Integra® Ruggles®-Redmond™ RN6202Recalls & reported problems
Integra Lifesciences Production Corporation · Integra® Ruggles®-Redmond™ Rhoton Style Dissector, 7-1/2", 2mm Diameter, Style No. 2, Round · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10381780458906), received between Feb 26, 2022 and Jul 1, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Dull, Blunt1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2025Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT THE RUGGLES MICRO DISSECTOR ROUND 7 1/2 INCHES (RN6202) WAS "BLUNT AND NOT AS SHARP AS IT SHOULD BE". IT WAS REPORTED THAT THERE WAS AN INCREASE IN SURGERY TIME OF OVER 30 MINUTES, AS THE "DOCTOR HAD TO OPEN HIS OLD SET TO USE" TO FINALIZE THE SURGERY. NO PATIENT INJURY OR DEATH OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22378951 ↗InjuryFeb 26, 2022Break; Break
THE FOLLOWING WAS REPORTED VIA MEDWATCH MW5105526: ¿PATIENT UNDERWENT SURGERY FOR REMOVAL OF ACOUSTIC NEUROMA. DURING THE RESECTION OF THE TUMOR, THE TIP OF THE RHOTON DISSECTOR BROKE OFF AND WENT INTO THE SURGICAL FIELD. AFTER A CT DETERMINED THE LOCATION OF THE TIP. THE PATIENT RETURNED TO SURGERY ON (B)(6) 2021 TO REMOVE THE TIP FROM THE BRAIN. SMALL METALLIC FOREIGN BODY IS NOTED INTRACRANIALLY IN THE REGION OF THE RIGHT CEREBELLOPONTINE ANGLE. PATHOLOGY COMMENT AFTER REMOVAL- CONSISTS OF 0.2 CM IN GREATEST DIMENSION FLATTENED SOMEWHAT OVOID PORTION OF THE METALLIC FOREIGN MATERIAL.¿ IT WAS SUBSEQUENTLY REPORTED THAT FOLLOWING THE EVENT, THE SURGEONS INSPECTED THE SURGICAL FIELD AND DID NOT FIND THE BROKEN PIECE. AN INTRA-OPERATIVE X-RAY WAS PERFORMED, AND THE RADIOLOGIST RECOMMENDED A POST-OPERATIVE COMPUTERIZED TOMOGRAPHY (CT) SCAN WHICH THEN IDENTIFIED THE MISSING PIECE. ALL PARTS WERE COMPLETELY RECOVERED IN THE 2ND SURGERY ON (B)(6) 2021. IT WAS REPORTED THAT THERE WAS NO SURGICAL DELAY, AND AS OF (B)(6) 2021, THE PATIENT WAS RECOVERING AT HOME DUE TO HER UNDERLYING REASON FOR INITIAL SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 13609090 ↗
Other INSTRUMENT, MICROSURGICAL devices
- Synergetics 20.04.25SSynergetics USA, Inc. · INSTRUMENT, MICROSURGICAL5 reports
- Integra® Ruggles®-Redmond™ RN6206Integra Lifesciences Production Corporation · INSTRUMENT, MICROSURGICAL1 report
- Symmetry Surgical 57-2610Aspen Surgical Products, Inc. · Instrument, Microsurgical1 report
- Avalign VM82-7222Avalign Technologies, Inc. · INSTRUMENT, MICROSURGICAL1 report
- Integra® Jarit® 277274TIIntegra Lifesciences Production Corporation · INSTRUMENT, MICROSURGICAL1 report
- V. Mueller NL3785-009Steris Corporation · INSTRUMENT, MICROSURGICAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page