CereLink™ 826854Recalls & reported problems

Integra Lifesciences Production Corporation · CereLink™ ICP Sensor Ventricular Catheter Kit with ICP Sensor · Class 2

All models of CereLink (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10381780520702), received between Mar 29, 2022 and May 8, 2026 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress2
  • No Device Output1
  • Nonstandard Device1
  • Obstruction of Flow1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Nonstandard Device

    A FACILITY REPORTED A CERELINK VENTRICULAR CATHETER WITH ICP SENSOR (ID 826854) GAVE AN ERROR MESSAGE ("SENSOR ERROR REPLACE¿). THE MONITOR AND THE CABLE WERE REPLACED, HOWEVER, REPLACING THE SENSOR FIXED THE PROBLEM. IT IN UNKNOWN IF THE EVENT HAPPENED DURING OR AFTER IMPLANTATION PROCEDURE. NO ADDITIONAL INFORMATION WAS RECEIVED AFTER SEVERAL ATTEMPTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25123265 ↗
  • InjurySep 12, 2022Obstruction of Flow; Obstruction of Flow

    A FACILITY REPORTED A CERELINK MONITOR GAVE ERROR THAT SENSOR OR PATIENT CABLE IS FAULTY. TROUBLESHOOTING WAS DONE AS PER IFU. DIFFERENT MONITOR WAS CONNECTED TO PROBE AND SAME ERROR PERSISTED. WHEN SURGEON REMOVED THE PROBE IN OPERATIVE ROOM IT BECAME APPARENT THAT THE VENTRICULAR DRAIN WAS OCCLUDED. A NEW SENSOR WAS INSERTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15406885 ↗
  • InjuryAug 17, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    A FACILITY REPORTED A CERELINK VENTRICULAR CATHETER SHOWED A KINK THEREFORE, NO MEASUREMENTS WERE POSSIBLE. THE CATHETER WAS EXPLANTED AND REPLACED ON (B)(6) 2022 TWO TO THREE DAYS AFTER IMPLANTATION. NO PATIENT SYMPTOMS DUE TO MALFUNCTION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15243365 ↗
  • InjuryJun 1, 2022No Device Output; No Device Output

    A FACILITY REPORTED A CERELINK ICP VENTRICULAR CATHETER WITH NO PRESSURE VALUE AFTER INSERTION AND MONITORING WAS NOT POSSIBLE. THE CATHETER WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14565607 ↗
  • InjuryMay 9, 2022Output Problem; Output Problem

    A FACILITY REPORTED NO VALUE OF A CERELINK ICP VENTRICULAR CATHETER. THE ICP MONITORING WAS DISCONTINUED, MORE MANIPULATION OF EVD SYSTEM. THE PATIENT IS BEING MONITORED WITH ANOTHER SYSTEM. THE CATHETER WAS REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 14337830 ↗
  • InjuryMar 29, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    A FACILITY REPORTED A CERELINK ICP VENTRICULAR CATHETER WAS KINKED AND THE SENSOR COULD NOT GIVE MEASUREMENTS. THE SENSOR WAS IMPLANTED ON (B)(6) 2022, EXPLANTED AND REPLACED ON (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13938920 ↗

Other Device, monitoring, intracranial pressure devices

All Device, monitoring, intracranial pressure devices →