Skip to content

Sundt™ NL8505070Recalls & reported problems

Integra Lifesciences Corporation · Sundt™ External Carotid Endarterectomy Shunt · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780026259), received between Mar 27, 2025 and Mar 27, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2025Use of Device Problem; Use of Device Problem

    A FACILITY REPORTED THAT WHEN PLACING THE SUNDT EXTERNAL SHUNT (NL8505070), THE SURGEON OBSERVED AN ISSUE AND HAD TO CHANGE THE SHUNT QUICKLY. THE CAROTID ARTERY WAS CLAMPED AT THE TIME OF THE OCCURRENCE. THERE WAS NO ALLEGATION OF PATIENT INJURY. ADDITIONAL INFORMATION RECEIVED INDICATING THE FOLLOWING: DO YOU HAVE MORE DETAILS ON THE ISSUE FACED? SHUNT WAS FLATTENED. - DID THE ISSUE CAUSE ANY INCREASE OF SURGICAL TIME EQUAL OR SUPERIOR TO 30 MINUTES)? NO INCREASED TIME REPORTED AS DEVICES CHANGED AT THE BEGINNING OF THE PROCEDURE. HOW DID THEY FINALIZE THE SURGERY? USE OF ANOTHER DEVICE AVAILABLE IN STOCK? >> YES - "AS PER RISK ASSESSMENT PERFORMED BY OUR CLINICIAN TEAM, THIS CASE IS CONSIDERED AS REPORTABLE. POTENTIAL FOR DEATH OR SERIOUS INJURY IS NOT REMOTE. FLATTENED CAROTID SHUNT MAY IMPEDE OR REDUCE BLOOD FLOW TO THE HEAD AND BRAIN. THEREFORE, POTENTIAL RISK FOR SERIOUS INJURY IS NOT REMOTE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21708976 ↗

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page