The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10381780267553), received between Feb 4, 2022 and Sep 5, 2024 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Device Dislodged or Dislocated1
- Device Slipped1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Pressure Sore/Ulcer1
- Spinal Cord Injury1
- Ulcer1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS IS 4 OF 4 REPORTS LINKED TO MFG REPORT NUMBERS: 3004608878-2024-00106 3004608878-2024-00107 3004608878-2024-00108. A FACILITY REPORTED THAT FOUR INCIDENCES OF PRESSURE INJURY OCCURRED AFTER USING THE MAYFIELD SWIVEL HORSESHOE HEADREST (A1012) AND THE MAYFIELD PEDIATRIC HORSESHOE HEADREST (A1051). IT IS UNKNOWN IF THERE WAS A DELAY IN SURGERY; HOWEVER, ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONALLY, THE FACILITY HAS INDICATED THERE WAS NO MALFUNCTION OF THE DEVICE AND IS PERFORMING A MULTIFACTORIAL ROOT CAUSE ANALYSIS.
Patient problems: Spinal Cord Injury
MAUDE report 20153887 ↗MalfunctionJan 2, 2024Device Slipped; Device Slipped
THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3004608878-2023-00241: A FACILITY REPORTED THAT THE MAYFIELD MODIFIED SKULL CLAMP (A1059) WAS BEING USED FOR CRANIAL STABILIZATION PROCEDURE, AND A SLIP OCCURRED. NO PATIENT INJURY OR SURGICAL DELAY HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18428697 ↗MalfunctionDec 22, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
A FACILITY REPORTED THAT THE LEFT AND RIGHT SIDES OF THE MAYFIELD PEDIATRIC HORSESHOE HEADREST (A1051), DID NOT STAY TOGETHER EVEN WITH SCREW TIGHTENED DOWN AS HARD AS POSSIBLE. DURING A CASE, THE HEADREST FAILED TO STAY LOCKED, CAUSING THE BABY'S HEAD TO SLIP AS THE TWO SIDES SEPARATED. FORTUNATELY, THERE WAS NO INJURY, BUT THERE WAS A 5-MINUTE DELAY AS THEY REPLACED IT WITH ANOTHER UNIT THAT DID NOT SLIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18388858 ↗InjuryFeb 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPORTED THE MAYFIELD PEDIATRIC HORSESHOE HEADREST (A1051) WAS USED IN AN UNSPECIFIED SURGERY AND THE PATIENT DEVELOPED A PRESSURE ULCER ON THE FACE. THE PATIENT¿S PRESSURE ULCER HAS NOT DISAPPEARED FOR MORE THAN 2 YEARS AND HAS FILED A LAWSUIT AGAINST THE DISTRIBUTOR. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Ulcer
MAUDE report 13451425 ↗
Other HEADREST, NEUROSURGICAL devices
- NA Head SupportSteris Corporation · HEADREST, NEUROSURGICAL1 recall5 reports
- NA Head SupportSteris Corporation · HEADREST, NEUROSURGICAL1 recall1 report
- MAYFIELD® A1012Integra Lifesciences Corporation · HEADREST, NEUROSURGICAL6 reports
- DORO® Base Unit 3001-00Black Forest Medical GmbH · Headrest, Neurosurgical3 reports
- DORO® Adaptor 3002-00Black Forest Medical GmbH · Headrest, Neurosurgical2 reports
- MAYFIELD® A1112Integra Lifesciences Corporation · HEADREST, NEUROSURGICAL1 report
- MAYFIELD® A1011Integra Lifesciences Corporation · HEADREST, NEUROSURGICAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page