CODMAN® 801457Recalls & reported problems
Integra Lifesciences Production Corporation · CODMAN® Surgical Strips 3 1/2" x 6" (89mm x 152mm) · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780515135), received between Apr 19, 2022 and Apr 19, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Peeled/Delaminated” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K880402); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 20, 2024Open, Classified
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
Z-0256-2025 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2022Peeled/Delaminated; Peeled/Delaminated
A FACILITY REPORTED DELAMINATION OF THE SURGICAL STRIPS ((B)(4)) DURING PROCEDURE. THE NURSE MANAGED TO RETRIEVE ALL THE PIECE THAT CAME OFF. HOWEVER, THIS CAUSED AN IMPACT IN THE MANAGEMENT OF THE PATIENT BECAUSE IT WAS IMPOSSIBLE TO USE THIN PIECES (THEY SWELLED UP TOO MUCH ONCE PUT IN THE SALINE AND THEY SPLIT: DELAMINATE). THE EVENT LED TO 15 MINUTES SURGICAL DELAY, WITH NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14141531 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K880402. One submission often covers many models, so check the product and lot details in each record.
- Z-0095-2011 · Jul 16, 2010 · TerminatedSurgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card) Product code: 24-5403
- Z-0094-2011 · Jul 16, 2010 · TerminatedCodman 1/4 x 1/4" Surgical Patties (20 pouches of 10 patties per box) Product code: 80-1399
- Z-0093-2011 · Jul 16, 2010 · TerminatedCodman 5/16" Round Surgical Patties (20 pouches of 10 patties per box) Product code: 80-1396
Other Neurosurgical paddie devices
- CODMAN® 801400Integra Lifesciences Production Corporation · Neurosurgical paddie2 recalls9 reports
- CODMAN® 801407Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall6 reports
- MEDLINE NEUROSPNG05Medline Industries, Inc. · Neurosurgical paddie1 recall5 reports
- CODMAN® 801451Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall5 reports
- CODMAN® 801404Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall4 reports
- CODMAN® 801406Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall2 reports
- CODMAN® 801403Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall2 reports
- CODMAN 80-1407Cerenovus, Inc. · NEUROSURGICAL PADDIE1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.