Skip to content

Integra Universal Flexible Arm part number REF 1362275 recallZ-0906-2022 · Integra LifeSciences Corp. · initiated Feb 28, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Excess force used to tighten the Universal Flexible Arm can cause the internal cable to break at the tip, resulting in an unusable condition.

FDA root cause category: Device Design

What the company is doing

Urgent Voluntary Medical Device Recall Notice dated 02/28/2022 was sent via FexEx. Customers are instructed to remove the affected product from service, return it to Integra and complete and submit the attached acknowledgment form. Contact Customer Service with questions: Phone: 1-800-654-2873, Email: [email protected]

Recall details

Recall number
Z-0906-2022
Status
Open, Classified
Initiated by the company
Feb 28, 2022
Recalling firm
Integra LifeSciences Corp.
Quantity
54
Product code
FXG
Distribution
US Nationwide Distribution: AL, AZ, FL, GA, IL, MA, MD, NY, OH, OR, PA, TX

Product description

Integra Universal Flexible Arm part number REF 1362275

Code information (models, lots, serial numbers)

UDI#10381780074670, Lots: AC2109, AC2103, AC2106, AC2110, AC2101, AC2008, AC2009

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page