Omni-Tract® 11516Recalls & reported problems
Integra Lifesciences Corporation · Omni-Tract® Volkman Retractor 8-1/2", Hollow Handle, 4 Blunt Prongs · Class 1
All models of Omni-Tract (447)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 5, 2026Class IIOpen, Classified
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Z-2714-2026 · Root cause: Labeling design · Recall details · FDA record ↗
Other RETRACTOR, SELF-RETAINING devices
- Omni-Tract® 11306Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 11880Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 11481Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 3927Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 11560Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 11615Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 3720Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
- Omni-Tract® 3888Integra Lifesciences Corporation · RETRACTOR, SELF-RETAINING1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page