DuraGen® Suturable DURS2291Recalls & reported problems
Integra Lifesciences Corporation · DuraGen® Suturable Dural Regeneration Matrix, 2 in x 2 in (5 cm x 5 cm) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10381780023500), received between Nov 9, 2023 and Jul 2, 2025 — classified as 5 injuries.
The most frequently reported specific device problem is “Device Handling Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K043427); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device Handling Problem1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Material Puncture/Hole1
Patient problems
- Cerebrospinal Fluid Leakage3
- Bacterial Infection1
- Fistula1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT IS 2 OF 2 AND LINKED TO MFG REPORT NUMBER: 3003418325-2025-00032: A FACILITY REPORTED THAT A PATIENT DIAGNOSED WITH SUBDURAL HEMATOMA UNDERWENT SURGERY ON (B)(6) 2025. HOWEVER, IN THE FOLLOWING DAYS, HE DEVELOPED A FISTULA THAT REQUIRED REOPERATION. THE REOPERATION WAS PERFORMED TO TREAT THE AFOREMENTIONED COMPLICATION. THE SAME PRODUCT REFERENCES WERE USED IN THE SECOND SURGERY, AND THE PATIENT IS IN STABLE CONDITION.
Patient problems: Fistula
MAUDE report 22385704 ↗InjuryJan 31, 2025Device Handling Problem; Device Handling Problem
InjuryApr 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A DISTRIBUTOR INITIALLY REPORTED THE FOLLOWING: "THE PATIENT WAS INITIALLY OPERATED ON WITH A SURGICAL TECHNIQUE FOR CHIARI MALFORMATION IN WHICH THE SPECIALIST USED A CLASSIC PATCH THAT WAS IN THE INSTITUTION'S STOCK AND SUTURED IT. GIVEN THE PRESENCE OF A FISTULA, THE PATIENT MUST BE RE-OPERATED URGENTLY. FOR THIS CASE, A SUTURABLE PATCH AND A DURASEAL ARE SENT. IN THIS SECOND POSTOPERATIVE PERIOD, THE CSF LEAK PERSISTS AND A THIRD INTERVENTION IS CONTEMPLATED, WHICH IS NOT CARRIED OUT SINCE THE LEAK STOPS. IT IS NECESSARY TO CLARIFY THE INDICATIONS FOR USE OF THE DIFFERENT DURAGEN TO AVOID FUTURE ADVERSE EVENTS.". ADDITIONAL INFORMATION WAS SUBSEQUENTLY RECEIVED INDICATING THE FOLLOWING: 1. CAN YOU VERIFY THE TYPE OF PROCEDURE PERFORMED DURING THE INCIDENT? - IT WAS A RESECTION OF A CEREBELLAR TUMOR. 2. WHAT MEDICAL CHECKUP/INTERVENTION WAS PERFORMED? IF YOU HAD TO DO A REVIEW, WHAT DID YOU DO? -YES. THE DUROPLASTY WAS REVISED, A SUTURABLE PATCH AND SEALANT WERE PLACED. 3. HOW MUCH CSF WAS LEAKED DURING THE INCIDENT (IF AN EXACT AMOUNT CANNOT BE PROVIDED, PLEASE PROVIDE AN APPROXIMATE AMOUNT: SMALL, MODERATE, ETC.)? -MODERATE. 4. WAS THE PATIENT TRANSFERRED OR BEING TRANSFERRED
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 19089670 ↗InjuryDec 7, 2023Material Puncture/Hole; Material Puncture/Hole
InjuryNov 9, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
Recalls under the same FDA submission
These recalls involve devices cleared under K043427. One submission often covers many models, so check the product and lot details in each record.
- Z-1283-2013 · Mar 28, 2013 · TerminatedSuturable DuraGen Dural Regenerative Matrix Suturable DuraGen Dural Regenerative Matrix is an absorbable implant for repair of dural defects. Suturable DuraGen is an easy to handl
- Z-1081-2010 · Jan 13, 2010 · TerminatedSuturable DuraGen Dural Regeneration Matrix; Non-Pyrogenic; Sterility guaranteed unless package is damaged or opened. Rx only. Manufacturer: Integra LifeSciences Corporation, Plain
Other DURA SUBSTITUTE devices
- Durepair® 62100Medtronic Ps Medical, Inc. · DURA SUBSTITUTE1 recall6 reports
- Durepair® 62105Medtronic Ps Medical, Inc. · DURA SUBSTITUTE1 recall5 reports
- Durepair® 62110Medtronic Ps Medical, Inc. · DURA SUBSTITUTE1 recall3 reports
- DuraMatrix Onlay CDSLM45Collagen Matrix, Inc. · Dura Substitute1 recall3 reports
- Durepair® 62111Medtronic Ps Medical, Inc. · DURA SUBSTITUTE1 recall1 report
- Durepair® 62106Medtronic Ps Medical, Inc. · DURA SUBSTITUTE1 recall1 report
- Biodesign G31091Cook Biotech Incorporated · DURA SUBSTITUTE1 recall0 reports
- Biodesign G31091Cook Biotech Incorporated · DURA SUBSTITUTE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page