPORTED THAT DURAGEN PLUS DURAL REGENERATION MATRIX (DP1045) WAS IMPLANTED IN A PATIENT DIAGNOSED WITH MENINGIOMA BRAIN TUMOR. AT THE TIME OF PASSING THE SUTURES INTO THE PATCH, IT BEGAN TO TEAR AND EVEN CAME LOOSE FROM THE TRACTION POINTS. INTRAOPERATIVELY, THE PATIENT DID NOT HAVE ANY DAMAGE/INJURY FROM THE EVENT THAT OCCURRED, AND IT IS HOPED THAT IT WILL NOT CAUSE A FISTULA LATER. THE PATIENT WAS REPORTED TO BE IN SATISFACTORY RECOVERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18897384 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032693. One submission often covers many models, so check the product and lot details in each record.

  • Z-1284-2013 · Mar 28, 2013 · Terminated
    DuraGen Plus Dural Regeneration Matrix DuraGen Plus Adhesion Barrier Matrix is an absorbable implant for repair and restoration of dural defects and as an adhesion barrier for the

Other DURA SUBSTITUTE devices

All DURA SUBSTITUTE devices →