Skip to content

Integra® 1362275Recalls & reported problems

Integra Lifesciences Production Corporation · Integra® Universal Flexible Arm · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780074670), received between Mar 11, 2022 and Mar 11, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Feb 28, 2022Open, Classified

    Excess force used to tighten the Universal Flexible Arm can cause the internal cable to break at the tip, resulting in an unusable condition.

    Z-0906-2022 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2022Break; Break

    THIS 3 OF 3 REPORTS LINKED TO MFG REPORT NUMBERS: 3014334038-2022-00049, 3014334038-2022-00048. A FACILITY REPORTED THAT THE UNIVERSAL FLEXIBLE ARM (1362275) BROKE IN THE MIDDLE OF A TYMPANOPLASTY PROCEDURE. ALL PARTS WERE RECOVERED AND WERE TAKEN OFF THE SURGICAL FIELD; NO FRAGMENTS FELL INTO THE SURGICAL SITE. IT WAS REPORTED THAT AN X-RAY WAS PERFORMED TO ASSURE THAT NO PARTS WERE LEFT IN THE PATIENT FOLLOWING THE EVENT. THE EVENT DID NOT LEAD TO PATIENT INJURY, DEATH, OR DELAY IN PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13737749 ↗

Other SPECULA ACCESSORIES devices

All SPECULA ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page