Skip to content

MAYFIELD® A1012Recalls & reported problems

Integra Lifesciences Corporation · MAYFIELD® SWIVEL HORSESHOE HEADREST · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10381780267461), received between May 28, 2021 and Jul 31, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Detachment of Device or Device Component1
  • Device Slipped1
  • Mechanical Problem1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Neck Stiffness1
  • Pressure Sore/Ulcer1
  • Spinal Cord Injury1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22680264 ↗
  • InjuryJul 21, 2025Unstable; Unstable

    THIS IS 2 OF 4 REPORTS FOR THE MAYFIELD SYSTEM LINKED TO MFG REPORT NUMBERS: 3004608878-2025-00148, 3004608878-2025-00150, 3004608878-2025-00151. A FACILITY REPORTED THAT THE MAYFIELD SWIVEL HORSESHOE HEADREST (A1012) NEEDS TO BE CHECKED AS IT GOT LOOSE/BECAME UNLOCKED DURING SURGERY, RESULTING IN WHIPLASH FOR THE PATIENT. IT IS UNCLEAR IF THE EVENT LED TO INCREASED SURGERY TIME. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Neck Stiffness

    MAUDE report 22555441 ↗
  • InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS 1 OF 4 REPORTS LINKED TO MFG REPORT NUMBERS: 3004608878-2024-00107 3004608878-2024-00108 3004608878-2024-00109. A FACILITY REPORTED THAT FOUR INCIDENCES OF PRESSURE INJURY OCCURRED AFTER USING THE MAYFIELD SWIVEL HORSESHOE HEADREST (A1012) AND THE MAYFIELD PEDIATRIC HORSESHOE HEADREST (A1051). IT IS UNKNOWN IF THERE WAS A DELAY IN SURGERY; HOWEVER, ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONALLY, THE FACILITY HAS INDICATED THERE WAS NO MALFUNCTION OF THE DEVICE AND IS PERFORMING A MULTIFACTORIAL ROOT CAUSE ANALYSIS.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 20153886 ↗
  • InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS 3 OF 4 REPORTS LINKED TO MFG REPORT NUMBERS: 3004608878-2024-00106 3004608878-2024-00107 3004608878-2024-00109. A FACILITY REPORTED THAT FOUR INCIDENCES OF PRESSURE INJURY OCCURRED AFTER USING THE MAYFIELD SWIVEL HORSESHOE HEADREST (A1012) AND THE MAYFIELD PEDIATRIC HORSESHOE HEADREST (A1051). IT IS UNKNOWN IF THERE WAS A DELAY IN SURGERY; HOWEVER, ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONALLY, THE FACILITY HAS INDICATED THERE WAS NO MALFUNCTION OF THE DEVICE AND IS PERFORMING A MULTIFACTORIAL ROOT CAUSE ANALYSIS.

    Patient problems: Spinal Cord Injury

    MAUDE report 20153885 ↗
  • MalfunctionSep 7, 2023Device Slipped; Device Slipped

    A FACILITY REPORTED THAT THE MAYFIELD SWIVEL HORSESHOE HEADREST (A1012) WAS ¿SLIPPING¿. NO PATIENT INJURY OR SURGICAL DELAY HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17705223 ↗
  • MalfunctionMay 28, 2021Mechanical Problem; Mechanical Problem

    A FACILITY REPORTED THAT PRIOR TO POSTERIOR CERVICAL DISCECTOMY, IT WAS DISCOVERED THAT THE MAYFIELD SWIVEL HORSESHOE HEADREST (A1012) HAD BAD THREADS AND WOULD NOT CONNECT TO THE MAYFIELD SWIVEL UNIT. THERE WAS A DELAY FOR ABOUT 30 MINUTES WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11900967 ↗

Other HEADREST, NEUROSURGICAL devices

All HEADREST, NEUROSURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page